Questions and Answers NIST Solicitation No. AMD-TC-RQ-22-01938-2 Fourier-Transform Infrared Spectrometer (FTIR) System

Question 1: Please confirm the minimum 4Hz scan rate specification. Would the Government consider procuring a system with less than the minimum scan rate specification?

Answer 1: No, the Government will not consider a system with less than the minimum scan rate specification. The FTIR must be able to achieve a 4 Hz scan rate.

Question 2: Regarding the gases list, are the required detection and concentration limits also at 4Hz?

Answer 2: Yes, the FTIR must be able to achieve a 4 Hz scan rate.