36C24221B0066.docx
DOCX document 18 KB Posted
- Attached to
- Z1DA--528A8-18-806 REPLACE 1. 5T MRI-INSTALL DEDICATED FEEDER Federal contract opportunity
- Solicitation number
- 36C24221B0066
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Pre-Solicitation Notice Project No. 528A8-18-806 Install dedicated feeder for MRI upgrade Stratton VAMC
| SUBJECT* |
| 528A8-18-806 REPLACE 1.5T MRI-INSTALL DEDICATED FEEDER |
GENERAL INFORMATION
| CONTRACTING OFFICE’S ZIP CODE* |
| 12208-3410 |
| SOLICITATION NUMBER* |
| 36C24221B0066 |
| RESPONSE DATE/TIME/ZONE |
| EASTERN TIME, NEW YORK, USA |
| ARCHIVE |
| 90 DAYS AFTER THE RESPONSE DATE |
SET-ASIDE
| PRODUCT SERVICE CODE* |
| Z1DA |
| NAICS CODE* |
| 236220 |
| PLACE OF PERFORMANCE |
| Department Of Veterans Affairs |
Stratton VA Medical Center
113 Holland Ave
Albany NY
| POSTAL CODE |
| 12208 |
COUNTRY
CONTACT INFORMATION
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs Network Contracting Office 2 20 Madison Ave Ext Albany NY 12203
POINT OF CONTACT*
ADDITIONAL INFORMATION
| AGENCY’S URL |
| http://vaww.visn2.va.gov/albany/index.asp |
| URL DESCRIPTION |
| Department Of Veterans Affairs |
| AGENCY CONTACT’S EMAIL ADDRESS |
| Tony.Clemente@va.gov |
| EMAIL DESCRIPTION |
| Contracting Officer |
Presolicitation Notice Presolicitation Notice
DESCRIPTION
| *=Required Field |
| Presolicitation Notice |
Presolicitation Notice
THIS PRE-SOLICITATION NOTICE IS TO INFORM YOU THAT THE STRATTON VA MEDICAL CENTER, LOCATED AT 113 HOLLAND AVE, ALBANY, NY 12208 WILL BE POSTING A SOLICITATION ON OR ABOUT AUGUST 13, 2021 FOR THE INSTALL DEDICATED FEEDER FOR THE MRI UPGRADE PROJECT.
PROJECT DESCRIPTION
The Stratton VA Medical Center (VAMC) in Albany, NY is replacing its existing 1.5T MRI machine. Philips Healthcare is contracted with the Stratton VA Medical Center to furnish and install an Ingenia Ambition 1.5T X MRI. The VA has contracted Construction Technology Group (CTG) to conduct most of the site prep for the new MRI, including installing a new circuit breaker for the MDU. However, CTG’s contract does not cover Philips’ requirement of isolating the feeder for the MRI’s MDU from existing Panel, NPP-1.
Per Philips’ Pre-Order Site Preparation Support Document, the MDU requires the following:
· Supply Configuration
· 3 phase, 3 wire power, unity ground, and bonded ground
· Nominal Line Voltage
· 480 VAC, 60 Hz
· Branch Power Requirement
· 80 kVA
· Circuit Breaker
· 3 pole, 100 Amp (480 VAC) This procurement is 100% set-aside for Service Disabled Veteran Owned Small Businesses. The NAICS code is 236220. The small business standard is $39.5 Million. The magnitude of construction is between $100,000 and $250,000. A 20% Bid Bond must be submitted with the bid. Information regarding the scheduled site visit will be stated in the Solicitation.
Solicitation materials consisting of the Information for Bid Details, Specifications, Drawings and Contract Forms will be available on https://beta.sam.gov in electronic format only.
The POC for this project will be Tony Clemente and can be reach at 518-258-5483 or email at Tony.Clemente@va.gov.
Page 1 of Page 1 of
File details come from the government source that posted it. Updated .