SOW.pdf

PDF 297 KB Posted

Attached to
Environmental Test Chambers (2) for research/testing Federal contract opportunity
Solicitation number
80NSSC24881418Q
Issued by
National Aeronautics and Space Administration Shared Services Center

About this file

This document is a Statement of Work (SOW) for two environmental test chambers required for electrochemical testing. The chambers must have a stainless steel internal construction, 20 ft3 internal volume, temperature range of -50C to 150C with fluctuation of +/- 1C, built-in nitrogen purging, viewing window, and other specified features. Vendors must provide details on the temperature range, control tolerance, electrical requirements, dimensions, controller model, and weight limits.

This SOW is related to a federal contract opportunity from the National Aeronautics and Space Administration Shared Services Center, with Solicitation Number 80NSSC24881418Q for the procurement of the two environmental test chambers described in the SOW.

View the file

Other files for this federal contract opportunity

Other files attached to Environmental Test Chambers (2) for research/testing, newest first.
File Type Posted
Questions and Answers.pdf PDF
Questions and Answers.pdf PDF
RFQ.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Requirements for Environmental Test Chambers

Two (2) environmental test chambers are sought for electrochemical testing. The desired systems shall provide the following features and capabilities.

1. Stainless steel internal construction

2. 20 ft3 internal chamber volume

3. Temperature control from at least -50°C to 150°C

4. Temperature fluctuation at steady state (after stabilization) shall be no greater than plus or minus 1°C

5. Built-in gaseous nitrogen purging capability

6. Window in door

7. Internal chamber light

8. Exterior height less than or equal to 80 inches (including casters).

9. Self-contained refrigeration system

10. Minimum of two openings of at least 3 inches in diameter for instrumentation feed-through in side wall.

11. One adjustable stainless steel shelf with a weight limit of at least 25 pounds

12. Failsafe backup of main controller to provide a redundant temperature limiting system (limit controller). Shall be adjustable over temperature range of chamber. Shall have an alarm feature (can be visual) to indicate to the operator when an out-of-tolerance temperature has been reached.

13. 110/120 volt or 208 volt single-phase AC operation

Vendors shall specify the following:

o Temperature range o Temperature control tolerance o Electrical requirements of chamber (current, voltage, number of phases) o Interior and exterior dimensions o Manufacturer and model number of main temperature controller o Weight limit of shelf and total weight limit of chamber

File details come from the government source that posted it. Updated .