About this opportunity

The Department of Veterans Affairs Veterans Health Administration (VA VHA) is seeking to procure a Thyroid Uptake System with Well Counting System for the John J. Pershing VA Medical Center in Poplar Bluff, Missouri. The medical spectrometer is a multi-purpose spectrum analysis instrument designed for nuclear medicine applications, specifically supporting bioassay programs for measuring staff thyroid burden and performing various radiological tests. The system must be capable of measuring isotopes such as I-123, I-125, and I-131, with advanced reporting capabilities including automatic calculations, patient information storage, and customizable reporting. The solicitation requires the system to be compatible with Windows 10, meet DICOM standards, and comply with VA technical requirements. The proposal responses are due by May 16, 2025, with a total delivery timeline of 30 days from the award notice.

This contract opportunity is set aside exclusively for Service-Disabled Veteran-Owned Small Businesses (SDVOSB) under NAICS code 334510 with a 1,250-employee size standard. The procurement is classified under the Electromedical and Electrotherapeutic Apparatus Manufacturing category, with a Product Service Code of 6515 for medical and surgical instruments. The system will be installed in the Nuclear Medicine Department at the Poplar Bluff VAMC, with requirements including a minimum 2-inch well container, a 34-inch vertical range of travel, and a 220-degree collimator rotation. The contractor must provide initial installation training, two operator's manuals, two service manuals, and offer a one-year warranty with two years of parts and labor coverage. The total system weight must not exceed 350 pounds, and it must operate on 120 VAC, 60 Hz power with a 2 AMP circuit breaker.

Notice text

STATEMENT OF REQUIREMENT
A. GENERAL INFORMATION
The Nuclear Medicine service at John J. Pershing VA Medical Center (VAMC) in Poplar Bluff, MO will better serve its Veterans by acquiring this Thyroid Uptake System with Well Counting System with 2 Inch Well Shielding, a multi-purpose spectrum analysis instrument designed for diverse nuclear medicine applications. The system performs Uptake studies, wipe tests, and other user-defined tasks that are accomplished with speed and precision using this fully integrated computer-controlled instrument and its comprehensive selection of application software.
Engineered for mobility, durability and operational efficiency, the unit handles clinical tests, safety compliance tasks, and system administration procedures quickly and accurately. Built-in menus follow conventionally established nuclear medicine procedures, providing step-by-step guidance throughout defined procedures. The system automatically performs calculations, stores patient information and test results, and outputs clear, concise reports. User-defined uptake protocols can be initiated.
B. SCOPE
Delivery of the Thyroid Uptake system will allow a bioassay program for quick and efficient measurement of the staff thyroid burden for radiopharmaceuticals, I-123, I-125 and I-131. Measurement results can be reported in cpm, dpm, uCi, or Bq. Action levels may be set by the user for each isotope. System reporting facilitates single bioassay report, summary of multiple bioassays on an individual staff member, and summary report on multiple staff members. Reports are easily printed or saved as PDF files
The Thyroid Uptake program supports multiple time-stamped uptake measurements, auto decay correction, user-defined uptake protocols, on-screen spectrum acquisition and analysis. The system provides reports that include normal ranges and allow notes/comments entered by the facility physician and technologists.
The wipe test application automatically calculates net contamination levels and reports results in dpm, uCi and Bq. The program allows compliance with regulations for counting and reporting results of wipes taken for contamination surveys. The system allows user to define an unlimited number of “Area” names and wipe “Locations” entered within the areas for testing and documentation. Each area may be designated as “restricted”, “unrestricted”, “package” or “sealed source”. The user may specify count times, activity units, wide window, and individual isotopes and geometric efficiencies or simply utilize the wide range of factory default options.
The program displays and stores the full spectrum with each wipe, automatically calculates net contamination levels and reports results in dpm, uCi or Bq. Wipes above the user defined trigger levels are easily identified. A detailed spectrum analysis may
also be performed for any wipe to help identify isotopes causing contamination. Customized reports can be printed or saved in PDF format.
The application specialist will provide training and guidance to ensure safe use of the equipment and the study and administration applications provided. The contractor will provide cybersecurity best practices information and training, when applicable.
The contractor will provide application update to ensure information security in the current equipment and other factors that interconnects with the imaging system including the network, other current peripheral applications and other new tools and software that will be developed or installed to work with the current work station and imaging equipment hardware and software.
C. REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS
Common Nomenclature: Medical Spectrometer Hardware
Kind of material: ALL IN ONE Touch Screen Computer
Electrical Data:
Power: Low voltage transformer with hospital grade power cord and plug.
Line Voltage: 120 VAC, 60 Hz, 2 AMP circuit breaker; 230 VAC, 50 Hz, 1 AMP circuit breaker
Dimensions, size, capacity:
Minimum articulating arm with a minimum of 34-inch vertical range of travel with a minimum of 220-degree collimator rotation.
Principles of operation:
Multi-Channel Analyzer:
Channels: 1024
Spectral Resolution: FWHM 8%
Count Rate: (Maximum) 150,000 cps for Tc-99m
Count Rate Stability: 99%
Gross Count Rate Linearity: Within 5% up to 150,000 cps
Pulse Height Linearity: Within 2% (independent of detector)
Detector High Voltage Adjustment: Automatic adjustment for both probe and well; uses 10 μCi Cs-137 as the calibration source.
Restrictive environmental conditions: User Set: Unlimited user defined Isotopes, setting ROI, half-life, name, efficiency and range.
Intended use:
Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, F-18, Fe-59, Ga-67, Ga-68, Gd-153, Ge-68/Ga-68, Hg-197, I-123, I-125, I-131, In-111, Ir-192, K-42, Lu-177, Na-22, Na-24, Pd-103, Ra-223, Se-75, Sr-85, Tc-99m, Tc-99m/Tl-201, Tl-201, Yb-169, Wide Window.
Equipment with which the item is used, Job related tools:
Probe: 2" x 2" Nal (Tl) integral line scintillation detector
Other pertinent information that describes the item, material or service required:
Uptake Stand:
Dimensions:
Arm down and retracted: 34" l x 26" w x 55" h (86.4 x 66 x 139.7 cm)
Maximum Height: 60" (152.4 cm)
Maximum Length: Arm Extended: 61" (154.9 cm)
Collimated Shield: Flat field collimator meeting IAEA specifications
Positioning: Distance bar and positioning light
Casters: 4" Total locking
Maximum weigh of 300 lbs.
A maximum combined weight with Well Counter: 350 lbs.
Power: Low voltage transformer with hospital grade power cord and plug.
Line Voltage: 120 VAC, 60 Hz, 2 AMP circuit breaker; 230 VAC, 50 Hz, 1 AMP circuit breaker
OTHER UNIQUE REQUIREMENTS: Minimum of two-inch well container
Compatibility with Existing Equipment: TRM Compliance with VA Directives
WINDOWS 10 compatible or Newest Windows Version compliance with VHA standards.
Must be DICOM compliance, Directives or IT Policies, Required.
Coordination with On-going Construction Project (provide the project number): N/A
Warranty: 1 year from date of receipt and verification of installation/operation. Two years part/labor warranty
Training: Initial installation and launch training staff and service support
Service and/or Maintenance: Preventive maintenance and training support, to include two Operating Manuals and two Service Manuals
Safety/ISO/IT/BioMed/SPS & Other Departmental Compliance: Must meet Technical Reference Model (TRM) VA/BioMed Compliance.
D. DELIVERY
Contractor must deliver all components of the system within 30 days of award notice.
Standard FOB Destination to:
John J. Pershing VA Medical Center
1500 North Westwood Blvd
Poplar Bluff MO 63901
Government-Furnished Property, Material, Equipment, or Information (GFP, GFM, GFE, or GFI) N/A
SECURITY N/A
SPECIAL MATERIAL REQUIREMENT: Two Operator’s Manuals and two Service Manuals
E. PLACE OF PERFORMANCE
The system will be installed, serviced, and updated at the facility, John J. Pershing VAMC, Poplar Bluff, MO, Nuclear Medicine Department. Servicing, training, and support of personnel will be per stipulations on the final contract.

Attachments

Files attached to this notice, newest first
File Type Posted
36C25525Q0328.pdf PDF

Notice history

Notices posted for this opportunity, newest first
Notice Type Posted
Thyroid Uptake System, Poplar Bluff, MO VAMC This notice · Latest solicitation Solicitation
Thyroid Uptake System, Poplar Bluff, MO VAMC Original Pre-Solicitation

On GovTribe

Work this opportunity on GovTribe

  • Track it in your pipeline
  • Find teaming partners
  • Similar opportunities
  • Ask GovTribe AI about this opportunity