Performance Work Statement Digital Radiographic Imaging System

This item is a digital radiography (x-ray) system suitable for Nondestructive Inspection (NDI) use in a military aircraft maintenance environment. It replaces traditional chemical-based x-ay film with a storage phosphor imaging plate. Technical requirements are contained in Product Description PD-AFLCMC-EZPT-16-01. It is used to detect fatigue cracks, water entrapment, foreign objects, and welding defects. This item uses commercial technology with accessories especially kitted to meet Air Force needs, however it is not a “Commercial-Off-The-Shelf” item. The ability to identify weld defects is the critical performance requirement. The item should meet and exceed process control standards as described in TO 33B-1-2, NONDESTRUCTIVE INSPECTION – General Procedures and Process Controls, WP 106 01, Computed Radiography Process Control.

The Digital Radiographic Imaging System system shall be capable to perform to the requirements of this PD throughout temperatures from 55 to 120 degrees Fahrenheit at 95% relative humidity, and from 0 to 10,000 feet above sea level. All proposals shall include test results conducted IAW MIL-PRF-28800F, paragraph 3.8.2, by an independent laboratory. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.

The Digital Radiographic Imaging System system shall be capable of being stored from 15 – 160 degrees Fahrenheit. All proposals shall include test results conducted IAW MIL-PRF-28800F, paragraph 3.8.2.1, by an independent laboratory. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.

The Digital Radiographic Imaging System shall be capable of operation when supplied with 100 – 250 Volt Alternating Current – 50 or 60 Hertz (100-250 VAC – 50/60 Hz) automatically selected. All proposals shall include test results conducted IAW MIL-PRF-28800F, paragraph 3.5.1, by an independent laboratory. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal.

Shipping Containers: The CR system shall include hard-shell carrying cases and be suitable for storing and shipping the unit with accessories. Carrying cases shall be appropriately padded to ensure potential shipping damage to the complete systems would be minimized. All proposals shall include test results from a 46 centimeter (18”) drop in the hard cases conducted IAW the procedure in MIL-PRF-28800F, paragraph 4.5.5.4.2, by an independent laboratory. Prior testing accomplished to a previous revision of MIL-PRF-28800 is acceptable, but shall be identified in the proposal. Cases shall have wheels and handles. (Note, it is desirable for cases be of similar size to facilitate stacking & palletizing for shipment).