This document is the Specifications for a Gas Chromatograph (GC) system that the U.S. Department of Agriculture (USDA) desires to procure. The GC system must be capable of simultaneously measuring methane, nitrous oxide, carbon dioxide, oxygen, and nitrogen at various concentration levels. It must be a commercial off-the-shelf product that can accommodate at least three detectors (FID, TCD, and ECD) and have automated headspace sampling capabilities. The system must meet specific performance requirements, including analytical ranges and precisions, and have on-site installation, training, and calibration.

This is a sources sought notice only, not a solicitation. The USDA plans to award a single firm-fixed-price contract on a small business set-aside basis, provided at least two qualified small businesses respond with information to support the set-aside. The anticipated NAICS code is 334516 with a size standard of 1,000 employees. Interested small businesses should provide their company information, small business status, and confirmation that they can meet the stated specifications with a Made in America GC system.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications:

Gas chromatograph (GC) system capable of simultaneous measurement of gases: methane (CH4 – ppm levels), nitrous oxide (N2O; ppb to ppm levels), carbon dioxide (CO2: ppm to % levels), oxygen (O2) and nitrogen (N2) at percent levels (1-100% by volume). The requirement is an instrument that can analyze gas mixtures collected in 12-mL headspace vials from both laboratory and field sites. Thus, the gases will have a wide range of concentrations and the instrument must be calibrated across a range of analyte concentrations. This type of analysis is conducted by using 3 detectors (TCD, ECD and FID) with appropriate columns allowing for the detection of the suite of gases listed above.

The GC must have the ability to automatically analyze the headspace vials, typically achieved using an automated headspace sampler. The sampler must have at least the capability to hold 250 vials per loading to ensure overnight and weekend analytical operation. The GC will need to be interfaced through a 14-port (or two 10 port sample valves) to allow the simultaneous injection of the same sample into three separate analytical columns.

Gas chromatograph (GC) system (i.e., analytical laboratory instrument, computer, required software and license, and other peripheral items such as gas regulators, column, etc.) to analyze the gases in the table below:

Table 1. Gas analysis requirements

Chemical Formula Analytical range required Desired analytical precision

Methane CH4 0.3 to 100 ppm +/- 2% or better Carbon dioxide CO2 200 ppm to 55% +/- 2% or better Nitrogen N2 1-100% +/- 2% or better Oxygen O2 1-21% +/- 2% or better Nitrous oxide N2O 100 ppb to 100 ppm +/- 2% or better

• The GC system shall be a commercial off-the-shelf product with a model or product number and arrive pre-assembled or with only minor assembly needed to fit in the desired lab space.

• POWER SOURCE must be 208 to 240 VAC, 50/60 HZ.

• Footprint of the system shall fit on a typical laboratory benchtop (approx. 24” x 36”)

• The GC must be able to accommodate at least 3 GC detectors (FID, TCD, and

ECD) simultaneously and they must be individually temperature controlled with electronic pressure control (EPC) modules for support gases and housed in the same GC unit (not an external box or module)

• The GC shall have one thermal conductivity detector, one flame ionization detector, and one electron capture detector. All three detectors must be capable of analyzing column flow from packed (1/8”) columns and not capillary columns.

• On-site installation by factory-trained technician.

• In-lab training and familiarity with the system provided and included shall be required.

• Technician must calibrate instrument as part of on-site installation process, and ensure instrument is performing to specifications listed in Table 1 in order for installation to be complete.

• The GC shall have an interface with a keypad for manual override operation.

• The GC shall have programming capabilities (temperature, time, valve switching, etc.)

• The GC shall be able to perform automatic leak checks.

• The GC software shall have all instrument controls from one monitor/computer screen to include all configuration, GC analytical conditions, batch automated sample handling software to minimize initial startup training time and future user training.

• The GC software shall have all data processing controls accessible from one window.

• QA/QC software for the GC shall be included that can monitor all operational parameters such as resolution and baseline values.

• GC software shall include monitoring of routine maintenance parts such as columns, injection liners and septa to insure constant proper performance.

• For data processing and analysis, the GC software shall allow for peak integration manipulation, identification, quantitation, and manual calibration curve creation for all 3 detectors simultaneously.

• The software shall be able to export sample data in digital formats compatible with Microsoft Excel such as .xlsx, .csv, or .txt files.

Number of units: 1