Attachment 1 Specifications.pdf
PDF 71 KB Posted
- Attached to
- Standing Prevacuum Steam Sterilizer (Autoclave), Riverside, CA Federal contract opportunity
- Solicitation number
- 12905B23Q0147
About this file
This document contains specifications for a prevacuum standing steam sterilizer and accessories. The specifications call for a small prevacuum steam sterilizer no larger than 20" W x 20" H x 38" D to fit existing lab space, plumbing, electrical, and ventilation. Key required features include a steam-jacketed chamber, manual vertical sliding door, steam-activated door seal, electric steam generator, and minimum pre-programmed cycles for gravity, prevacuum, liquid, and waste bag. The related federal contract opportunity is for a standing prevacuum steam sterilizer to be located in Riverside, California, issued by the Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area. Installation, training, warranty, and disposal services are also required.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Questions and Answers.pdf | ||
| SF30_12905B23Q0147_0001.pdf | ||
| RFQ_12905B23Q0147 Terms and Conditions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Atachment 1 - Specifica�ons
Prevacuum Standing Steam Sterilizer and Accessories
Prevacuum Steam Sterilizer (autoclave) for steriliza�on including, but not limited to, biowaste for disposal. Autoclave must fit exis�ng lab space, plumbing, electrical and ven�la�on. The prevacuum cycle is required to decontaminate wastes.
Brand Name or Equal. Must be equal to model Steris AMSCO 250LS Small Steam Sterilizer, loading rack and two shelves (for 20x20x38 sterilizer; single door), seismic kit htps://www.sterislifesciences.com/products/steam-sterilizers/steamsterilizers/ small-steam-sterilizers
PERFORMANCE REQUIREMENTS AND SALIENT CHARACTERISITCS
Small Sterilizer
Dimensions (sliding door) Interior chamber: 20” W x 20” H x 38” D Exterior: 30” W x 74.5”H x 45.375” D
Accessories for small sterilizer Loading rack and two shelves (for 20x20x38 sterilizer; single door) Seismic Kit
Required Features
Steam-jacketed Manual ver�cal sliding door Steam-ac�vated door seal Standard piping: brass and copper threaded Electric steam generator Minimum pre-programmed cycles: gravity, prevacuum, liquid, waste bag Program shutdown
Installa�on
Uncrate and inspect for damage
Assemble and re-erect equipment Assemble equipment piping package and accessories Install standard final panels Move equipment to facility loca�on and level Make all air, steam, water, and drain connec�ons Make electrical connec�on Install cabinet package on the Unit (perform opera�onal check-out and adjustments; demonstrate equipment for facili�es personnel) Install seismic anchoring mee�ng local building codes
Deinstalla�on
Deinstall exis�ng small sterilizer Dispose of exis�ng small sterilizer off site.
One-year Warranty