26Q7873-SALIENT CHARACTERISTICS.pdf

PDF 217 KB Posted

Attached to
VISCOMETER
Solicitation number
SPMYM226Q7873
Issued by

About this file

This document is a detailed service description for the procurement of one (1) automatic kinematic viscometer. The equipment must measure liquids with viscosity in the range of 0.5-10,000 centistokes (cSt) using a capillary viscometer submerged in a recirculating silicone bath. The instrument shall be calibrated at 40°C using certified viscosity standards traceable to NIST and must maintain temperature accuracy of less than ±0.02°C. Results must be reportable according to ASTM D445 standards.

The viscometer must be capable of measuring sample volumes less than 20 mL and feature a multiple sample carousel with sample vial washing capabilities including 2-solvent cleaning. The device shall use thermistor sensors for temperature measurement and must be operable both with and without a computer, as well as compatible with USB thumb drive data storage. Additionally, the equipment must have a unique identifier such as a serial number for tracking and identification purposes.

View the file

Other files

File Type Posted
SPMYM226Q7873.pdf PDF
26Q7873_EXHIBIT B-CDRL-NRTL.pdf PDF
26Q7873_EXHIBIT C-CDRL-WARRANTY.pdf PDF
26Q7873_EXHIBIT A-CDRL-MANUALS.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DETAILED DESCRIPTION OF SERVICE: PURCHASE OF ONE (1) AUTOMATIC KINEMATIC

VISCOMETER

THE EQUIPMENT AND ACCESSORIES MUST HAVE THE FOLLOWING SALIENT

CHARACTERISTICS:

1. Must be able to measure liquids with a viscosity in the range of 0.5-10,000 cSt

2. Must use a capillary viscometer submerged in a recirculating silicone bath.

3. Must be calibrated at 40°C using certified viscosity standards traceable to NIST.

4. Must have a temperature accuracy of less than ±0.02°C.

5. Must be able to report results according to ASTM D445.

6. Must be capable of measuring sample volumes less than 20 mL.

7. Must be usable with and without a computer.

8. Must be usable with a USB thumb drive.

9. Must have a multiple sample carousel.

10. Must have sample vial washing capabilities with 2 solvent cleaning.

11. Must use thermistor sensors.

12. Must have a unique identifier such as a serial number.

File details come from the government source that posted it. Updated .