36C26119Q0621-001.docx
DOCX document 22 KB Posted
- Attached to
- NEW MRI MOBILE TRAILER VA Fresno Federal contract opportunity
- Solicitation number
- 36C26119Q0621
About this file
36C26119Q0621 36C26119Q0621_1.docx
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| -17022.docx | DOCX document | |
| 36C26119Q0621-003.docx | DOCX document | |
| 36C26119Q0621-002.docx | DOCX document | |
| 36C26119Q0621-000.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
FedBizOpps Modification to a Previous Notice
CLASSIFICATION CODE
SUBJECT
CONTRACTING OFFICE'S
ZIP-CODE
SOLICITATION NUMBER
BASE NOTICE TYPE
RESPONSE
DATE (MM-DD-YYYY)
ARCHIVE
DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
SET-ASIDE
NAICS CODE
CONTRACTING OFFICE
ADDRESS
POINT OF CONTACT
(POC Information Automatically Filled from User Profile Unless Entered)
DESCRIPTION
See Attachment
AGENCY'S URL
URL DESCRIPTION
AGENCY CONTACT'S EMAIL
ADDRESS
EMAIL DESCRIPTION
ADDRESS
POSTAL CODE
COUNTRY
ADDITIONAL INFORMATION
GENERAL INFORMATION
PLACE OF PERFORMANCE
* = Required Field FedBizOpps Modification to a Previous Notice Rev. March 2010 W Amendment to MRI MOBILE TRAILER 95652-2609 36C26119Q0621 05-15-2019 N 621512 Department of Veterans Affairs VA Sierra Pacific Network (VISN 21) VA Northern California HealthCare System 5342 Dudley Blvd, Bldg McClellan CA 95652-2609 CS Deborah Sosa No changes deborah.sosa@va.gov Amendment to Solicitation 36C261-19-Q-0621
QUESTIONS AND ANSWERS
QUESTION: On page 4 under “Gradient System” the standard strength is listed at 35 .. is this a typo? Generally, the strength and stew rate are at 33120. I just want to clarify if this was a misprint or not.
ANSWER: Every gradient system is characterized by its degree of linearity, maximum amplitude (strength, measured in mT/m), and rise time (i.e., slew rate, measured in T/m/sec).
The SOW required that the strength is 35 (must have been a typo, should be 33), but it’s not unreasonable to require such (our new Signa Artist will have a strength of 45 mT/m at 200T/m/sec! – this is today’s standards). These factor impact the speed of imaging.
Included the strength and slew rate of some 1.5T MRI systems below.
The question above states a measurement of “33120” which refers to a standard strength in the z-axis of 33 mT/m and a slew rate of 120 T/m/sec, which one be the GE Signa Creator or GE Signa Explorer below.
| Manufacturer |
| GE Healthcare |
SIGNA Artist 1.5T GE Healthcare SIGNA Creator 1.5T GE Healthcare SIGNA Explorer 1.5T
GRADIENT SYSTEM
| Standard name |
| eXtreme |
| Creator |
| Explorer |
| Standard strength, z-axis, mT/m |
| 45 |
| 33 |
| 33 |
| Standard slew rate, z-axis, T/m/sec |
| 200 |
| 120 |
| 120 |
QUESTION: On page 13 the solicitation has the Gray Market Prevention Language - Are you wanting to lease a brand-new machine for only 4-6 months? I ask this because we have an MRI that is fully remanufactured and certified through the OEM. Would this be acceptable?
ANSWER: There is no brand-new system that can be outfitted into a trailer for our purpose. Almost mobile MRI are “pre-owned” or “repurposed”. We do require OEM parts for repairs.