B09_ATCH_1_UPDATED_SALIENTS_S24019_RS5_11-27-23_0002.pdf

PDF 180 KB Posted

Attached to
CURRENT PROFILING SYSTEM IAW SALIENTS
Solicitation number
140G0124Q0021
Issued by

About this file

This document outlines the salient characteristics for a current profiling system for discharge measurements. The system must include a portable, self-contained, five-beam acoustic Doppler current profiler no larger than 9.5" x 2.2" and weighing less than 1.0 lb. It must have bottom tracking from 0.1 to 6 m, operate at 3 MHz, and include a vertical beam. The system shall include batteries, charger, cables, a GNSS rover smart antenna, tilt sensors, compass, and a ruggedized transport case. A flotation platform no larger than 2' x 3' and weighing less than 11 lbs is required along with Bluetooth connectivity up to 100 m. The system must be waterproof with an IP67 rating, measure velocities from +/- 5.0 m/s with 0.001 m/s resolution, sample at 1 Hz, and profile from 0.1 to 6 m with accuracy of 1% or better. It should operate from -5 to 45°C on internal batteries for up to 7 hours and measure 128 cells from 2.5-30 cm.

The related federal contract opportunity is solicitation 140G0124Q0021 from the Department of Interior's US Geological Survey Office of Acquisitions and Grants to procure a current profiling system meeting the outlined salient characteristics.

View the file

Other files

File Type Posted
Sol_140G0124Q0021_Amd_0002.pdf PDF
Sol_140G0124Q0021_Amd_0001.pdf PDF
Sol_140G0124Q0021.pdf PDF
B08_ATCH_3_FISMA_S24019_OAG_12_14_22.pdf PDF
B08_ATCH_1_SALIENTS_S24019_RS5.pdf PDF
B08_ATCH_2_508_S24019.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics S24019

CURRENT PROFILING SYSTEM FOR DISCHARGE MEASUREMENTS

Salient Characteristics

1 ea. Current Profiling System:

1. Shall consist of a portable, self-contained, five-beam, Broadband, Acoustic Doppler Current Profiling (ADCP) system for discharge measurements in rivers and streams.

2. The system shall be comprised of an ADCP, transport Bluetooth radio, GNSS rover, smart antenna ADCP, transport case, flotation platform with backpack carrying case, and accessories.

3. Shall include Broadband and Pulse-coherent acoustic processing technologies.

4. Shall have bottom tracking capability with a depth range of at least 0.1 to 6 m.

5. Shall include a 5-beam configuration, including a vertical beam, capable of operating at 3 MHz frequency.

6. The ADCP shall be no larger than 24 cm (9.5”) x 5 cm (2.2”) in size and weight no more than 0.45 kg

(1.0 lbs.).

7. System shall be supplied with all applicable accessories including, but not limited to, batteries, battery charger/wall adapter, cables (i.e. radio), and antenna.

8. Shall include GNSS rover smart antenna.

9. Shall include built-in temperature, 360 degrees magnetic compass, and 2-axis tilt sensors.

10. Shall include a ruggedized transport case for housing the ADCP, GNSS smart antenna, and accessories.

11. Shall include a flotation platform. Platform shall include facilities for mounting the ADCP and smart antenna. Shall be no larger than 2 ft. wide x 3 ft. long and weigh no more than 11 pounds.

12. Shall have integral Bluetooth low energy radio communications with a range of at least 100 m.

Bluetooth shall include an automatic reconnection feature with the ability to buffer data in the event of communication loss.

13. Shall have an Ingress Protection (IP) rating of at least IP67 (waterproof) or better.

14. Shall have a water velocity range from at least +/- 5.0 m/s.

15. Shall have a velocity resolution of 0.001 m/s, or better.

16. Shall have a sampling frequency of at least 1 Hz.

17. Shall have a profiling range from at least 0.1 to 6 m with an accuracy of 1% +/- 0.002 m/s, or better.

18. Shall be capable of operating over a temperature range from at least -5 to 45o C.

19. Shall be capable of operating on an internal rechargeable, replaceable battery with up to at least 7 hours operating life on a single charge.

20. Shall be capable of measuring at least 128 cells throughout the water column with a cell size of 2.5-30 cm.

File details come from the government source that posted it. Updated .