Specification Document.pdf

PDF 271 KB Posted

Attached to
Oxidation Stability Analysis Machine Federal contract opportunity
Solicitation number
12505B24Q0008
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Midwest Area

About this file

This specification document outlines requirements for an oxidation stability analysis machine. The Department of Agriculture Agricultural Research Service Midwest Area seeks to purchase an instrument that can measure the oxidative stability index of fats, oils, and biodiesel according to AOCS official method Cd 12b-92 and ISO 6886. The machine must have a minimum of eight sample measuring locations, temperature range of 50-220 degrees Celsius adjustable in 1 degree increments, built-in air pump with 1-25 liter per hour flow range, and conductivity measuring vessels capable of holding 50ml of water and covering an aeration tube by at least 5cm. The instrument is to interface with a computer via USB to monitor conductivity of each probe, automatically or manually calculate endpoints, and save data. Delivery is required no later than 120 days after receipt of a purchase order.

View the file

Other files for this federal contract opportunity

Other files attached to Oxidation Stability Analysis Machine, newest first.
File Type Posted
Sources Sought Response Form.pdf PDF
NDAA 889 Certification.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPCECIFICATION DOCUMENT

Project Oxidation Stability Analysis Machine

General Information

1.0 Scope of Work

Purchase an oxidation stability analysis machine for the purpose of determining the “Oil Stability Index” of food fats and oils for ARS research according to AOCS official method Cd 12b-92. Since the instrument may occasionally also be used for biodiesel analysis by the BOR research unit, the biodiesel method sample programming was also requested as an add-on.

2.0 Background

An instrument is required for the purpose of determining the Oil Stability Index of fats and oils according to the AOCS official method Cd 12b-92. The requirements and specifications of the instrument are laid out by the method and are copied below.

3.0 Technical Requirements / Tasks

• instrument that can measure the oxidative stability index of fats and oils and biodiesel according to AOCS official method Cd 1b-92, ISO 6886.

• A minimum of eight sample measuring locations

• Temperature range of 50°C to 220°C that can be set in steps of 1°C increments.

• Temperature calibration/correction capability, and maximum deviation of < 0.3°C from the set temperature.

• Built-in air pump with a flow range of 1-25 L/h

• Conductivity measuring vessels with a conductivity measuring capability of 0-400 µs/cm.

• Conductivity measurement vessels capable of holding 50 ml of water, and capable of covering the tip of an aeration tube by 5 cm or more.

• Must interface with a computer through a USB or similar port interface.

• Software that when operated through a computer, should monitor the conductivity of each probe and be able to automatically calculate the end point, or be configured to manually calculate the endpoint, and save data.

5.0 Deliverables / Schedule

Key Deliverables

Item No. Deliverable Objective Due 1 Delivery of instrument N/A No later than 120 days after purchase order is provided to the vendor

SPCECIFICATION DOCUMENT
Project
General Information
1.0 Scope of Work
2.0 Background
3.0 Technical Requirements / Tasks
5.0 Deliverables / Schedule