This document contains questions and answers regarding specifications for a benchtop centrifuge being procured by the Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Midwest Area under solicitation number 12505B23Q0179. The centrifuge must be able to spin at approximately 25,000 x g with fixed angle rotors and 7,000 x g with swinging rotors. It must include adapters for 96-well plates in both standard and deep well formats as well as 15mL and 50mL conical tubes. While adapters for 150mL bottles are preferred, quotes should specify if this capacity is not available. Both angle and swinging rotors must be supported to accommodate different tube sizes.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Statement of Work.pdf | ||
| Provisions and Clauses.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Ques�ons and Answers
Q: Do you have a loading dock?
A: Yes
Q: Do you require li�gate service?
A: No
Q: How fast does the centrifuge need to spin?
A: approximately 25,000 x g with fixed angle and 7,000 x g with the swinging rotors
Q: Do you need angle rotor or swing rotor?
A: Both poten�ally since the larger tubes likely can’t be swing rotors
Q: Is this only for 1 centrifuge?
A: Yes
Q: The request men�oned different capaci�es (96 deep well, 96 well standard), adapters for 15 and 50 mL conical tubes, as well as 150 mL botles. What capacity are you looking for in the single centrifuge if it’s just want? (Considering that 150 mL botles is 10 �mes of 15ml tube capacity)
A: We need adapters for all of the following tubes listed. If we can’t get an adapter for 150mL botle with a par�cular unit then please state that when the quote is submited. We’ll need to assess whether we can find an alterna�ve for our 150mL botles. However, rotors for 96 well plates (standard and deep well) as well as 15 and 50mL tubes must be included for the benchtop centrifuge.