36C26220Q0508-002.docx

DOCX document Posted

Attached to
6640--Anatomic Pathology Grossing Stations Federal contract opportunity
Solicitation number
36C26220Q0508
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22

About this file

36C26220Q0508 Attachment B - Salient Characteristics.docx

View the file

Other files for this federal contract opportunity

Other files attached to 6640--Anatomic Pathology Grossing Stations, newest first.
File Type Posted
36C26220Q0508-0001001.pdf PDF
36C26220Q0508-0001000.docx DOCX document
36C26220Q0508-003.docx DOCX document
36C26220Q0508-001.xlsx XLSX spreadsheet
36C26220Q0508-000.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment B Brand Name or Equal Description/Requirements:

The Brand Name for the below grossing station is Mopec.

PRODUCT DESCRIPTION
PRODUCT #
QUANTITY
Maestro, Grossing Station, 60" LG, RH
MB1060R
2

Salient characteristics –

1. Each station must be no larger than 70” L x 30” W x 84”H in order to fit both stations into the allotted space in the grossing room.

2. Stainless steel construction is required to eliminate risk of rust from exposure to moisture and for ease of cleaning.

3. Must include lighting above work area and at counter area.

4. Must include ventilation system with exhaust connections less than or equal to 12” round, produce maximum exhaust of 700 cubic feet per minute, and must push air from the front to the back of the cutting surface to prevents formalin exposure to the operator.

5. Must include system controller to monitor air flow and formalin collection in grossing station.

6. Must include fume detection system with audible alarms to alert users of obstruction to airflow.

7. Height adjustability to accommodate users of varying heights between 5’ to 6’4”.

8. Must include sink with running water with at least ½ horsepower garbage disposal which connects to facility plumbing located on right-hand side of grossing station.

9. Must include pneumatic foot pedals for hands-free on/off control of faucet.

10. The sink must have an attached hand-held rinse fixture to allow easy cleaning.

11. Must include on-board formalin collection system on each station.

12. Must include customizable pegboard with the following accessories:

a. Computer monitor mount

b. Requisition holder

c. Marking dye holder

d. Must have approximately 14” shelf and 7” shelf for supplies, for a total of 20” – 24” shelf space

e. Glove bin

13. Must include CPU adjustable mounting bracket

14. Must include magnetic instrument bar

15. Must include customizable mount for the following accessories:

a. Document holder adjustable to eye level

b. Right wing peg board

c. Keyboard arm

d. Monitor arm

16. Work surface must be high quality stainless steel with removable perforated panels and cutting boards with removable legs.

17. Must include cabinet doors to enclose leg space.

18. Must include ventilated system trashcan with wheels to protect user from chemicals fumes.

19. Must include writing ledge.

20. Must include 2 small drawers.

21. Must include 1 large drawer.

22. Must include dictation stand.

File details come from the government source that posted it. Updated .