Questions and Answers.pdf
PDF 2 MB Posted
- Attached to
- Standing Prevacuum Steam Sterilizer (Autoclave), Riverside, CA Federal contract opportunity
- Solicitation number
- 12905B23Q0147
About this file
This document contains questions and answers regarding solicitation number 12905B23Q0147 from the Department of Agriculture Agricultural Research Service for a standing prevacuum steam sterilizer (autoclave) in Riverside, California. Key details include that 208V 3-phase electrical power is available on site, compressed air can be supplied to the autoclave's location, appropriate feedwater quality will be provided, and photos as well as a dimensioned drawing of the autoclave room are included to ensure proper fit. The document provides logistical information to support bidding on the standing prevacuum steam sterilizer opportunity.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SF30_12905B23Q0147_0001.pdf | ||
| Attachment 1 Specifications.pdf | ||
| RFQ_12905B23Q0147 Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
RFQ #12905B23Q0147 Amendment 1 Standing Prevacuum Steam Sterilizer (Autoclave), Riverside, CA
Ques ons and Answers
1. Is a loading dock available for delivery? yes
2. Is there a forkli on site? yes
3. Will the autoclave be installed on the ground floor? If not, is an elevator available for transport to its final loca on? Elevator not required.
4. If an elevator is required for transport, can you confirm its door opening size, internal cab dimensions and weight limit? n/a
5. Are there any obstacles in the pathway to the autoclave’s final loca on? Such as stairs, ledges, ght turns, etc.? doorways; see drawing.
6. What is the hallway size in the pathway to the autoclave’s final loca on? 5’6” See drawing.
7. What is the smallest doorway size in the pathway to the autoclave’s final loca on? 36” wide; 7’ high
8. Can “house compressed air” (90 PSI minimum) be made available at the autoclave’s final loca on? yes
9. Will the “USDA-ARS” supply the appropriate feedwater quality to the autoclave’s electric steam generator? This may require a reverse osmosis system, to reduce “hardness” and “total dissolved solids.”
Please confirm the appropriate feedwater quality will be provided, or if the autoclave vendor will need to provide a pre-treatment system, as part of their scope of work? yes
10. 3-Phase electrical power (208V, 240V or 480V) is required for opera ng the electric steam generator.
Which 3-Phase voltage is available on site? 208V
11. The bid states a “manual ver cal sliding door.” Will the “USDA-ARS” accept an automa c ver cal sliding door, with built-in safety devices? yes
12. The bid states a “steam-ac vated door seal.” Will the “USDA-ARS” accept compressed air ac vated door seal, which extends the life of the door gasket? yes
13. The bid states a “One-Year Warranty.” We assume you are referring to a One-Year Warranty on both parts and labor. Please confirm. yes
14. Please provide photos of the exis ng autoclave, since it is the vendor’s responsibility for de-installa on, removal and disposal of the exis ng unit. If possible, please also provide photos of the exis ng u li es. Photos and room measurements are especially important, since the bid states “Autoclave must fit exis ng lab space, plumbing, electrical and ven la on.” Photos provided.
15. Can a dimensioned drawing of the autoclave room be made available? This informa on would be helpful to have, prior to bidding, to ensure a proper fit, and verify there is adequate space for service access. Drawing provided.
HEADHOUSE
jerald.bentley Highlight jerald.bentley Highlight
| Questions and Answers |
| Autoclave Measurment BP |
| 0002.pdf |