Attachment 1 Specification Requirements for Capillary Electrophoresis System.pdf

PDF 161 KB Posted

Attached to
Capillary Electrophoresis System - Name Brand or Equal Federal contract opportunity
Solicitation number
12805B23Q0146
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Plains Area

View the file

Other files for this federal contract opportunity

Other files attached to Capillary Electrophoresis System - Name Brand or Equal, newest first.
File Type Posted
Sol_12805B23Q0146.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Minimum Specifications

Item Description:

Capillary Electrophoresis System - Maurice OBM Name Brand or Equal

- Combines two parallel capillary electrophoresis detection schemes (icIEF and CE-SDS) into one fully automated system.

- The full system should include the instrument, computer workstation, Compass for iCE software, reagents and cIEF or CE-SDS cartridges.

- The system must provide identity, purity and heterogeneity data on a wide range of protein samples.

Product Specifications:

- System must deliver isoelectric focused charge and SDS size data via capillary electrophoresis in a minimum sample volume of 50 uL.

- Must get Isoelectric point and charge heterogeneity data in less than 10 minutes.

- Must give high resolution and wide molecular weight range (~35 minutes).

- Must have effective and safe handling methods for use with toxic or sensitive samples.

- Must have proper and functional design to prevent cross-contamination.

- Must have the ability to develop platform methods to use across various molecular makeups.

- Must be highly dependable with Coefficients of Variation (CVs) under 4% daily across various samples, assays, users, instruments, or labs.

- Must have the resources available to maintain compliance and traceability.

- Must have the ability to process up to 200 samples injections/cartridge.

- Must have verified and high-quality dynamic range, linearity, and sensitivity.

- Installation and operational training by a certified technician must be included.

- The operational training should cover instrument startup, operation, maintenance, and troubleshooting.

- A development kit and guide to assist with the calibration and selection of reagents required for method development on the system must be included.

- A training package including one day of on-site training class provided by a protein analysis expert, and related applications scientist must be provided.

Please fill out the matrix below and return a sign copy as part of the submission. The Vendor SHALL comment on each specification as instructed in the Buy Terms, Technical Evaluation of Quotes (Brand Name or Equal).

If bidding exact item, it is acceptable to write “Exact Match” in each of the categories below”. Vendors offering “Equal” shall NOT write “Complies”, “Equal”, “Yes”, etc in the Vendor Submission column of this matrix, vendors shall write the exact specification they are offering as an equal for each line of this matrix. Failure to meet this requirement will result in your offer being found “Technically Unacceptable”.

Brand Name requested: Quip Laboratories Inc.

Signature:

Specifications (Do not include Pricing in matrix) Vendor's Submission: Brand Name or Equal, Please read provision 52.211-6 and all buy terms.

Item No 001: Total eclipse controller – peristaltic pump calibrated in a way to properly dose detergents

Item No 001: Acidulate 45T – safe inorganic acid, safe on Glass and PET, biodegradable cleaning agents, no alkyl phenol ethoxylates, no phosphates, no butyl-based glycol ethers

Item No 001: Enviro-Kleen 1200 – Biodegradable cleaning agents, no alkyl phenol ethoxylates, no butyl-based glycol ethers, no chlorine, no phosphates

Item No 001: AccuPoint Advanced HC System – device with information display to show site name, room number, testing type. Over 900 testing sites able to be tested.

Item No 001:

Item No 001:

Item No 001:

Item No 001:

Item No 001:

Item No 001:

File details come from the government source that posted it. Updated .