Text version

Specifications

V127A152/5M-HV-Q-X

Description: Large volume Nal(TI) detector with built-in high voltage power supply and special connector

Scintillation crystal:Nal*TI), 127 x 127 x 152 mm
Design drawing:VS-548-10
Housing:1.0 mm thick stainless steel
Optical window:Quartz, 12.7 mm thick
Photomultiplier tube:127 mm diameter, ET 9390
Internal high voltage generator:Corkcroft Walton type PS1822/5
High voltage regulation:By means of external control voltage via white Wire (pin 4 of MOLEX connector)
Internal HV supply HV control:accessible via hole in backcap
HV test point voltage:1V = 1kV
Temperature sensitivity:Cathode-to-anode voltage changes: < 0.03, 0.01 typical (+20 degrees Celsius to 50 degrees Celsius)
Warm-up time:0.3% of final output after 10 s. 0.02% of final output after 10 min.
High voltage discharge time:< 60 s to < 30 V
Line regulation:0.005% per volt power supply voltage
Gain stability:< 1% up to 20 uA anode current
Power requirements:+5 V, power requirement 100 mW
Electrical Connections:Negative high voltage circuit (anode at ground potential). Readout at anode of PMT

45 cm long flying leads cables through two openings in backcap of detector to MOLEX connector type 90156-0146 WM 8021-ND (Farnell)

FunctionCable colourPin number
+5V input VoltageRed1
Common groundBlack2
High Voltage control (INT)Yellow3
Control VoltageWhite4
Anode outputCoax RG 174 core5
Signal groundCoax RG 174 braid6

Performance criteria Energy resolution (662 keV) < 8.52 % FWHM

Note: The High voltage should be chosen as large as possible given the required dynamic range. This ensures the optimum signal-to-noise ratio.

File details come from the government source that posted it. Updated .