Table3A. CR Systems Qualified forWelder Certification

CR System Component Component Information CR System

CR Scanner (Reader) manufacturer GE GE Carestream/Kodak Durr VMI model number CRx Tower

CR25NDTor

CR35NDT

CR Flex HPX-1 HD-CR35 5100MS equipment settings gain 150 PMT gain = 40 PMT gain = 600 PMT gain = 4.5 laser power laser power = HIGH laser power = 4 laser power =14 other

(specify) speed = 50 Beam Deflect Spd

= 2000 other

(specify) sampling resolution 50 micron 50 micron 25 micron 16 micron 30 micron maximum pixel value 65504 65504 4096 65535 65500

Imaging Plates (IP) manufacturer/type GE IPU or

Carestream XL

Blue

GEIPU Carestream XL Blue Carestream XL

Blue

Carestream XL Blue

Processing Cassette (if applicable) manufacturer GE n/a n/a n/a n/a type hard n/a n/a n/a n/a model very high resolution n/a n/a n/a n/a

Viewing Monitor display size (inches, diagonal)

21 21 21 15 21 resolution (megapixels, MP)

3 3 3 or 5 5 3

NOTE

CR systems shall pass USAF CR Process Controlsfor crack detection systems per TO 33B-1-2WP 106 01 with two exceptions:

1) Spatialresolution shall be 8 Ip/mm in bothorientations (TO 33B-1-2 WP 106 01 para 4.6) when viewed at 2Xabsolute magnification

2) Table 3B shall be used in lieuof TO 33B-1-2 WP 106 01 Table 4. "Technique and PV Range for EPS Test" o o

ISJ

OI

w n n

TO 00-25-252J

Table 3B. Technique and PV Range for TO 33B-I-2 EPS Test (welder certification systems)

CR scanner IP kV mA

Time (sec)

Allowable Pixel Value Range

Min Max

GE CR Tower GElPUor

Carestream XL Blue 70 5 180 12000 62000

GE CR Flex GEIPU 75 5 180 10000 62000

Carestream HPX-l Carestream XL Blue 70 5 180 2500 3900

Durr HD CR-35 Carestream XL Blue 75 5 180 10000 62000

VMI5100MS Carestream XL Blue 65 5 180 13000 62000

29. WP 005 04, Page 12,Table 4 ofthe basic manual has been amended to read as follows:

Table 4. Recommended Inspection Setting

MATERIAL MATERIAL

THICKNESS

KV MA TIME IQI FILM

Steel Tube 1.070.050" thick tube 120 4 4m, 14 sec

FE.25 5x7 M-8Pb

Stainless Steel

Tube

1.070.050" thick tube 120 4 4m, 14 sec

SS.25 5x7M-8Pb

Inconel Tube 1.070.050" thick tube 130 4 4m, 14 sec

IN .25 5x7 M-8Pb

Aluminum

Tube

1.070.050" thick tube 50 4 4m, 14 sec

AL.25 5 x 7 M-8

Magnesium Tube

1.070.050" thick tube 40 4 5m, 38 sec

MG.25 5 x 7 M-8

Titanium Tube 1.070.050" thick tube 90 4 4m, 14 sec

TI.25 5 x 7 M-8

Cobalt Tube 1.070.050" thick tube 140 4 8m, 27 sec

IN .25 5x7M-8Pb

Steel Sheet 0.250" thick sheet 150 4 12m, 39 sec

FE.43 5x7 M-8Pb

Steel Sheet 0.032" thick sheet 120 4 2m, 5 sec

FE.25 5 x 7 M Pb

Steel Sheet 0.125" thick sheet 120 4 5 min FE.25 5 x 7 M Pb

Stainless Steel

Sheet

0.020" thick sheet 120 4 lm,27 sec

SS.25 5 x 7 M Pb