Specifications.pdf
PDF 137 KB Posted
- Attached to
- Purchase Time-Lapse Benchtop Incubator Federal contract opportunity
- Solicitation number
- 12805B21Q0318
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SF-18 with Clauses and Provisions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications Time-Lapse Tri-Gas Incubator
GENERAL INFORMATION
1.0 Background and Purpose:
Scientists at the Roman L. Hruska U.S. Meat Animal Research Center (USMARC) are developing scientific information and new technology to solve high priority problems for the U.S. beef, sheep, and swine industries. As part of the effort, they require a time-lapse tri-gas incubator with a pH monitor system to monitor oocyte competency and early embryonic development.
CONTRACTOR REQUIREMENTS
2.0 Technical Requirements/Tasks:
The requirement is for one time lapse tri-gas incubator with pH monitoring. The time-lapse tri-gas incubator with a pH monitor shall have the following capability:
Six (6) individual chambers for controlled culturing environment.
Continuous pH measurement to monitor culture media.
Overall Dimension: 786 x 596 x 380 mm (30.9 x 23.5 x 15.0 Temperature Control Range: 25 – 40 °C Gas Consumption (CO2): < 2 L/h Gas Consumption (N2): < 10 L/h CO2 Control Range: 1.9 – 10% O2 Control Range: 5 – 20% Input Gas Range: 0.6 bar (8.7 psi) Built-in Microscope: Zeiss 20x objective has number aperture of 0.35, specialized for
635 nm illumination.
Embryo Illumination: 0.064s per image, using IW single red LED *(635nm) Camera Resolution: 1280 x 1024. Monochrome, 8-bit, IDS system.
Optics Tube Ratio: 2.22 px/um Imaging Focal Planes: 5 min. image interval in 3 to 7 focal planes.
3.0 Deliverable/Schedule/Address:
Delivered 90 days after award. Facility address is listed below:
USMARC
844 Rd 313 Clay Center, NE 68933
Key Deliverables
Item No. Deliverable Objective 1 One (1) time-lapse tri-gas incubator with a pH monitor system the technical requirements.
One (1) time-lapse tri-gas incubator with a pH monitor system per the technical requirements and within the 90 days award
4.0 Performance Requirement Measures:
A one-time delivery One (1) time-lapse tri-gas incubator with a pH monitor system per the technical requirements within the expected delivery time.
5.0 Government Point of Contact:
Alexandria Snider, Phone: (402)762-4178, alex.snider@usda.gov