Attachment 4 Specifications.pdf
PDF 88 KB Posted
- Attached to
- Steris Equipment Federal contract opportunity
- Solicitation number
- 12805B20Q0117
About this file
This document contains specifications for equipment and services required by the Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Plains Area. The solicitation seeks four VHP M100SX Mobile Skid units with dehumidifiers and Vaprox storage mounted on stainless steel carts. Each skid requires a 208 VAC, 30 AMP circuit and includes a 50ft power cord. Two Drager X-am 5100 portable hydrogen peroxide detectors per unit and four adapter and hose connection kits are also required. Startup, calibration, and operator training services for the equipment are included. The response due date is not provided. The equipment and services specified appear to be for continuous decontamination of research spaces using vaporized hydrogen peroxide.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 2 FAR Class Deviation to Implement the United States Mexico Canada Agreement.pdf | ||
| Attachment 3 Bid Schedule (MS Excel Interactive Document).xlsx | XLSX spreadsheet | |
| Solicitation 12805B20Q0117.pdf | ||
| Attachment 1 FAR Deviation Accelerated Payments to Small Business.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SSQ20200716509 M100SX Mobile Skid with Dehumidifier and Vaprox storage 208 VAC Quantity 4
The VHP M100X Biodecontamination Unit provides a flexible, safe, environmentally friendly method for continuous decontamination of larger spaces such as clean rooms, production filling line isolators, food and beverage lines, animal holding rooms, biological (BL2, BL3, BL4) safety labs, and chambers. The VHP M100X is designed for “single-pass” peroxide injection and airflow. This Open-Loop system can be integrated with the facility as a utility to produce shorter cycle times. An operator friendly touch-screen PLC control provides fast and easy operation and cycle data is logged on an impact printer or through an optional Electronic Data Security Package. Multi language control contains English, French, Italian, German, Spanish or Dutch operator screens. The VHP M100SX will be mounted on SS mobile cart with a dehumidifier and H2O2 safety sensor. The skid will require a 208 VAC, 30 AMP circuit for maximum output. The unit will include 50ft power cord with 30 AMP plug.
VD0003 Drager X-am 5100 H2O2 Handheld Sensor, Alkaline Battery Quantity 2
The Drager X-am 5100 portable single-gas detector guarantees you use the safest methods possible for handling H2O2.
VH80 Adaptor Kit, 1.5" Tri-Clamp Quantity 4
1.5" Stainless Steel Tri-Clamp Adapter for flexible connection hose, Tri-Clamp swivel clamp, Tri-Clamp gasket, and hose band clamp. Inlet/Outlet requires 1 kit each.
VD0046 1.25IN BANJO HOSE CONNECTION KIT Quantity 4
The 1.25 in banjo quick connect hose connection kit includes two (2) 1.25 in banjo to 1 in MNPT male adapters for the inlet and outlet of the unit. Two (2) 1.25 in banjo to 1.25 in hose barb female adapters, four (4) hose clamps, and two (2) 10 ft (4 m) 3.18CM hoses are also included.
SE0600225 Start-Up - VHP M100 Biodecontamination Unit Quantity 4
• Execute the STERIS Installation/Start-up Checklist for the equipment
• Verify that building utilities are to the design specification of the equipment
• Verify that the installation was adequately performed and documented
• Verify that the operational requirements are met
Modifications to the structure and/or content of our standard documentation are available for an additional charge. STERIS requires that a minimum of three (3) week’s notice be given to the responsible District Service Manager to allow for scheduling of personnel.
STERIS requires that all of the necessary utilities be available and operational without interruption for the duration of the start-up.
• VHP M100
SE6020322 Calibration - VHP M100/100X Generator Quantity 4
The STERIS Calibration program provides NIST or ISO traceable on-site calibration of process measuring equipment* used in manufacturing and manufacturing-support environments. Qualified STERIS Service personnel using NIST or ISO traceable instrumentation perform all program services according to documented STERIS protocols and procedures.
*Some equipment/components are excluded from our standard calibration services and require NIST or ISO traceable calibration to be performed by the OEM or 3rd party.
• Calibration
• LIFE SCIENCE
VHPAPPTRAIN VHP Onsite Operator Training Quantity 1
VHP operator training includes basics on how to safely operate VHP units. The training includes how to operate a VHP unit including a short demonstration, safe handling of Vaprox hydrogen peroxide sterilant, and facility safety considerations.
File details come from the government source that posted it. Updated .