Specifications: Water Jet
· Environment: 50-90 °F (10-32 °C) ambient temperature with 95% or less humidity. Recommended slab requirement: 4 inch minimum slab thickness, 3000 PSI concrete with steel reinforcement. For certain machine table models, the floor flatness underneath machine needs to be flat within 1 inch.
· Cutting Footprint Minimum: 10’6”x6’6” (Length x Width)
· Desired Cutting Head Style: Carriage
· Air Supply:
· Pressure – 75-120 psi (517-827 kPa)
· Flow – 16.0 cfm (27 Nm3/hr) Minimum
· Air Quality: Clean and Dry
· Water Supply (Each Pump:
· Temperature – Required: 45-70oF (7-21oC) Recommended: 45-60oF (7-16oC)
· Pressure/Flow
· Pressure (w/ Reverse Osmosis System) – 50-90 psi (345-620 kPa)
· Flow – 3.2 gpm (12.1 l/m) Minimum
· Water Quality
· pH Level – 6.5-8.5
· Impurity Content Limits
· Total Dissolved Solids - < 250 ppm
· Calcium - < 17 ppm
· Magnesium - < 6 ppm
· Iron - < 0.3 ppm
· Manganese - < 0.05 ppm
· Chloride - < 100 ppm
· Sulfate - < 200 ppm
· Silica - < 10 ppm
· The machine requires water to be filtered at a very high level and contains fine filter to do so. Additional water filtering may be required to prevent premature clogging of these filters if the incoming water is especially dirty.
· Drain Flow (Each Pump): 5 gpm (19 l/m) Minimum
· Electrical Grounding: The machine is provided with a ground strap that must be attached to a recognized ground.
· System Electrical:
· Three Phase Electrical Requirements
· 40 HP Pump
· 80A, 3 Phase, Mains Disconnect
· 80A, 3 Phase, Pump Disconnect (Optional)
· 36000 BTU Chiller
· 460V, 12A, 3 Phase Mains Disconnect
· Closed Loop System
· 460 VAC, 3 Phase, Disconnect
· Single Phase Electrical Requirements
· Table Controller
· 230/115 VAC, 10/20A. Must have GFCI Protection
· Reverse Osmosis System (optional)
· 230/115 VAC, 10/20A
· Solids Removal System (optional)
· 200-240 VAC, 10A. Must have GFCI Protection
· Water Softener
· 115 VAC, 10A
· Laminar Filter
· 230/115 VAC, 5/10A. Must have GFCI Protection