This is a detailed specification document listing the salient characteristics required for two particle counter devices being procured by the Defense Logistics Agency Land and Maritime. The particle counters must be capable of measuring liquids with viscosity up to 1600 cSt, including hydraulic oils, gear oils, engine oils, and water-based hydraulic fluids, with external pressure supply and precise flow rate control.
The devices must meet specific calibration requirements according to ISO 11171 and ISO 4402 standards, with calibration ranges of 4-70 μm(c) and 2-100 μm respectively. Additional requirements include reporting capabilities for ISO 4406, SAE AS 4059, and NAS 1638 standards, ability to measure volumes from 1-1,000 mL, compatibility with sample bottles up to 7 cm diameter and 18 cm height, and maximum dimensions of 400mm x 280mm x 600mm. The units must have SEA 04Q (LQAO) approval, include a printer, operate with or without a computer, have unique serial numbers, and come with suitable transport cases for shipping and calibration.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SOLE SOURCE 42938701_r.pdf | ||
| 25Q0471 ANALYZER SYSTEM.pdf | ||
| CDRL WARRANTY.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
JML#s 42938701 THRU 8707 SUPPLEMENTAL INFORMATION
DETAILED DESCRIPTION OF SERVICE: PURCHASE OF TWO (2) PARTICLE COUNTERS
THE EQUIPMENT AND ACCESSORIES MUST HAVE THE FOLLOWING SALIENT
CHARACTERISTICS:
1. Must be able to measure liquids with a viscosity of up to 1600 cSt with an external pressure supply and precise flow rate control for numerous fluids that include hydraulic oils, gear and engine oils, and water based hydraulic fluids.
2. Must be calibrated according to ISO 11171 with test dust ISO MTD with a calibration range of 4-70 µm(c).
3. Must be calibrated according to ISO 4402 with test dust ACFTD with a calibration range of 2 to 100 µm.
4. Must be able to report results according to the following standards: ISO 4406, SAE AS 4059, and NAS 1638 as well as raw data.
5. Must be able to use any sample bottle of a maximum diameter of 7 cm and a height of 18 cm.
6. Must include automatic sample flow and volume control.
7. Must be capable of measuring volumes of 1 to 1,000 mL.
8. Must be usable with or without a computer.
9. Must include printer.
10. Must be a maximum of 400 mm by 280 mm by 600 mm (W x D x H).
11. Must be model with SEA 04Q (LQAO) approval.
12. Must have an unique identifier such as a serial number.
13. Must include transport case that is suitable for shipping particle counter for service and calibration.