1333ND19QNB190478_-_Attachment_III_Statement_of_Work.pdf

PDF 95 KB Posted

Attached to
180L Liquid Nitrogen Dewars
Solicitation number
1333ND19QNB190478
Issued by

About this file

Statement of Work

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

13333ND19QNB190478

Attachment III – Statement of Work

BACKGROUND

The National Institute of Standards and Technology (NIST) Storeroom and Warehouse is a

11,500 square foot facility, with a 2,700 square foot sales floor, located in Building 301 on the

NIST Gaithersburg Campus. The Storeroom is a part of the NIST Facilities Services Division.

Their mission is to the best customer service plus the highest quality, variety and convenience at a competitive price.

The NIST campus has a fleet of approx. 100 NIST-owned Liquid Nitrogen Dewars in various sizes and ages that range from 25L-300L. The NIST Gaithersburg Main Storeroom is responsible for picking up, filling and delivering the various Liquid Nitrogen Dewars to the end users that use Liquid Nitrogen throughout the NIST Gaithersburg Campus.

Recently, NIST upgraded the NIST owned Liquid Nitrogen Dewar fleet to a universal sized

180L Liquid Nitrogen Dewar. These have quickly become the dewar of choice for most users within the NIST Gaithersburg community, as they allow for faster pick-up, delivery and filling as well as including a number of safety features.

PURPOSE

NIST is looking to purchase additional 180L Liquid Nitrogen Dewars, meeting the following specifications, to add to its existing fleet. All equipment shall be new and not refurbished and must become the property of the government.

Required Specifications:

Brand name or equal to Cryofab CLPB180, 180 Liter portable liquid nitrogen storage Dewar with 198 L gross capacity;

Dewar shall:

a. Be constructed of all “non-magnetic” stainless steel, including any post-annealing steps to remove induced magnetization due to fabrication;

b. Be mounted on square base with a pull handle;

c. Have two (2) 5-6 inch diameter fixed casters and two (2) 5-6 inch diameter swivel casters with wing style lock;

d. Have lifting rings or use existing reinforced oblong holes;

e. Have an outer diameter of approx. 24 inches and a height of approx. 53 inches;

f. Have an estimated empty weight of approx. 240 pounds;

g. Have a 0.5-inch top vertical liquid withdrawal port with quick coupling and ball valve;

h. Have a self-pressurizing system with regulator valve and pressure building coil;

i. Have 22 PSIG operating pressure;

ii. Have 25 PSIG M.A.W.P. (Maximum Allowable Working Pressure);

13333ND19QNB190478

Attachment III – Statement of Work

i. Have liquid and vent access ports pointed in opposite directions, and gas access port at 90 degrees to the vent/access ports;

i. Have a 0.5-inch male flare (CGA 295) on liquid outlet;

ii. Have low pressure liquid service only,

j. Have a liquid level gauge and pressure gauge to monitor system, as part of the standard plumbing package

k. Have a static evaporation rate of less than 2.8 L per day;

i. Have a vacuum jacket with a maximum pressure of 15 micrometers of Hg

ii. Have a vacuum jacket that maintains the above maximum pressure for a minimum of one year.

WARRANTY

A two-year parts, labor and shipping warranty shall be provided on new dewars only.

Shipping charges both to and from the manufacturer shall be covered under warranty.

The government would also like a one-year extended warranty past the one-year standard warranty to include parts, labor and shipping warranty shall be provided on new dewars only. Shipping charges both to and from the manufacturer shall be covered under extended warranty.

DELIVERY

National Institute of Standards and Technology

100 Bureau Drive

Bldg. 301, Loading Dock

Gaithersburg, MD 20899

PERIOD OF PERFORMANCE

The period of performance will be three years from the date of award.

File details come from the government source that posted it.