Salient Characteristics
Benchtop SEM for Vandenberg AFB, CA
06 Apr 2018
CLIN 0001 – Scanning Electron Microscope (SEM)
- SEM suitable for a standard sized desktop and a fully-integrated system to include all necessary components for plug and play
- Magnification Range up to 100,000x
- Solid state crystal (CeB6) high brightness, long life electron source - the source must last 3 years before requiring replacement. If it does not last 3 years from date of delivery, the vendor must come on-site and replace the CeB6 at no additional cost to the Government.
- No effect on image quality in the presence of an 8 milligauss magnetic field
- Better than 14 nm resolution
- Capable of high resolution images at least 2048 x 2048
- Automatic image adjustment functions: Autofocus, Auto-brightness and Auto-contrast
- Backscatter detector, fully integrated with software
- x/y self-centering Motorized Stage
- 16x Color Zoom Color Microscope Camera with the same point of view as the detector
- Customizable accelerating voltage up to 18kV
- Interactive relationship between optical and electron image
- Accommodate samples up to 100 mm x 100 mm in size
- Backing Pump for Turbomolecular Pump
- Total pump down time from air to SEM image <45s
- 19"(min) monitor/mouse/keyboard
- Separate computer with necessary software for fully integrated energy-dispersive X-ray spectroscopy (EDS) for elemental analysis. Must include elemental identification software and be LN2 free.
CLIN 0002 – Training
- On-site SEM Training (1 day): Training instruction on instrument use and testing of SEM system(s). Training should include start up, sample preparation and loading, software use, sample analysis, shutdown and troubleshooting.
CLIN 0003 – Secondary Electron Detector
- Secondary Electron Detector (SED), fully integrated with software
CLIN 0004 – Elemental Mapping Software
- Elemental Mapping application software for Energy Dispersive Spectrometer (EDS): Provide elemental information over the sample