This sources sought notice from the Department of Veterans Affairs seeks responses from qualified service-disabled veteran-owned small businesses, veteran-owned small businesses, and small or emerging businesses to provide a Vision RT AlignRT Surface Guided Radiation Therapy System to the Clement J. Zablocki VA Medical Center in Milwaukee, Wisconsin. Interested firms must respond no later than April 22, 2021 with their company information, capability statement addressing the requirement, and documentation showing authorization to distribute the manufacturer's product. The medical center requires the system to verify patient setup reproducibility, reduce additional imaging needs, and automatically pause treatment for expected or unexpected motion. The system must track patients in real-time, interface with Elekta systems, and include three cameras, a workstation, calibration and adjustment accessories, and support various therapies including breast and head and neck treatments.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Contract Opportunity Sources Sought Notice
PRODUCT SERVICE CODE
SUBJECT
CONTRACTING OFFICE'S
ZIP-CODE
SOLICITATION NUMBER
RESPONSE DATE/TIME/ZONE
ARCHIVE
DAYS AFTER THE RESPONSE DATE
RECOVERY ACT FUNDS
SET-ASIDE
NAICS CODE
CONTRACTING OFFICE
ADDRESS
POINT OF CONTACT
(POC Information Automatically Filled from User Profile Unless Entered)
DESCRIPTION
See Attachment
AGENCY'S URL
URL DESCRIPTION
AGENCY CONTACT'S EMAIL
ADDRESS
EMAIL DESCRIPTION
ADDRESS
POSTAL CODE
COUNTRY
ADDITIONAL INFORMATION
GENERAL INFORMATION
PLACE OF PERFORMANCE
* = Required Field Contract Opportunity Sources Sought Notice AlignRT: Surface Guided Radiation Therapy System 53214-1476 36C25221Q0572 04-22-2021 11:59pm
CENTRAL TIME, CHICAGO, USA
N 334510 Department of Veterans Affairs Great Lakes Acquisition Center (GLAC) 115 S 84th Street, Suite 101 Milwaukee WI 53214-1476 Matthew Hohl Matthew.hohl@va.gov
THIS IS A SOURCES SOUGHT NOTICE ONLY. THIS IS NOT A REQUEST FOR QUOTE.
The NCO 12, Great Lakes Acquisition Center, 115 S. 84th Street, Suite 101, Milwaukee, WI 53214, is performing market research to determine if there is a sufficient number of qualified: (1) Service Disabled Veteran Owned Small Business (SDVOSB); (2) Veteran Owned Small Business (VOSB); or (3) Small or emerging small business firms, or (4) Other Than Small business firms, to provide Vision RT AlignRT Surface Guided Radiation Therapy System [Brand Name or Equal] to the Milwaukee VA Medical Center per the Statement of Work.
General Requirements:
All interested firms who can meet the requirements stated below should respond, in writing, INCLUDING ALL THE FOLLOWING:
1. Company name.
2. Address.
3. Point of contact (name/title/telephone number/e-mail address).
4. FSS / SEWP Contract number, if applicable.
5. DUNS number.
6. Tax ID number.
7. Indication of which business category (SDVOSB, VOSB, Small Business or Other Than Small Business) for which the organization qualifies.
8. A capability statement that addresses the organizations qualifications and ability to provide the requirement depicted above.
9. Documentation from manufacturer (i.e. correspondence from manufacturer) of your firm being an authorized distributor for the manufacturer.
Submit responses to the Contracting Officer, Matthew Hohl, at Matthew.Hohl@va.gov no later than 11:59 PM (Central Standard Time), 22 April 2021.
“36C25221Q0572 Surface Guided Radiation Therapy System” shall be referenced on all correspondence regarding this announcement.
36C25221Q0572 AlignRT
Purpose The Clement J. Zablocki VA Medical Center Radiation Oncology department requires the purchase of the Align RT Surface Guidance system manufactured by Vision RT. As part of new construction efforts to expand Radiation Oncology, new linear accelerators have been purchased which work in conjunction with surface guidance systems and applications. These systems are brought in place to verify patient setup reproducibility, reduce the need for additional imaging and imaging dose, and automatically pause treatment due to expected (planned breathe hold) and unexpected (patient adjustments mid-treatment) motion.
With the projected increase in women veteran population, this system will also minimize concerning cardiac doses when treating certain breast cancer cases. It will also reduce patient setup times by %50 with the use of automated interlocking systems which will increase patient safety and satisfaction. All equipment must follow the guidelines and meet all specifications listed below.
Project Scope:
A. Functional Requirements
· Surface Guided Radiation Therapy (SGRT) system that allows real-time tracking of patients before and during radiation therapy.
· Functions must use contact-free techniques to track patients skin surfaces with sub-millimetric accuracy ranging below 0.5mm in all six degrees of freedom.
· Ensures radiation is only delivered when patients are correctly positioned.
· System must ensure accurate and reproductible patient setup and monitoring during dose delivery.
· System must have the ability to automatically pause treatment due to expected and unexpected patient adjustments made during dose delivery and treatment.
· Must allow control to the Elekta Couch Move functions
System must be able perform the following therapies and examinations
· DIBH (Deep Inspiration Breath Hold)
· General Breast
· SRS (Stereotactic Radiosurgery)
· SBRT (Stereotactic Body Radiation Therapy)
· Head and Neck
· Breath Hold SBRT
· Sarcoma
· Prostate
· Pediatrics
B. Software & Interfacing Requirements
· System must be able to interface with Elekta’s Gating (Beam Hold) “Response” Interface to the Elekta Linac and continue to integrate with any future upgrades Elekta has to offer.
· Must be able to automatically open/review patients in synchronization with MOSAIQ
i. Sends and receives new patient appointments and updates existing patient appointments while sending reports to MOSAIQ.
· Software must support a 3-camera system and include a 3D photo module that will map the overlay of the texture-map of the patient on a 3D surface.
· Must include a postural video module that can provide live stream video from any of the three (3) camera units with a high contrast position outline for use during patient setup and treatment.
· Patient tracking software must provide multi-angle, real-time view of patient’s overall alignment relative to their reference position
· Must include analysis software that will work in conjunction with a stereotactic calibration cube phantom and leveling plate.
C. Additional Features/Accessories The system must include and meet the following features/guidelines
· Three (3) HD Camera Units
· Power Supply Unit (PSU) included for camera
· One (1) Workstation with a remote console in the control room
· Keyboard, mouse and monitor provided
· Three (3) Portable devices that will allow for remote control of the key functions of the system software.
· One (1) Head adjuster for applying pitch, yaw and roll corrections for cranial treatments.
· One (1) Stereotactic calibration cube phantom and leveling plate.
· One (1) Workstation hardware for an Offline Workstation
Provided Bill of Materials from Vision RT The following list of equipment provided meet our requirements for the SGRT System needed
| Part Number |
| Description |
| Qty |
| ALRT-PS3C-STD |
| AlignRT: 3 Camera System Real Time Patient Positioning, Tracking and Surveillance |
| 1 |
| Included |
| AlignRT HD camera unit |
| 3 |
| AlignRT software upgrade: 3 camera support |
| 1 |
| AlignRT workstation; Remote console in control room |
| 1 |
| AlignRT Patient Tracking Software |
| 1 |
| DICOM RT Import Module |
| 1 |
| AlignRT calibration plate |
| 1 |
| PSU for AlignRT HD Camera |
| 1 |
| Portable device to allow the remote control of key functions of AlignRT software |
| 3 |
| ALRT-ELEKTA RESPONSE |
| Interface to Elekta’s Gating (Beam Hold) “Response” Interface |
| 1 |
Vision RT's interface to Elekta's Gating (Beam Hold) "Response" Interface to the Elekta Linac. Customer must purchase the "Response" interface from Elekta.
| SW-ELEKTA-PPS |
| SW-ELEKTA-PPS Interface to Elekta PPS |
| 1 |
Enables opening patients in AlignRT from MOSAIQ and controlling Elekta Couch Move functions from AlignRT, for systems configured with the Elekta PPS interface.
| SW-ACO-CALIB |
| Advanced Camera Optimization |
| 1 |
Advanced optical set-up and verification of the AlignRT system.
| ALRT-3DP-NOEXP |
| Pricing for perpetual access to 3D Photo module |
| 1 |
Monochrome texture mapping to overlay photo-realistic texture-map of patient on 3D surface.
| SRS-HEADADJ |
| 3Degree Head Adjuster for cranial radiotherapy treatments |
| 1 |
3Degree head Adjuster for applying pitch, yaw and roll corrections for cranial treatments.
| SRS-CALIB-Mk2-STD |
| Stereotactic calibration cube phantom, levelling plate and software |
| 1 |
Stereotactic calibration cube phantom and levelling plate. Stereotactic calibration module includes analysis software
| ALRT-RPW |
| Workstation hardware for Offline Workstation |
| 1 |
Workstation hardware for Offline Workstation. AlignRT v.5x software for Offline Workstation
| LA-POSTURAL-NOEXP |
| Pricing for perpetual access to Postural Video module |
| 1 |
Monochrome live video of patient from any camera unit with overlay of high contrast reference position outline for use during patient setup and for visualization during treatment
Deliverables or Delivery Schedule The contractor shall deliver and install the product. After contacting COR (or COR designee) at the VA, a specific time will be scheduled for the delivery of the product and the installation in accordance with the project schedule award. All quantities are listed above that should be delivered.
They shall be delivered to the following address:
Attn: VA Biomedical Engineering Clement J. Zablocki VAMC Building 111, warehouse 5000 West National Avenue Milwaukee, WI 53295
The Vendor shall confine all operations (including storage of materials) on Government premises to areas authorized and approved by the Contracting Officer. Working space and space available shall be as determined by the COR.
Workmen are subject to rules of the Medical Center applicable to their conduct. All personal automobiles and contractor trucks shall be parked at the contractor's staging area or offsite of the VA premises.
Execute work to interfere as little as possible with normal functioning of the Medical Center as a whole, including operations of utility services, fire protection systems and any existing equipment, with work being done by others. Do not store materials and equipment in other than the designated contractor storage areas.
Daily, Vendor shall keep work, storage, and staging areas clean and neat. Vendor shall provide enough trash containers so that there is no debris lying around. Debris will be removed daily by the Vendor unless otherwise directed.
Hours of Operation Normal hours of coverage are Monday through Friday, 7:00 AM to 4:30 PM Central Standard Time excluding national holidays. All services shall be performed during normal hours unless otherwise scheduled with the COTR or his/her designee, except for scheduled training. Contractor shall always be professional and shall take all necessary precautions to maintain a safe environment for the occupants. Contractor shall fully minimize noise possible while working in all buildings and wear all required Personal Protective Equipment (PPE).
Federal Holidays observed by VA Milwaukee are:
| New Year’s Day | Labor Day | ||
| Martin Luther King Day | Columbus Day | ||
| Presidents’ Day | Veterans’ Day | ||
| Memorial Day | Thanksgiving Day | ||
| Independence Day | Christmas Day |
Place of Performance The place of performance will be at Clement J Zablocki VA Medical Center, Milwaukee, WI.
Period of Performance The Contractor warrants that the items will be free from defects from materials and workmanship under normal use for a period of not less than one (1) year for equipment defined herein. This start date of this period shall be one-year from go-live of the system and shall be considered the base-year. Contractor will replace or repair equipment if necessary, to the satisfaction of the Government. Contractor shall provide a copy of the written warranty upon acceptance of the equipment, which will be establish base on the go-live date.
All labor, time, materials, equipment, travel, and supplies necessary to keep equipment operational during the warranty period shall be included. Phone response to warranty issues within 1 hour and on-site response within 48-hours is required.
TELEPHONIC INQUIRIES WILL NOT BE ACCEPTED.
The NAICS Code 334510, Electromedical and Electrotherapeutic Apparatus Manufacturing, is applicable to this acquisition; the size standard is 1,250 employees. Interested firms are reminded that in accordance with FAR 4.12, prospective contractors shall complete electronic annual representations and certifications in conjunction with FAR 4.11 required registration in the Systems for Award Management (SAM) database prior to award of a contract. The Government will use responses to this notice to make an appropriate acquisition decision. This is the only notice; no other copies of this notice will be provided.