High Field Split-Coil ElectroMagnet

Awarded Award Notice Posted

Solicitation number
SB1341-15-RQ-0468
Agency
National Institute of Standards and Technology Department of Commerce
Awarded
to GMW Associates
Set-aside
No set-aside

Opportunity facts

Contract number
SB1341-15-SU-0762 Federal contract award
NAICS code
334516 Analytical Laboratory Instrument Manufacturing
PSC
Not on record
Place of performance
100 Bureau Drive Gaithersburg, Maryland 20899, United States

Notice details come from SAM.gov. Updated .

Notice text

3 versions

Update #3 · Latest ·

Added: Jun 11, 2015 3:57 pm AMENDMENT A002:

The purpose of this amendment is to answer questions:

(1) Are iron pole and yoke part of 200 kg?

Mass limit applies to the entirety of the magnet system, including iron pole and yoke,  and any cooling mechanism for the magnet, which shall be included in the total cost of the system. It does not apply to the sample cryostat,  which we are not asking for in this solicitation.


(2)Should the close cycle refrigerator for magnet be included in the system quotation?  Is the weight of cryocooler for magnet, included in 200 kg?


Mass limit applies to the entirety of the magnet system, including mass of the magnet cryocooler.


(3) There is no heat leak and dimension specification for magnet cryocooler, such cryocooler appears not part of cost or delivery?

Cryocooler is part of the magnet system. 

 With regard to dimensions, see MINIMUM REQUIREMENTS 1.A.b-d.

In this section "magnet" is meant to mean the entirety of the magnet system, including magnet cryostat. Our own sample cryostat (which is not part of this solicitation) is not to be considered in these dimensions.


(4) Cryocooler for superconducting magnet is not specified, and the crycoolers for sample is not specified either.  Please specify coolers and and whether magnet coolers is part of deliverables.


A separate cooling system is in place for samples, and does not factor into the mass limit. The magnet cryocooler is not to be combined with the sample cryocooler, meaning combined heat load is not a consideration. Cooling is a of magnet coils is a deliverable (see section MINIMUM REQUIREMENTS 1.A.l). Pulse tube coolers are necessary.

All other terms remain unchanged.


 

Update #2 ·

Added: Jun 02, 2015 5:08 pm AMENDMENT A001:

THE PURPOSE OF THIS AMENDMENT IS TO INCORPORATE ATTACHMENT NAMED "BEAMLINE SETUP". THE ATTACHMENT IS HEREBY PART OF THIS SOLICITATION AND IS ATTACHED.

ATTACHMENTS:

The attached pictures and/or drawings are hereby provided as attachments. The purpose of these attachments is for reference and to provide specification requirements to the Contractor under this solicitation.


All attachments included in this solicitation are owned by the Government and shall not be reproduced, released, distributed, or otherwise utilized for any purpose other than submission of a quotation and the execution of the required work without the prior written approval of the NIST Contracting Officer.


Attachment # 1: Beamline Setup

All other terms remain unchanged.

Update #1 ·

Added: May 29, 2015 12:07 pm See attached Combined Synopsis/Solicitation

Attachments

Files attached to this notice, newest first
File Type Posted
beamlinesetup.pdf PDF
FINALSOLICITATION.pdf PDF