ATTACHMENT A - SALIENT CHARACTERISTICS - CHEMIDOC IMAGING SYSTEM.docx

DOCX document 17 KB Posted

Attached to
6630--CHEMIDOC IMAGING SYSTEM Federal contract opportunity
Solicitation number
36C26223Q1366
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22

About this file

This document outlines the salient characteristics for a ChemiDoc imaging system procurement by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22. Key requirements for the imaging system include compatibility with the agency's existing Bio-Rad equipment, 16-bit data acquisition capabilities, a CCD with enhanced sensitivity matching X-ray film, over 4 orders of linear dynamic range, a mechanical zoom maintaining optical resolution, and sample trays for blot, UV, stain, trans-white, and blue applications. The system must provide various illumination sources including UV, white light, blue, green, red, far red, and near infrared as well as chemiluminescent, multiplex imaging, and software compatible with Mac or PC in common file formats. The related federal contract opportunity is solicitation number 36C26223Q1366 for a ChemiDoc imaging system by the listed agency.

View the file

Other files for this federal contract opportunity

Other files attached to 6630--CHEMIDOC IMAGING SYSTEM, newest first.
File Type Posted
36C26223Q1366.docx DOCX document
ATTACHMENT B - SCHEDULE OF PRICING - CHEMIDOC IMAGING SYSTEM.xlsx XLSX spreadsheet

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient characteristics list- ChemiDoc Imaging system Section 1.

1) Unit must be compatible with the current Bio-rad equipment in our laboratory:

a. Bio-Rad PowerPac™ Basic Power Supply/1645050

b. Bio-Rad Trans-Blot Turbo Transfer System/1704150

2) Needs to have the following features for data acquisition, analyses and quality controls:

a. 16-bit data acquisition for publication and data analysis

b. Charge-coupled device (CCD) must reduce background and enhance sensitive to match X-ray film

c. CCD resolution must allow for resolving closely spaced bands on a gel or blot

d. Unit must have >4 orders of linear dynamic range for all samples

e. Mechanical zoom that maintains optical resolution over any field view without compromising sensitivity.

f. Sample trays

i. Sample trays must be Blot/UV/stain-free

ii. White tray for trans-white illumination (for Coomassie silver stain)

iii. Blue tray for nucleic acid applications

iv. Sample tray size: 21 cm x 16.8 cm

g. Illumination source:

i. Trans-UV: 302 nm excitation

ii. Epi-white

iii. Trans-white

iv. Trans-blue: 450-490 nm excitation

v. Epi-blue: 460-490 nm

vi. Epi-green: 520-545 nm

vii. Epi-red: 625-650 nm

viii. Epi-far red: 650-676 nm

ix. Epi-near infrared: 755-777 nm

h. Chemiluminescent filter:

i. 518-546 nm for blue

ii. 577-613 nm for green

iii. 675-725 nm for red

iv. 700-730 nm for red

v. 813-860 nm for infrared

i. Unit must have a safety attachment that protects users from unintentional UV exposure.

j. Must provide software that allows images to be exported in the following file formats: SCN/MSCN, TIF (16-bit or 8-bit), JPG

k. Multiplex image acquisition that allows user to image/assess several antibodies at the same time on the same image

l. Image software must be compatible with MAC or PC and no requirement for license registration

m. Publication ready images with high publishing resolution (DPI)