Att 1 Tech Specs_Temperature and Humidity Chamber.pdf
PDF 132 KB Posted
- Attached to
- Temperature and Humidity (Conditioning) Chamber Federal contract opportunity
- Solicitation number
- SP4706-25-Q-0008
- Issued by
- Defense Logistics Agency
About this file
This document contains technical specifications for a Temperature and Humidity Chamber for the DLA Product Test Center's TA Lab. The procurement objective is to replace a 20-year-old chamber that no longer meets testing requirements, enabling the lab to continue providing high-quality material testing for leather, paper, and textile materials.
Key specifications include a chamber with 40 cubic feet interior capacity, stainless steel interior construction, temperature range of 20°C to 60°C, humidity range of 20%-80%, two hinged glass doors, six adjustable shelves, and a programmable controller with LED display, data logging, and communication capabilities. Additional requirements include a one-year warranty, NIST traceable certification, compatibility with reverse osmosis water, and delivery within 120 days to the DLA Product Test Center in Philadelphia, PA. The contractor must provide complete installation instructions and notify specific DLA contacts upon delivery.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SF1449_SP470625Q0008.PDF |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Technical Specifications Temperature and Humidity Chamber
DLA Product Test Center, TA Lab
Objective:
To obtain a Temperature and Humidity Chamber to be used for testing materials subjected to dynamic conditions over time. Currently, the Product Test Center (PTC) has a chamber that is nearly 20 years old and is no longer capable of achieving required testing parameters. This model chamber is no longer manufactured, and repairs are not as cost effective as a complete replacement.
Procurement of a Temperature and Humidity Chamber will enable the Product Test Center (PTC) to regain the capability of subjecting test specimens to stable environmental conditions for leather, paper, and textile materials designated for testing, which will allow the DLA Product Test Center in continuing its mission in providing high quality reporting to its customers.
The DLA Product Test Center-TA Lab has listed the requirements for the procurement of a Temperature and Humidity Chamber. The following specifications are a list of minimum functions and features required by the DLA Product Test Center.
Minimum Requirements:
1. Specifications:
Temperature & Humidity Stability Chamber -
a) Conditioning: Heating, cooling, dehumidifying, humidifying
b) Volume: Humidity and Temperature Stability Chamber with a minimum interior capacity of 40 Cu. Ft.
c) Exterior dimensions (W x D x H): Maximum dimensions 60” x 43” x 82” (maximum height to include caster wheels)
d) Interior construction: Stainless steel
e) Temperature range: 20°C – 60°C
f) Humidity range: 20%-80%Rh
1. Compatibility - Humidification system must be compatible with Reverse Osmosis (RO) water and must accept a direct plumbed RO water supply
g) Programmable controller: LED display, data logging, adjustable temperature and humidity with ramp and soak function, USB data port, RS-485 communication (MODBUS-RTU), battery backup, and alarm indication of hi/low temp & Rh, and power failure
h) Casters: Equipped with caster wheels
i) Doors: Two hinged, clear view glass doors
j) Shelves: Minimum supplied six (6), adjustable, removable
l) Light: Interior mounted light fixture preferred, but not required
m) Electrical: 115v or 208v, 60Hz, 30A maximum
1. Installation: Contractor shall provide complete instructions to uncrate/unpack, install, setup and operate the chamber. The contractor shall include any components and/or accessories required to operate the chamber (i.e. component connections, cables, power connections, controller firmware setup, OEM manual, etc.)
2. Certification/verification/calibration: Provide instrument certification or verification of instrument performance, to include NIST Traceable ANSI/ NCSL Z540.1 with data certificate for all components and sub-components, as applicable
3. Training: Not required
4. Warranty: Warranty period will cover parts, labor, and travel, for one year
5. Pricing: To include shipping/delivery
6. Notice of delivery: Contractor shall give notice to Edward Francoeur
(Edward.Francoeur@dla.mil 215-737-3244) and Jamie Hieber (Jamie.Hieber@dla.mil 215- 737-3241) when instrument is to be delivered
7. Delivered to/ Mark for:
Attn: Edward Francoeur DLA Product Test Center 700 Robbins St., Bldg. 5-D Philadelphia, PA 19111
8. Period of performance: 120 days from date of award to include delivery
9. Invoices: All invoices must be submitted through Wide Area Workflow (WAWF), See
DFARS Clause 252.232-7006 Wide Area Workflow Payment Instructions for detailed instructions on how to submit invoices
10. DoDAAC: SL0701 (TA Lab) mailto:Edward.Francoeur@dla.mil mailto:Jamie.Hieber@dla.mil
File details come from the government source that posted it. Updated .