Grain NIR Specifications.docx

DOCX document 22 KB Posted

Attached to
NIR Analyzer for Pullman, WA Federal contract opportunity
Solicitation number
12905B23Q0192
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area

About this file

This document outlines specifications for a grain analyzer and references a related federal contract opportunity. The specifications require a small footprint, benchtop grain analyzer capable of obtaining whole wheat kernel percent protein and moisture content. It must come with protein and moisture calibrations for wheat kernels as well as maintenance and customer support services. The analyzer footprint must be 16.5 x 18.5 x 16.5 inches or less. It should provide filter-free, monochromator-based optics with a wavelength range of 400-1100nm, optimal for protein and moisture analysis. The optimal bandwidth is 7nm or less with a minimum of 1400 data points per scan and a scan speed of 1 second or less. The total analysis time must be 60 seconds or less. Easy data export via Ethernet or USB is preferred.

The related federal contract opportunity is solicitation number 12905B23Q0192 for an NIR analyzer in Pullman, Washington issued by the US Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area.

View the file

Other files for this federal contract opportunity

Other files attached to NIR Analyzer for Pullman, WA, newest first.
File Type Posted
Terms and Conditions.docx DOCX document
RFQ - Combined Syn-SOL - SAM Description Notice - UNDER SAT - Supply (8).docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATIONS

Grain Analyzer

Requirements:

A small footprint, benchtop grain analyzer for obtaining whole wheat kernel percent protein and moisture content.

· MUST come with protein and moisture calibrations for wheat kernels.

· MUST come with maintenance services and customer support services.

· MUST have a Machine footprint of 16.5 x 18.5 x 16.5 inches or less.

· Provide filter-free, monochromator-based optics.

· Wavelength range 400 - 1100 nm (optimal for protein and moisture) must have full range in one device.

· Optimal bandwidth of 7 nm or less.

· Minimum of 1400 data points per scan.

· Scan speed of 1 second or less per scan.

· Analysis time of 60 seconds or less.

· Easy data export/output for upload to local database (ethernet or USB ports preferred).

File details come from the government source that posted it. Updated .