Salient Characteristics.pdf

PDF 448 KB Posted

Attached to
ICS6000 SYSTEM Federal contract opportunity
Solicitation number
SPMYM223Q2001
Issued by
Defense Logistics Agency Land and Maritime

About this file

This document outlines the salient characteristics required of an ion chromatograph system to be procured through solicitation SPMYM223Q2001 issued by the Defense Logistics Agency Land and Maritime. The system must include modules for pumping, detection, autosampling, eluent generation, mass spectrometry, nitrogen generation, and auxiliary pumping. Key requirements are that the system be new, modular, NRTL listed, compatible with the Navy's DADMS software, able to interface with a computer and support a mass spectrometer detector, and include specific pump, detector, autosampler, eluent generator, mass spectrometer, nitrogen generator, and auxiliary pump models meeting size, capability, and technology specifications laid out in the document.

View the file

Other files for this federal contract opportunity

Other files attached to ICS6000 SYSTEM, newest first.
File Type Posted
31418700 THRU 8746 - CDRL MANUALS.pdf PDF
23Q2001 ICS6000 SYSTEM.pdf PDF
31418700 THRU 8746 - CDRL WARRANTY.pdf PDF
SOW 31418700 THRU 8746.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

JMLS 31418700 THRU 8747 – SALIENT FEATURES

Salient Characteristics

Must be an ion chromatograph system Must be new Must be a modular system Must be NRTL listed Must be able to support a mass spectrometer detector Must have a software that is pre-approved by the Navy (DADMS) Must be able to operate on a wired connection Must be compatible with a computer system System must include a pumping system that… o Has a dual channel configuration option o Has isocratic elution (not gradient elution) for both channels o Has microbore format on the cation channel o Has microbore format on the anion channel o Can continuously operate up to 5000 psi

System must include a DC detector that… o Is capable of having a microbore compartment o Has an injection value o Has dual temperature zones at low temperatures o Have a temperature stability of no more than 0.05°C

System must include an autosampler that… o Can be automated using a software that is pre-approved by the Navy (DADMS) o Can hold up to 81 x 10mL vials and up to 120 x 1.5 mL vials o Can perform simultaneous and sequential injections

System must include an eluent generator that… o Has the ability to support analytical and capillary eluent generation o Has Reagent-Free Ion Chromatography technology o Does not include an EG Degasser o Has dimensions of up to 16” x 8.75” x 21.5” (HxWxD)

System must include a mass spectrometer that… o Is a single quadrupole mass spectrometer o Has a mass range (m/z) of up to 1250 o Has a scan rate of up to 20,000 Da/s o Has a heated electrospray ionization (HESI) system o Has dimensions of up to 20.5” x 16.5” x 36” (HxWxD)

System must include a nitrogen generator that… o Has a flow rate up to 35 L/min o Has a max pressure of up to 116 psi o Has dimensions of up to 25.6” x 22.4” x 28” (HxWxD) o Has electronic compressor optimization (ECO) technology o Has up to 2 internal compressors

System must include an auxiliary pump that… o Has dimensions of up to 16.5 cm X 25.4 cm x 16.5 cm (HxLxW) o Weighs up to 15 pounds o Is compatible with a single quadrupole mass spectrometer

File details come from the government source that posted it. Updated .