This document is a Statement of Work (SOW) for specialized refrigeration repair services. The SOW outlines the required services for three refrigeration units: a ThermoScientific minus 80 chest freezer, a Powers Scientific Refrigerated Chamber, and a Powers Scientific Stability Chamber. The required services include replacing compressors, repairing or replacing refrigerant lines, installing dryer filters, and recalibrating the systems.
The related federal contract opportunity is a Request for Quotation (RFQ) for the same specialized refrigeration repair services. The solicitation, number 12805B24Q0581, is issued by the USDA Agricultural Research Service, Plains Area, and has a NAICS code of 811210 with a $34 million small business size standard. The government intends to award a firm fixed-price purchase order to the responsible offeror whose quote conforms to the specifications and is the most advantageous based on technical acceptance, past performance, and price. Quotes are due by 4:00 PM Central Time on August 15, 2024 and should be submitted via email to the specified contracting officer.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFQ 12805B24Q0581 Clauses and Provisions.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
ThermoScientific minus 80 chest freezer:
Replace 1 of 2 compressors for multi-stage cooling system and recalibrate system. This ThermoScientific minus 80 chest freezer requires a compressor replacement and recalibration of the intricate computer system.
Existing refrigerant must be captured and contained prior to replacing existing compressor with new compressor. Installation of the new compressor will require refrigerant lines to be reconnected, resoldered, and refilled with refrigerant. The multi-stage cooling system will require calibration to ensure both compressors are functioning correctly, and the system works properly.
Powers Scientific Refrigerated Chamber DROS33SD2:
Replace compressor, install dryer filters, and recalibrate system. This Powers Scientific Refrigerated Chamber DROS33SD2 requires a compressor replacement, dryer filter installation, and system recalibration.
Existing refrigerant must be captured and contained prior to replacing existing compressor with new compressor. Installation of the new compressor will require refrigerant lines to be reconnected, resoldered, and refilled with refrigerant. The Powers Scientific Refrigerated Chamber DROS33SD2 will require testing to ensure compressor is functioning correctly and the system works properly.
Powers Scientific Stability Chamber B105936 Repair or replace leaking refrigerant lines for Powers Scientific Stability Chamber, install dryer filter, and recalibrate system. This Powers Scientific Stability Chamber B105936 requires the highly sensitive refrigerant lines to be repaired or replaced, dryer filter installation, and system recalibration.
Existing refrigerant must be captured and contained, and refrigerant lines repaired or replaced.
Repair or replacement of the refrigerant lines will require refrigerant lines to be reconnected, resoldered by a highly skilled welder with specialty welding technique, and refilled with refrigerant.
The Powers Scientific Stability Chamber B105936 will require testing to ensure refrigerant lines are functioning correctly and the system works properly.