Specifications

Product

CLIN
Manufacturer
Description
Unit
Qty
1
Ted Twietmeyer
Driver box with 5 zero-crossing BNC outputs+ Digital size control front panel, OLED display,

Scan size: rear panel BNC input, 1U rack mount Power: 110/220VAC 50/60Hz

· Seamless integration with custom EOPC resonance scanner required with 10 to 16 kHz scanning range present by EOPC

· Storage of 1 to 10 present scan sizes

· Automatic dimming with room lights

· Digital display brightness control

· Real time resonant scanner frequency display

· 100 second resonant frequency stability self test to +/- 1 Hz

· Zero crossing outputs: 5 BNC rear panel, 1 BNC front panel

· Zero crossing AC characteristics: rise/fall time 3 ns

· Enclosure: standard 19” rack mount, 1U rack height

· Fuses: glass 3AG and .4A

· Power: 115/120 V AC with AC line filtering

Each
2
2
Ted Twietmeyer
4 channel inverting amplifier 1U rack mount

Gain: All channels are 1x (1:1)

· Channel input and output connections: 50 ohm BNC female x4

· Bandwidth: DC to 130 MHz

· Input voltage range: +/- 2 Vpp

· Input impedance: 50 ohms

· Output impedance: 50 or 75 ohms

· Noise: < 10 mv

· Enclosure: standard 19” rack mount, 1U rack height

· Fuses: .4A x2

· Power: 115/120 V AC with AC line filtering

Each
3

Delivery Location National Institutes of Health National Eye Institute ATTN: Johnny Tam 9000 Rockville Pike Building 10 Room 10N226 Bethesda, MD 20892

Delivery Date Within 6 weeks after receipt of order.

Quote Evaluation Process Sole Source

Acquisition Contact

TBD