Statement Of Work_ADCP System Repair Refurb CD0001273110_June2023.docx

DOCX document 34 KB Posted

Attached to
Teledyne ACDP Repair - EEB Federal contract opportunity
Solicitation number
NM-AN2120-23-01367
Issued by
Department of Commerce National Oceanic and Atmospheric Administration

About this file

This statement of work outlines repair and refurbishment services required for Teledyne-RDI acoustic Doppler current profilers. The National Oceanic and Atmospheric Administration requires repair of a 75kHz transducer and electronics chassis, replacement of defective components as identified, cleaning, refurbishment and performance of acceptance testing. Deliverables include a level 1 service report identifying repair needs, replacement of parts on the transducer and chassis, and providing test results to NOAA demonstrating full performance specification compliance. Work must commence within 30 days of award and be completed 60 days post-award at Teledyne-RDI's facility in Poway, California, with repaired systems delivered to NOAA in Newport, Oregon.

View the file

Other files for this federal contract opportunity

Other files attached to Teledyne ACDP Repair - EEB, newest first.
File Type Posted
FAC 23-04 Provisions and Clauses.docx DOCX document
Wage Determination 2015-5575 rev 21.pdf PDF
Equivalent Rate 113R-23Aug2022.pdf PDF
SAP JOFOC for Solicitation.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK (SOW)

FOR

TELEDYNE-RDI ADCP REPAIR/REFURBISHMENT

US Department of Commerce National Oceanic and Atmospheric Administration Office of Marine and Aviation Operations

1. Introduction and Overview (Describe the Departments purpose/mission and the current environment)

NOAA’s Office of Marine and Aviation Operations manages and operates a fleet of 16 Oceanographic Survey ships and 4 regional port offices to support ship operations and maintenance. Of these 16 ships, the deep-water vessels have hull-mounted Teledyne-RDI Acoustic Doppler Current Profilers (ADCP) as part of their oceanographic instrumentation; they are either model Ocean Surveyor 75 kHz (OS75) or Ocean Surveyor 150 kHz (OS150) systems. These systems acoustically measure the ocean currents beneath the ship to a depth of up to 1,000m.

NOAA has had complete systems fail during 2019 through early 2023 and has a need for them to be repaired/refurbished/upgraded and put back into the Parts Depot to ready to replace installed systems the next time they fail. There is one transducer identified repair components that must be replaced and other refurbishment tasks that must be completed to restore full functionality to that transducer. These are very sophisticated systems that have components, specifically the transducer that must be replaced while the ship is in dry-dock. A typical repair of the system cannot be done within the time frame of our usually 4-to-8-week dry-docking periods. Having a repaired, tested and installation-ready system on hand allows NOAA to minimize the time in dry-dock and thus the cost of the dry-docking.

Information on what the Items are to be performed / provided:

The following summarizes the required items necessary to fulfill this procurement request:

1) Per the OO/OS Level 1 Service Report of conditions found and recommended repair actions against RMA 098973; clean, replace defective parts and complete refurbishment of P/N 71A-1029-00 OS75 Transducer S/N 2167. (The serial number of the sonar head was deteriorated to the point of non-recognition. Teledyne has provided the software recovery of the serial number.) Perform an in-water final acceptance test to verify full performance per manufacturer specification; provide copy of completed test results to NOAA with the transducer.

2) Perform OS/OO Chassis OR DRUC Level 1 Service on P/N 71A-2010-02 Electronics Chassis S/N 1047 (CD0001273110) and interconnecting cable. Provide a Level 1 Service Report of conditions found, recommended replacement parts/materials, and a quote to complete repairs. Replace defective parts and complete refurbishment P/N 71A-2010-02 Electronics Chassis S/N 1047. Perform the in-water final acceptance test using chassis s/n 1047 in conjunction with transducer s/n 2167 to verify full performance per manufacturer specification; provide copy of completed test results to NOAA with the transducer and electronics chassis.

2. Scope of Work

The scope of this work is for the repair/refurbishment and final acceptance testing of the transducer listed in 1.1) above and for complete OO/OS Level 1 Service assessment of one complete system listed in 1.2) above with OO/OS Level 1 Service Report of conditions found, recommended replacement/repair parts identified, and a quote to complete those repairs delivered to NOAA. Replace defective parts and complete refurbishment of P/N 71A-1029-00 OS75 Transducer and P/N 71A-2010-02 Electronics Chassis identified above.

Final acceptance shall include thorough functional testing of the Electronics Chassis in all operational modes with associated transducer testing to include a tank/lake test to ensure the transducer works in water and water does not intrude into the transducer. Copy of final acceptance test results shall be delivered to NOAA with the returned systems and components.

The shipping packages used to send the equipment to Teledyne-RDI shall be used for return of the repaired/refurbished equipment.

3. Objectives

The overall objective of this Statement of Work is to procure the services to repair and make ready for deployment the Teledyne-RDI ADCP systems and components listed in Paragraph 1 above.

4. Tasks

Perform the following:

1) Per the Level 1 Service Report of conditions found and recommended repair actions against RMA 098973; replace defective parts and complete refurbishment of P/N 71A-1029-00 OS75 Transducer S/N 2167. Complete a DRUC Level 1 service on P/N 71A-2010-02 Electronics Chassis S/N 1047 and interconnecting cable. Then perform an in-water final acceptance of transducer s/n 2167 together with chassis s/n 1047 test to verify full performance per manufacturer specification and provide copy of completed test results to NOAA with the transducer.

The Services to accomplish the work are:

Currently these are the identified repair work items required (identified) and other items may be possible after the full testing is completed:

· OSAF-FIX-LV01OS/OO/DRUC SYSTEM LEVEL 1 SERVICE
· AF-BRNCL-RMVLREMOVAL OF ANTI-FOULING PAINT OR BARNACLES from 75Khz transducer
· 713-1010-0075KHZ CERAMIC ARRAY replacement
· OSAF-FIX-ECCIOS END CAP CONNECTOR ASSEMBLY installation
· 73A-1010-00ASSY, CONNECTOR & WIRES, OS PHASE II END CAP (CONNECTOR ASSEMBLY) (STANDARD)
· 20828-128HANDLE, CHASSIS, OS
· 1165JAM NUT, 1IN-20, HEX SST
· 2-022O-RING, BORE SEAL/MINPCCP , DURO 70, E-0603 or E-0803
· 22-01-3107CONN, HOUSING, 10P
· 39-01-2245CONNECTOR, INTERNAL, OS TOP HAT CABLE ASSEMBLY
· 71AK6002-00KIT, CERAMIC ASSEMBLY, OS75, REPAIR
· 71AK6003-00KIT, CLOSE-UP, OS75, REPAIR
· 97Z-6017-00O-RING, 2-020 ,70 DURO, EDPM
· 97Z-6018-00O-RING, 2-021 ,70 DURO, EDPM
· 97Z-6020-00O-RING, 2-024 ,70 DURO, EDPM
· 97Z-6084-00O-RING, 2-015 ..070DIAX.551 ID, EPDM, DURO 90A,VENT PLUG
· 97Z-6084-01O-RING, 3-094, .072DIAX.351 ID, EPDM, DURO90A, VENT PLUG
· B65702-B0000-T001COIL FORMER, 1 SEC, P3019
· B65705-B0003-X000CARRIER, P3019 8-TERM, & YOKE
· OSAF-DIA-LAKEOS/OO LAKE TEST (Does not include repair)
· 43666Perform ECO POWER SUPPLY, OS CHASSIS, OS II, Lock-Tite; with Upgrade A2 to A6
·Return freight shipping

5. Government-Furnished Material / Government-Furnished Equipment

1) OS75 Transducer, P/N: 71A-1029-00, SN: 2167 & Chassis, P/N 71A-2010-02 S/N: 1047 (CD0001273110)

6. Security

There are no special security considerations associated with this Statement of Work.

7. Period of Performance Work shall begin within thirty (30) days of contract award and shall be completed no later than 60 days following date of contract award.

8. Place of Performance Teledyne-RDI 14020 Stowe Drive Poway, CA 92064.

9. End Results/Deliverables Upon completion of repairs/refurbishment and final acceptance testing of each system/component, they shall be delivered to:

NOAA MOC Pacific / MOE electronics ATTN: Gordon Manning (W122)

2002 SE Marine Science Drive
Newport, OR 97365-5300
Phone: 541-876-8717 or Cell: 541-270-2497

10. References RMA 09873 with work inspection report completed by Teledyne RDi Technician: Lue Vue