This statement of work outlines preventative maintenance and emergency repair services required for a BD FACSymphony A5 Cell Analyzer located at the USDA Agricultural Research Service in Ames, Iowa. The contractor shall provide two annual preventative maintenance visits including replacement parts, unlimited emergency and telephone support, and repair services using factory-trained engineers and original equipment manufacturer parts. The base year of performance is January 20, 2024 through January 19, 2025, with four optional one-year renewal periods. Preventative maintenance services shall include inspection, cleaning, alignment, calibration and repair of all covered components. Emergency repairs shall be completed within 48 hours. Two service visits will be provided annually with all work coordinated with the USDA site contact. Only factory-trained technicians are permitted to service the equipment.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Provisions Clauses Full and Open.pdf | ||
| Wage Determination.pdf | ||
| Provisions Clauses.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work (SOW)
BD FACSymphony A5 Cell Analyzer - SN: R66096400079
GENERAL INFORMATION
1.0 Scope of Work: Contractor will provide factory OEM trained engineers to perform two (2) preventive maintenance on site service visits and unlimited emergency visits on the BD FACSymphony A5 Cell Analyzer located at 1920 Dayton Ave. in Ames, IA.
Covered components will include:
1) BD FACSymphony A5 Cell Analyzer (SN: R66096400079)
2) One Computer Workstation
3) BD FACSFlow Supply System (SN: N3492775352)
4) All OEM replacement parts
5) All labor and travel expenses
6) Two (2) Annual Preventative Maintenance with PM kits
2.0 Background: Preventive maintenance has been procured for this instrument since it was purchased to ensure continued & reliable operation in support of the USDA- ARS-NADC research mission to be billed Quarterly in arrears.
The intent is to award a contract of a Base Year plus 4 Option Years.
Base Year – 1/20/24-1/19/25 Option Year 1 – 1/20/25-1/19/26 Option Year 2 – 1/20/26-1/19/27 Option Year 3 – 1/20/27-1/19/28 Option Year 4 – 1/20/28-1/19/29
Award will be made for a 1-year period with four optional 1-year annual renewal periods.
Renewal decisions will be made two months prior to the potential renewal date. The decision to renew will be at the sole discretion of the USDA.
CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:
Service tasks include all OEM ST Total Level coverage, but are not limited to:
1) 2 Preventative Maintenance Inspections to be performed during a 12 month period including 2 PM kits
2) Labor, travel expenses and parts (excluding consumables) are covered to include the five lasers: 355-60 w/BD UV Laser, 405—100 LX Violet Laser, 488-150 LX Blue Laser, 561 -100 LS Y/G laser, 637-140 LX Red Laser along with cell analyzer, computer workstation, and supply system
3) Software Revisions released during the agreement term
4) Unlimited service visits, Monday –Friday, excluding holidays
5) Unlimited telephone support for instrument, reagents, applications
6) 48 hour guaranteed response to a request for emergency on-site service
7) On-Site Repairs by certified original equipment manufacturer (OEM)
8) All repairs done with OEM Parts
9) All parts (including lasers) required for repair are covered except for consumables
10) Maintenance done in accordance with OEM specifications
11) Perform check of all portions of the unit that could require replacement parts during preventative maintenance visits per yearly contract.
12) Perform functional checks on system operation during preventative maintenance visits per yearly contract.
13) Execute all functional tests during preventative maintenance visits per yearly contract.
14) Completion of OEM ST Total Level PM Checklist once PM has been completed – provide to end-user for system records.
Contractor will provide telephone support and unlimited emergency visits to the customer location. Repair visits include factory OEM certified parts, factory OEM certified engineers and travel as applicable.
4.0 Government Furnished: Government-furnished support is limited to security access to the equipment in the NADC facility, replacement parts not otherwise covered under this agreement, AC power (110VAC) and standard office lighting.
5.0 Deliverables / Schedule:
Key Deliverables
Item No.
Deliverable
Objective
Due
1 Preventive Maintenance
Evaluate the instrument and provide preventative maintenance kit. Routine inspection, cleaning, alignment/calibration, lubrication and repair of covered OEM components including OEM lasers.
BASE YEAR: Once annually coordinated through the BD FACSymphony Operator
2 Emergency Maintenance
Emergency repair of defective components with OEM parts to bring the unit back on-line.
BASE YEAR: As necessary – coordinated through the BD FACSymphony Operator
3 Preventive Maintenance
Evaluate the instrument and provide preventative maintenance kit. Routine inspection, cleaning, alignment/calibration and repair of covered OEM components including OEM lasers
OPTION YR 1:
Once annually coordinated through the BD FACSymphony Operator
4 Emergency Maintenance
Emergency repair of defective components with OEM parts to bring the unit back on-line.
OPTION YR 1: As necessary – coordinated through the BD FACSymphony Operator
5 Preventive Maintenance
Evaluate the instrument and provide preventative maintenance kit. Routine inspection, cleaning, alignment/calibration, lubrication and repair of covered OEM components including OEM lasers.
OPTION YR 2:
Once annually coordinated through the BD FACSymphony Operator
6 Emergency Maintenance
Emergency repair of defective components with OEM parts to bring the unit back on-line.
OPTION YR 2: As necessary – coordinated through the BD FACSymphony Operator
7 Preventive Maintenance
Evaluate the instrument and provide preventative maintenance kit. Routine inspection, cleaning, alignment/calibration, lubrication and repair of covered OEM components including OEM lasers.
OPTION YR 3:
Once annually coordinated through the BD FACSymphony Operator
8 Emergency Maintenance
Emergency repair of defective components with OEM parts to bring the unit back on-line.
OPTION YR 3: As necessary – coordinated through the BD FACSymphony Operator
9 Preventive Maintenance
Evaluate the instrument and provide preventative maintenance kit. Routine inspection, cleaning, alignment/calibration, lubrication and repair of covered OEM components including OEM lasers.
OPTION YR 4:
Once annually coordinated through the BD FACSymphony Operator
10 Emergency Maintenance
Emergency repair of defective components with OEM parts to bring the unit back on-line.
OPTION YR 4: As necessary – coordinated through the BD FACSymphony Operator
6.0 Travel: Two (2) preventive maintenance service visit will be provided by the contractor. All service visits will be coordinated with the BD FACSymphony A5 Operator at NADC.
7.0 Contractor’s Key Personnel: Only factory OEM trained engineers are to perform any service procedure on the BD FACSymphony A5 Cell Analyzer.
8.0 Security Requirements: A copy of an official government photo ID (State-issued driver’s license or ID card, passport, green card, etc.) must be provided no later than one week before arrival of contract service technician(s) except in the case of emergency service, where only the name of the technician provided via email or telephone will be necessary. Security access to the NADC facility will be provided based on this identification requirement.
9.0 Place of Performance: USDA-NADC, 1920 Dayton Avenue, Ames, IA 50010.
10.00 Technical Point of Contract: Judi Stasko 515-337-7445 judith.stasko@usda.gov.
| CONTRACTOR REQUIREMENTS |
| 3.0 Technical Requirements/Tasks: |
| 6.0 Travel: Two (2) preventive maintenance service visit will be provided by the contractor. All service visits will be coordinated with the BD FACSymphony A5 Operator at NADC. |