Thermal Stimulator with MRI Compatible Thermodes
Closed Pre-Solicitation Posted
- Solicitation number
- 75N95021Q00397
- Agency
- National Institute on Drug Abuse National Institutes of Health, Department of Health and Human Services
- Responses due
- Set-aside
- No set-aside
Opportunity facts
- NAICS code
- 811219 Other Electronic and Precision Equipment Repair and Maintenance
- Place of performance
- Baltimore, Maryland 21224, United States
Notice details come from SAM.gov. Updated .
Notice text
The translational addiction medicine branch and neuroimaging research branch of the NIDA intramural research program (NIDA-IRP) needs to purchase an MRI compatible thermal stimulator for protocols studying pain.
Salient characteristics:
- be a thermal stimulator designed for pain research.
- be FDA approved.
- include the capability of MRI-compatible thermodes with any/all associated extra hardware.
- include or be compatible with non-MRI thermodes for use in purely behavioral experiments.
- be two channels so that we can perform pain discrimination experiments.
- Be fully controllable remotely through MATLAB.
- have a temperature range of at least 0C-50C, with a minimum thermal slew rate of 10C/sec. It must be accurate to at least +/- 0.5C and be controllable to at least 0.1C.
- be able to be calibrated locally by NIDA personal.
- have a TTL compatible output that we can integrate with our existing physiology recording systems.
Includes at least an eighteen-month manufacturer’s warranty for parts and labor