505889_CSR_MSPMRM_NOAA_12NOV2015.doc

DOC document 200 KB Posted

Attached to
PREVENTATIVE MAINTENANCE FOR AB SCIEX SPECTROMETER Federal contract opportunity
Solicitation number
EA-133C-16-RQ-0419
Issued by
Department of Commerce National Oceanic and Atmospheric Administration

About this file

Customer Service Report

View the file

Other files for this federal contract opportunity

Other files attached to PREVENTATIVE MAINTENANCE FOR AB SCIEX SPECTROMETER, newest first.
File Type Posted
505889_MSPMform_NOAA_12NOV2015.doc DOC document
Contractor_Question.docx DOCX document
Combined_Synopsis_Solicitation.doc DOC document
Specifications.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Customer Service Report mSPEC Job # 505889

For Service please have your Serial Number ready and call: 800-879-3758 or by email: service@mspecgroup.com or on the web: www.mspecgroup.com

Customer Ref #
N/A

Customer Info:

MS Job & Instrument Info: FORMCHECKBOX N/A

Hollings Marine Lab/NOAA

API 4000
S0660208
FORMDROPDOWN

Company:

Model:
Serial Number
MS Job Type:

331 Fort Johnson Rd.

B204
API4000
FORMDROPDOWN

Address:

Instrument Location:
Customer ID:
Instrument Status at Completion:
Charleston
SC
29412
USA

Service Form Attached: FORMCHECKBOX Yes FORMCHECKBOX N/A Raw Data Attached: FORMCHECKBOX Yes FORMCHECKBOX N/A

City:
State/Province:
ZIP/Postal Code:
Country:

HPLC Job & System Info: FORMCHECKBOX N/A

Contact Info:

FORMDROPDOWN

FORMDROPDOWN

Emily Pisarski
843-762-8880
System ID:
HPLC Job Type:
System Status at Completion:
Contact Name:
Contact Phone:

Service Form Attached: FORMCHECKBOX Yes FORMCHECKBOX N/A

Billing Info:

Component ID
Manufacturer:
Model:
Serial Number:

FORMDROPDOWN

N/A
N/A
Billing Type:
Alternate Billing Company Name:
Purchase Order #:
N/A
N/A
Billing Contact Name:
Billing Contact Phone:

Metrics Info:

11:26 06-Nov-15
11:26 06-Nov-15
04:30 12-Nov-15
Date/Time Job Received:
Date/Time Tele-Response:
Date/Time Job Completion:

Time Info:

Materials Info:

FORMCHECKBOX

N/A

Date:
Initials:
Arrival Time:
Labour

Hrs:

TravelHrs:

Part Number
Description
Serial Number
Qty:
U/R
WH or RMA #
09-Nov-15
DNC
00:00
0.00
8.00
1012179RX
TV301 pump, F
233583
1
FORMDROPDOWN

DC

10-Nov-15
DNC
09:30
8.00
1.00
MK0027S
4000 PM kit
1
FORMDROPDOWN

DC

11-Nov-15
DNC
10:00
8.00
1.00
FORMDROPDOWN
12-Nov-15
DNC
09:30
7.00
1.00
FORMDROPDOWN
13-Nov-15
DNC
00:00
0.00
8.00
FORMDROPDOWN
FORMDROPDOWN
1002920RX
TV301 controller, refund
223445
1
FORMDROPDOWN

18613a

not used
FORMDROPDOWN
FORMDROPDOWN
FORMDROPDOWN
FORMDROPDOWN
1012179RX
TV301 pumpF, defective
228436
1
FORMDROPDOWN

18613

Warranty
FORMDROPDOWN
FORMDROPDOWN
Labour Total:
23.00
19.00
FORMDROPDOWN

Fault(s) Reported/Fault(s) Observed:

- Customer reported low sensitivity and background contamination. MSPM due.

Actions Taken (list daily actions and major procedures performed/assemblies replaced and why):

- Nov. 9th : Travel from Utica, NY to Charleston, SC

- Nov. 10th : Performed Pre PM evaluation. Found that the sensitivity low and TIC was not stable. Cleaning of ion path necessary.

- Vented instrument and Thoroughly cleaned Q1, Q3, Q0, vacuum interface, and TIS source assemblies. Changed rough pump oil.

- Instrument had a vacuum failure on start up. The TV 301 turbo had a catastophic failure. Ordered replacement for the next day.

- Will have to clean the ion path again due to pump fragments/dust in the vacuum chamber. Instrument left vented.

- Nov. 11th : Removed ion path and throughly cleaned Q1, Q3, Q0, and all ion path assemblies. Cleaned vacuum chamber.

- Cleaned Cacuum interface assemblies. Found that when the turbo blew, it must have broken one of the stand offs on the orifice plate

- from the sudden vacuum failure and increased pressure. Epoxied the stand off back in place on the oriface. Re assembled the ion path.

- Replaced the TV 301 turbo pump and performed the soft start procedure. Instrument pumped down. Left to pump down over night.

- Faulty 301 turbo SN#228436 was installed on 14AUG2015, WO# 505492, Invoice# 17434INV. and should be billed as warranty.

- Nov. 12th : Performed Post PM voltage/performance tests, and mass calibration.

- Calibrated Q1, Q3 positive ion and Q1, Q3 negative ion. All specifications pass.

- Performed Reserpine performance tests. All Reserpine specifications pass. MSPM complete Instrument is operational.

- Nov. 13th : Travel from Charleston, SC to Utica, NY

Resolution/Summary:

Report Sign-off:

mSPEC Service Representative:

Dan Caruso

Signature & Date:
Name:

Authorized Customer Representative:

Emily Pisarski

Signature & Date:
Name:

Above is a detailed report of activities during this service call. Please insert a copy of this report in your maintenance logbook. The customer acknowledges the content of this report with their signature.