Q and A - Sol N00173-20-Q-CM02.pdf
PDF 276 KB Posted
- Attached to
- Vector Network Analyzer Federal contract opportunity
- Solicitation number
- N00173-20-Q-CM02
About this file
This document contains a request for quotations for a vector network analyzer system. The Naval Research Laboratory seeks to procure a Keysight Technologies N5290A 900 Hz to 110 GHz PNA MM-Wave system including an N5222B-205 26.5 GHz 2 port vector network analyzer with a 3 year calibration plan and extended warranty. Additional items include an 85059B 1mm precision calibration kit and its 3 year calibration plan. Quotations are requested by qualified vendors and must be received within 26 weeks of receipt of order, though earlier delivery is preferred. Upon inspection and acceptance at the specified Naval Research Laboratory location, a firm fixed price contract will be awarded.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Q and A - Sol N00173-20-Q-CM02 (Amended).pdf | ||
| RFQ - N00173-20-Q-CM02 (Amended).pdf | ||
| Exhibit A - Salient Characteristics (Amended).pdf | ||
| Exhibit A - Salient Characteristics.pdf | ||
| RFQ - N00173-20-Q-CM02.pdf | ||
| JA_13.501-20-CM01 (Redacted).pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Questions & Answers - Solicitation N00173-20-Q-CM02
Question 1: VNA: 26.5 GHz or 67 GHz?
Answer 1: Please provide a quote for the 26.5 GHz VNA.
Question 2: Time gating functions and rapid data collection feature software options: The software option S93010B and S93118B are recommended. Let us know if you want these added.
Answer 2: Please add these options.
Question 3: 1 mm cables and adapters: Please select one option from each group:
Test Port Cables
11500JK10 2 110 GHz, 1.0 mm (m-f) test port cable (10 cm) 3
11500JK13 2 110 GHz, 1.0 mm (m-f) test port cable (13 cm) 3
11500I 110 GHz, 1.0 mm (f-f) test port cable (8.8 cm) 11500J 110 GHz, 1.0 mm (m-f) test port cable (16 cm) 11500K 110 GHz, 1.0 mm (m-f) test port cable (20 cm) 11500L 110 GHz, 1.0 mm (m-f) test port cable (24 cm) 11500JK07 110 GHz, 1.0 mm (m-m) test port cable (30.0 cm)
Answer: 11500L is the preferred selection.
1.0 mm coaxial adapters
Y1900B 120 GHz, 1.0 mm (f) to 1.0 mm (f) ruggedized adapter Y1900C 120 GHz, 1.0 mm (m) to 1.0 mm (f) ruggedized adapter Y1910A 120 GHz, 1.0 mm (m) to 1.0 mm (m) standard adapter Y1910B 120 GHz, 1.0 mm (f) to 1.0 mm (f) standard adapter Y1910C 120 GHz , 1.0 mm (m) to 1.0 mm (f) standard adapter
Answer: Y1900B is the correct option.
Technical Review Comments:
Comment 1: The N5290A system will include a PNA or PNA-X series VNA, a test set controller, two frequency extenders, and interconnect cables required to connect the VNA to the test set controller. Some items in the spreadsheet are duplicated or already included in the N5290A-201.
Response 1: Correct, please disregard the duplication of information.
Comment 2: In the characteristics document, it mentioned the 67 GHz stand-alone VNA is preferred.
If so, you may choose N5290A-202, which includes N5227B with options 020, 205, test set N5292A with options 200, 224, and two frequency extenders N5293AX03. The N5227B is a PNA that works up to 67 GHz.
But the characteristics document also mentioned the RF ports are 3.5 mm. The RF ports connectors on N5227B are 1.85 mm, while the connectors on N5222B is 3.5 mm since the N5222B only works up to 26.5 GHz. Both N5222B and N5227B are compatible with 1mm mmWave extenders because the extenders are connected to the test set controller. The test controller and PNA are connected via interconnected cables that are included in the N5290A system. Therefore, please check to confirm which stand-alone PNA you will need.
Response 2: Please proceed based on the assumption of the N5222B with the 3.5mm ports.
Comment 3: The characteristics also require the time gating functions and rapid data collection feature. The software option S93010B and S93118B are recommended.
Response 3: You may add these software options.
Comment 4: Regarding the 1 mm cables and adapters, Keysight provide different models for different length and genders. You may choose the appropriate ones based on the connector types and genders of the device under test. The connectors on the mmWave frequency extenders are 1.0 mm male.
Response 4: Again, the 11500L with the Y1900B would be the correct option.