This document provides supplemental information for a federal solicitation seeking a total organic carbon analyzer. The solicitation requires an analyzer with high temperature combustion catalytic oxidation and non-dispersive infrared detection. The analyzer must be standalone or PC-controlled with a printer. It must also include an autosampler with a minimum 50 sample capacity using commercially available vials. The analyzer is required to analyze in total carbon, inorganic carbon, particulate organic carbon, and non-purgeable organic carbon modes with a measurement range of 0.050 mg/L to 1000 mg/L and reproducibility of 1.5% maximum coefficient of variation. Additional required capabilities include automatic dilution, up to 3 calibration curves, variable injection volumes from 10mL to 1000mL, automatic sample acidification and sparging, use of air or oxygen as carrier gases, data storage for 1000+ samples and 10+ calibration curves, and carrier gas problem detection. The related federal solicitation is issued by the Defense Logistics Agency Land and Maritime seeking a TOC-LSCH system.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SPMYM223Q2190 TOC-LCSH System.pdf | ||
| SSJM 1_Redacted.pdf | ||
| CDRL Exhibits A-C_Redacted.pdf | ||
| Performance Work Statement.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
JML# 31408700 THRU 8708 - SUPPLEMENTAL INFORMATION
DETAILED DESCRIPTION OF SERVICE: PURCHASE OF A TOTAL ORGANIC CARBON ANALYZER
THE EQUIPMENT AND ACCESSORIES MUST HAVE THE FOLLOWING SALIENT CHARACTERISTICS:
- HIGH TEMPERATURE COMBUSTION CATALYTIC OXIDATION WITH NON-DISPERSIVE
INFRARED (NDIR) DETECTION
- STANDALONE OR PC-CONTROLLED WITH PRINTER
- AUTOSAMPLER FOR AQUEOUS SAMPLES, WITH 50 SAMPLES MINIMUM CAPACITY, USING
COMMERCIALLY AVAILABLE VIALS
- ANALYZE IN TC, IC, POC, AND NPOC MODES
- MEASUREMENT RANGE TC/IC FROM 0.050 mg/L TO 1000 mg/L
- REPRODUCIBILITY TC/IC/NPOC CV=1.5%MAX
- MATRICE COMPATIBLE FROM PURE WATER TO INDUSTRIAL WASTEWATER AND
SEAWATER
- AUTOMATIC DILUTION, TO REDUCE SAMPLE ACIDITY, ALKALINITY, AND SALINITY,
EXTENDING THE PERIOD OF USE OF CATALYSTS AND COMBUSTION TUBE, AND
EXTENDING MEASUREMENT RANGE.
- UP TO 3 CALIBRATION CURVES CAN BE SET FOR SAMPLE MEASUREMENTS
- VARIABLE INJECTION VOLUMES 10ΜL TO 1000ΜL
- IC PRETREATMENT, SAMPLE ACIDIFICATION AND SPARGING BOTH PERFORMED
AUTOMATICALLY INSIDE THE ANALYZER
- CARRIER GAS: AIR OR OXYGEN
- 120 VAC
- NUMBER AND TYPE OF DATA SAVED: ALL MEASUREMENT RESULTS INCLUDING PEAK
DATA FOR 1,000 OR MORE SAMPLES, AS WELL AS 10 OR MORE CALIBRATION CURVES
- RINSING WATER: CAPABLE OF REPLACEMENT RINSING BY SAMPLE (RINSING WITH
SAMPLE), AND CLEANING BY RINSING WATER (ALSO USED AS DILUTION WATER)
- BLANK CHECK FUNCTION EVALUATES SYSTEM BLANKS BY MEASURING ULTRAPURE
WATER PROCESSED AUTOMATICALLY WITHIN THE INSTRUMENT
- CARRIER GAS PROBLEM DETECTION FUNCTION AVAILABLE