EndoDryRequirements.pdf

PDF 1 MB Posted

Attached to
Z1DZ--693-21-108 Prep Work for Storage Drying Cabinets Federal contract opportunity
Solicitation number
36C24421B0019
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 4

View the file

Other files for this federal contract opportunity

Other files attached to Z1DZ--693-21-108 Prep Work for Storage Drying Cabinets, newest first.
File Type Posted
5th Floor Mech Room Layout.pdf PDF
36C24421B0019 0001_1.docx DOCX document
SPSRoomLayout.pdf PDF
36C24421B0019_1.docx DOCX document
S02 Sol 36C24421B0019.docx DOCX document
S02 SafetyHealthDuringConstruction 18S-15-346.pdf PDF
EndoDryCabinets8x11SingleFinal.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ENDODRY™ Pre-Installation Customer Requirements | 1

This document provides the facility with requirements to ensure correct installation and subsequent proper performance of the ENDODRY™ Storage and Drying System.

READ THE ENTIRE DOCUMENT

Applicable Documents

• ENDODRY Storage and Drying System Users Manual

Customer Requirements These criteria must be met to ensure proper operation of the ENDODRY Storage and Drying System. It is the responsibility of the customer to provide qualified personnel to install proper electrical and air supply prior to install.

Air Supply

• The ENDODRY Storage and Drying System requires compressed air.

• The required air pressure is 29-145 PSI (2-10 BAR) for the ENDODRY Storage and Drying System used with the

ADVANTAGE™ Reprocessor and 72-145 PSI (5-10 BAR) for the ENDODRY Storage and Drying System used with the DSD reprocessor.

• The demand of compressed air is a maximum flow rate of 3.4 scfm (95 l/min ANR) for the ENDODRY Storage and

Drying System used with the ADVANTAGE Reprocessor and 7 scfm (200 l/min ANR) for the ENDODRY Storage and

Drying System used with the DSD Reprocessor.

• Air Quality: ISO 8573-1:2010, class 1.4.1.

• Maximum particle size is 1 micron.

• Maximum dew point is +3° C.

• Maximum oil content is 0.01 mg/m3.

• A flexible air pipe with ¼” or 6 mm inner diameter (minimum rating of 150 PSI (10.5 BAR) is required to connect the

ENDODRY Storage and Drying System with the air pressure system. The 5 foot (1.5 meter) line is included in the shipment.

• A facility supplied quick disconnect located on the wall above the ENDODRY Storage and Drying System, see Floor And Wall section.

Electrical Supply

• 100-240 VAC ± 10%, single phase, 50/60 Hz, 15 amp circuit (120v), 10 amp circuit (230v)

• Power input: 75W

• A NEMA 5-15 power cord for the North America market and a CEE 7 power cord for the European market is included with the ENDODRY Storage and Drying System. The cord is approximately 6 feet (1.8 meters) in length.

• A GFI/RCD protected outlet located on the wall above the ENDODRY Storage and Drying System, see Floor And Wall section.

• The power cable of the cabinet serves as a separation device from the power supply. If this cannot be achieved properly, a suitable device must be installed (e.g. central emergency stop).

• For other country-specific electrical requirements contact Medivators.

ENDODRY™

Storage & Drying System | Customer Requirements

Pre-Installation

2 | ENDODRY™ Pre-Installation Customer Requirements

Interface

• The ENDODRY™ Storage and Drying System can be connected to an ADVANTAGE PLUS™ Automated

Endoscope Reprocessor(s) or ENDORA™ Endoscope Tracking System by a LAN interface (RJ45, 10/100 Mbit).

• A 6.5 foot (2 meter) cable is included with the cabinet.

• A RJ45 wall mounted network jack located on the wall above the ENDODRY Storage and Drying System, see Floor and Wall section.

• A network connection to an ADVANTAGE PLUS Automatic Endoscope Reprocessor allows for the transfer of operator and endoscope information.

• Network access is highly recommended for remote access to the ENDODRY Storage and Drying System.

Dimensions Dimensions: H74.8” X W23.6” X D21.3” (190 cm x 60 cm x 54 cm)

Weight: 375 lbs (170 kg)

Floor & Wall

• The floor must be level.

• For optimal user-safety and operation, it is recommended that the ENDODRY Storage and Drying System always be placed directly on the floor, and not on an elevated platform.

• The back of the one-sided ENDODRY Storage and Drying System will be placed flush against the wall. All electrical, network and air connections must be placed a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.

• The Two-sided ENDODRY Storage and Drying System is designed to fit into an opening in the wall. The rough opening for a single unit is H75.2” X W24” (191 cm x 61 cm). All electrical, network and air connections must be placed a minimum of 78.7 inches (200 cm) off of the floor above the cabinet. The electrical, network and air connections must be located on the loading side of the ENDODRY Storage and Drying System.

• The facility is responsible for sealing and/or trimming the ENDODRY Storage and Drying System in the wall.

ENDODRY™ Pre-Installation Customer Requirements | 3

One-Sided Version Two-Sided Version

Minimum facility wall connections height Minimum facility wall connections height

60 cm (23.6”)

54 cm (21.3”)

30 cm (12”)

13 cm (5”)

Air, Electrical, Network

ConnectionsAir, Electrical, Network Connections

Loading Side

4 | ENDODRY™ Pre-Installation Customer Requirements

ENDODRY™ Storage And Drying System Requirements Checklist The ENDODRY Storage and Drying System Installation Checklist is utilized to ensure the installation site meets the required specifications. If all requirements are met, the site is suitable for the installation of the cabinet.

Section Customer Site Requirements Criteria

Air supply provides a minimum of 29 PSI (2 bar) for the ADVANTAGE cabinet and 72 PSI

(5 BAR) for the DSD cabinet, not to exceed 145 PSI (10 BAR)

Air supply provides a maximum flow rate of 3.4 scfm (95 l/min ANR) for the ADVANTAGE and

7 scfm (200 l/min ANR) for the DSD.

ISO 8573-1:2010, class 1.4.1. Air supply maximum particle size is 1 micron, maximum dew point is +3°C, maximum oil content is 0.01 mg/m3.

Air quick disconnect located on the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.

GFI/RCD protected outlet located on the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet.

RJ45 wall mounted network jack located on the wall a minimum of 78.7 inches (200 cm) off of the floor above the cabinet (ADVANTAGE ENDODRY only).

7 Installation site has a level floor.

Installation site accommodates the units dimensions of H74.8” x W23.6” x D21.3”

(190 cm x 60 cm x 54 cm)

By signing below, the customer acknowledges that; (1) The customer has reviewed the Site Requirements for the

ENDODRY Storage and Drying System; (2) The customer is solely responsible for meeting each of the specifications and requirements set forth in these Site Requirements, to ensure proper installation and performance for the cabinet; (3) The customer is able to meet each of these specifications and requirements; and (4) Failure of the customer to comply with these Site Requirements and any deviation from the specifications set forth herein may compromise the functionality of the cabinet, and/or may cause operational issues which will not be covered by the Medivators Inc. warranty, and hereby disclaimed by Medivators.

Facility Representative: Title:

Signature: Date:

Email:

Medivators Customer and Technical Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-421-7696

© 2017 Medivators Inc. All rights reserved. 50098-678-EN REV D

Medivators Inc.

14605 28th Avenue North Minneapolis, MN

55447-4822 USA

Tel: +1.763.553.3300 Fax: +1.763.553.3387

Medivators BV Sourethweg 11 6422 PC Heerlen The Netherlands Tel: +31.45.5.471.471 Fax: +31.45.5.429.595

ESCAD Medical GmbH Zur Öhmdwiesen 5 88633 Heiligenberg Germany Tel: +49 7554 9999 500 Fax: +49 7554 9999 558

ADVANTAGE™, ADVANTAGE PLUS™, ENDODRY™, and ENDORA™ are trademarks of Medivators Inc.

File details come from the government source that posted it. Updated .