This document outlines the salient characteristics for a high temperature gel permeation chromatography system sought under a Department of the Air Force solicitation. The system must be able to operate between 40-220 degrees Celsius, with an autosampler capable of injecting 10-500 microliters and holding 24 or more samples in 2-10mL vials. The dual, independent pumps must precisely deliver flows from 0.1-2.0mL/min and withstand pressures up to 15MPa while temperature controlled between 40-220 degrees Celsius. Additional requirements include a 7.8mm ID x 30cm oven, RI detector with viscometer capability, software for data analysis and report generation, installation, training, and a one-year warranty.

View the file

Other files for this federal contract opportunity

Other files attached to USAFA High Temperature Gel Permeation Chromatography System, newest first.
File Type Posted
Combined Synopsis Solicitation FA700023Q0072.pdf PDF
Attachment 3 - FA700023Q0072 Provisions and Clauses.pdf PDF
Attachment 2 - Buy American Certificate.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SALIENT CHARACTERISTICS

The HT GPC should meet or exceed the following technical specifications when evaluated as described for each value:

Operating Temperature Range

40 ⁰C up to 220 ⁰C

Solvent Reservoir

Capacity: at least 4 L

Temperature: maintained at 40 ⁰C

Autosampler

Injection volume: 10 to 500 mL

Number of samples: 24 or more

Vial size: 2-10 mL injection vials

Temperature: must be able to heat in the range of 40 ⁰C up to 220 ⁰C

Pump

Type: Dual, independent, and temperature controlled (40 ⁰C up to 220 ⁰C)

Max pressure: 15 MPa

Flow Rate: 0.1 to 2.0 mL/min

Precision: +/- 0.2%

Accuracy: +/- 3%

Oven

Capacity: 7.8 mm ID x 30 cm or smaller

Temperature range: 40 ⁰C up to 220 ⁰C

Detector

Type: RI dual flow with a tungsten light source, must be able to add a viscometer detector after initial purchase

Drift: 3 x 10-7 RIU/h

Noise: 1.5 x10-7 RIU/h

Volume: 10 mL

Additional Features/Requirements

All software and hardware (desktop, monitor, printer, etc.) required for data analysis must be included in the quotation. The data file format should easily allow sharing/transfer of data files as individual electronic documents, which are readable by the same data analysis package. The data analysis program should also include a .pdf generator, for the efficient export of analyzed plots.

The quotation must include an on-site installation and training.

Instrument must come with at least 1-year warranty, parts, and labor included, from date of installation.