Pre-Solicitation Notice
Project No. 528A8-18-806 Install dedicated feeder for MRI upgrade
Stratton VAMC
THIS PRE-SOLICITATION NOTICE IS TO INFORM YOU THAT THE STRATTON VA MEDICAL CENTER, LOCATED AT 113 HOLLAND AVE, ALBANY, NY 12208 WILL BE POSTING A SOLICITATION ON OR ABOUT AUGUST 13, 2021 FOR THE INSTALL DEDICATED FEEDER FOR THE MRI UPGRADE PROJECT.
PROJECT DESCRIPTION
The Stratton VA Medical Center (VAMC) in Albany, NY is replacing its existing 1.5T MRI machine. Philips Healthcare is contracted with the Stratton VA Medical Center to furnish and install an Ingenia Ambition 1.5T X MRI. The VA has contracted Construction Technology Group (CTG) to conduct most of the site prep for the new MRI, including installing a new circuit breaker for the MDU. However, CTG s contract does not cover Philips requirement of isolating the feeder for the MRI s MDU from existing Panel, NPP-1.
Per Philips Pre-Order Site Preparation Support Document, the MDU requires the following:
Supply Configuration
3 phase, 3 wire power, unity ground, and bonded ground
Nominal Line Voltage
480 VAC, 60 Hz
Branch Power Requirement
80 kVA
Circuit Breaker
3 pole, 100 Amp (480 VAC)
This procurement is 100% set-aside for Service Disabled Veteran Owned Small Businesses. The NAICS code is 236220. The small business standard is $39.5 Million. The magnitude of construction is between $100,000 and $250,000. A 20% Bid Bond must be submitted with the bid. Information regarding the scheduled site visit will be stated in the Solicitation.
Solicitation materials consisting of the Information for Bid Details, Specifications, Drawings and Contract Forms will be available on https://beta.sam.gov in electronic format only.
The POC for this project will be Tony Clemente and can be reach at 518-258-5483 or email at Tony.Clemente@va.gov.
Attachments
| File | Type | Posted |
|---|---|---|
| 36C24221B0066.docx | DOCX document |
Notice history
| Notice | Type | Posted |
|---|---|---|
| Z1DA--528A8-18-806 replace 1. 5T MRI-install dedicated feeder | Award Notice | |
| Z1DA--528A8-18-806 replace 1. 5T MRI-install dedicated feeder | Solicitation | |
| Z1DA--528A8-18-806 replace 1. 5T MRI-install dedicated feeder | Pre-Solicitation |
On GovTribe
Work this opportunity on GovTribe
- Track it in your pipeline
- Find teaming partners
- Similar opportunities
- Ask GovTribe AI about this opportunity