36C24421B0019_1.docx

DOCX document 32 KB Posted

Attached to
Z1DZ--693-21-108 Prep Work for Storage Drying Cabinets Federal contract opportunity
Solicitation number
36C24421B0019
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 4

View the file

Other files for this federal contract opportunity

Other files attached to Z1DZ--693-21-108 Prep Work for Storage Drying Cabinets, newest first.
File Type Posted
5th Floor Mech Room Layout.pdf PDF
36C24421B0019 0001_1.docx DOCX document
S02 SafetyHealthDuringConstruction 18S-15-346.pdf PDF
EndoDryCabinets8x11SingleFinal.pdf PDF
S02 Sol 36C24421B0019.docx DOCX document
EndoDryRequirements.pdf PDF
SPSRoomLayout.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

36C24421B0019

SUBJECT*
693-21-108 Prep Work for Storage Drying Cabinets

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE*
18711
SOLICITATION NUMBER*
36C24421B0019
RESPONSE DATE/TIME/ZONE
10-07-2021 4:00 EASTERN TIME, NEW YORK, USA
ARCHIVE
90 DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
N

SET-ASIDE

PRODUCT SERVICE CODE*
Z1DZ
NAICS CODE*
236220
CONTRACTING OFFICE ADDRESS
Department of Veterans Affairs

Network Contracting Office 4 Wilkes-Barre VAMC Acquisitions (049E) 1111 East End Blvd Wilkes-Barre PA 18711

POINT OF CONTACT*

Scot Plank Scot B Plank Scot.Plank@va.gov 570-824-3521

PLACE OF PERFORMANCE

ADDRESS
Department of Veterans Affairs

Wilkes-Barre VA Medical Center

1111 East End Blvd

Wilkes-Barre PA

18711

POSTAL CODE

COUNTRY
USA

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS
scot.plank@va.gov

EMAIL DESCRIPTION

Combined Synopsis/Solicitation Notice

Combined Synopsis/Solicitation Notice

DESCRIPTION

Combined Synopsis/Solicitation Notice

Combined Synopsis/Solicitation Notice

STATEMENT OF WORK

PREP WORK FOR STORAGE DRYING SYSTEM

PROJECT# 693-21-108

1. General:

a. The scope of work shall generally consist of, but not necessarily be limited to, providing all labor, materials, tools, equipment, permits, testing, and reports required to complete site prep for the installation of Endodry Storage and Drying System in Room N4-107 in the Sterile Processing area on the 4th floor of the Wilkes-Barre VA Medical Facility. Any work determined by the COR as having a significant impact on the daily operations of the facility shall be conducted on an after-hours schedule. All work shall be performed in accordance with VA Master Specifications. All federal, state, and local code/regulations/standards shall be followed.

b. The Endodry Storage and Drying System is a system that is already purchased by the VA. This project is for site prep only. The Endodry Storage and Drying System will be provided and installed by the manufacturer under a separate contract after site prep is completed. The scope of work does not require the contractor to install or handle the Endodry Storage and Drying System.

c. The work will consist of installing a medical grade compressed air system to provide the required compressed air for the storage and drying system. The compressors for the compressed air system shall be installed in the 5th Floor Mechanical Room C5-81. The compressed air shall meet quality standards of ISO 8573-1:2010 class 1.2.1.

· Particle Size: Incoming air filtered down to <1 micron in size

· Moisture Content: Pressure dew point of -40⁰C or better required and no liquid water.

· Oil Content: < 0.01 mg of oil per cubic meter of compressed air (liquid, aerosol, or vapor oil

d. For preparation within Room N4-99, there are electrical outlets and air outlets required. There shall be a total of 6 GFI/RCD protected electrical outlets and 6 air quick disconnects installed on the walls above each cabinet at a minimum height of 78.7 inches off the floor.

e. The compressed air system shall be capable of providing air to all 6 cabinets simultaneously and have a N+1 redundancy. Each cabinet has a minimum requirement of 7 scfm to operate. The manufacturer’s recommended compressors are BeaconMedaes LES20-145RD or Powerex LSE2005. Substitutes are permitted but contractor shall be responsible for confirming any substitutes meet the requirements for quality and quantity. All work on compressed air system shall be performed by ASSE certified installers. Before system is placed in service, the fully functioning system shall be inspected and certified.

f. The above scope of work is intended to describe the project in general. The contractor shall abide by all Infection Control Risk Assessment (ICRA) guidelines when performing this project.

g. The contractor shall follow all federal, state, and local codes, regulations, and standards, and all work shall be performed during normal business hours 7:00AM to 4:30PM Monday to Friday excluding national holidays. The work shall be coordinated through the designated Contracting Officer’s Representative (COR).

2. Electrical:

a. The contractor shall supply materials and labor for power connections needed for the compressors selected.

b. The contractor shall supply materials and labor for duplex electrical connections for each cabinet.

See attached document: S02 Sol 36C24421B0019.

See attached document: S02 SafetyHealthDuringConstruction 18S-15-346.

See attached document: SPSRoomLayout.

See attached document: EndoDryRequirements.

See attached document: EndoDryCabinets8x11SingleFinal.