26Q7873-SALIENT CHARACTERISTICS.pdf
PDF 217 KB Posted
- Attached to
- VISCOMETER
- Solicitation number
- SPMYM226Q7873
About this file
This document is a detailed service description for the procurement of one (1) automatic kinematic viscometer. The equipment must measure liquids with viscosity in the range of 0.5-10,000 centistokes (cSt) using a capillary viscometer submerged in a recirculating silicone bath. The instrument shall be calibrated at 40°C using certified viscosity standards traceable to NIST and must maintain temperature accuracy of less than ±0.02°C. Results must be reportable according to ASTM D445 standards.
The viscometer must be capable of measuring sample volumes less than 20 mL and feature a multiple sample carousel with sample vial washing capabilities including 2-solvent cleaning. The device shall use thermistor sensors for temperature measurement and must be operable both with and without a computer, as well as compatible with USB thumb drive data storage. Additionally, the equipment must have a unique identifier such as a serial number for tracking and identification purposes.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SPMYM226Q7873.pdf | ||
| 26Q7873_EXHIBIT B-CDRL-NRTL.pdf | ||
| 26Q7873_EXHIBIT C-CDRL-WARRANTY.pdf | ||
| 26Q7873_EXHIBIT A-CDRL-MANUALS.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
DETAILED DESCRIPTION OF SERVICE: PURCHASE OF ONE (1) AUTOMATIC KINEMATIC
VISCOMETER
THE EQUIPMENT AND ACCESSORIES MUST HAVE THE FOLLOWING SALIENT
CHARACTERISTICS:
1. Must be able to measure liquids with a viscosity in the range of 0.5-10,000 cSt
2. Must use a capillary viscometer submerged in a recirculating silicone bath.
3. Must be calibrated at 40°C using certified viscosity standards traceable to NIST.
4. Must have a temperature accuracy of less than ±0.02°C.
5. Must be able to report results according to ASTM D445.
6. Must be capable of measuring sample volumes less than 20 mL.
7. Must be usable with and without a computer.
8. Must be usable with a USB thumb drive.
9. Must have a multiple sample carousel.
10. Must have sample vial washing capabilities with 2 solvent cleaning.
11. Must use thermistor sensors.
12. Must have a unique identifier such as a serial number.
File details come from the government source that posted it. Updated .