36C25019B0025-0002001.pdf
PDF 4 MB Posted
- Attached to
- Pharmacy IV Prep Room Upgrade Chillicothe VA Medical Center 17273 St. Rt. 104 Chillicothe, OH Federal contract opportunity
- Solicitation number
- 36C25019B0025
About this file
36C25019B0025 0002 S02. Chilli IV Prep Room Bid Drawings Rev 1.pdf
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C25019C0185-000.docx | DOCX document | |
| 36C25019B0025-0003004.docx | DOCX document | |
| 36C25019B0025-0002000.docx | DOCX document | |
| 36C25019B0025-0001000.docx | DOCX document | |
| 36C25019B0025-014.pdf | ||
| 36C25019B0025-011.docx | DOCX document | |
| 36C25019B0025-013.pdf | ||
| 36C25019B0025-012.pdf | ||
| 36C25019B0025-000.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
N
EXTERIOR ELEVATION
INTERIOR ELEVATION
SEE 700 SERIES SHEETS
SECTION/DETAIL NUMBER
DRAWING NUMBER
EQUIPMENT TAG
COLUMN GRID
NOTE
WINDOW OPENING
DOOR OPENING
ROOM NAME
ROOM NUMBER
WALL TYPE
SEE SHEET A001
CEILING HEIGHT
FINISH DESIGNATION
X
10'-8"
X
X
XX.X-X
XX
X
1/A101
X
XXXX
X
SIM
XX
TOILET ACCESSORY TAGX SIGNAGE TAGX
U S 35
US 35
US 35
SR 104
EXIT
S R
N . H
IG
H
S T
R E
E T
C R
SCIO
TO R
IVER
S R
N
B
R
ID
G E
S T
R E
E T
Department of Veterans Affairs Medical Campus
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
JPA Project No.
Department of Veterans Affairs Medical Center
ARCHITECTURE PLANNING INTERIOR DESIGN
3131 NEWMARK DRIVE, SUITE 200
MIAMISBURG, OHIO 45342
937.461.3290 P
JOHN POE ARCHITECTS
HEAPY ENGINEERING
MECHANICAL ELECTRICAL COMMISSIONING TECHNOLOGY
1400 WEST DOROTHY LANE
DAYTON, OH 45409
937.224.0861 P
937.224.5777 F
Dwg. of
Terracon Consultants 611 Lunken Park Dr, Cincinnati, Ohio 45226 513.321.5816
Pharmacy IV Prep Room Renovation
10-26-2018
COVER SHEET
G001
SOJR
18003
538-18-201
Chillicothe VA
Pharmacy IV Prep Room Renovation
17273 State Route 104 Chillicothe, Ohio 45601 PROJECT #: 538-18-201
G001 COVER SHEET
G100 CODE DATA AND PHASING PLANS
A102 FLOOR PLANS
A103 ATTIC PLAN
AF101 FINISH PLANS
IF102-S SIGNAGE PLAN
FX001 LEGEND, ABBREVIATIONS, INDEX OF DRAWINGS
FX102 PARTIAL SECOND FLOOR PLAN
P001 LEGEND, ABBREVIATIONS, INDEX OF DRAWINGS
P102 PARTIAL FIRST & SECOND FLOOR PLANS
M001 LEGEND, ABBREVIATIONS, INDEX OF DRAWINGS
MD102 PARTIAL FLOOR PLANS - REMOVALS
MH102 PARTIAL FLOOR PLANS - DUCTWORK
MP102 PARTIAL FLOOR PLANS - PIPING
M501 DETAILS
M601 SCHEDULES
M701 CONTROLS
E001 LEGEND, ABBREVIATIONS, INDEX OF DRAWINGS
ED102 PARTIAL SECOND FLOOR PLAN - REMOVALS
E102 PARTIAL SECOND FLOOR PLANS
E103 PARTIAL FLOOR PLAN - ATTIC
E501 DETAILS & SCHEDULES
E601 TECHNOLOGY DETAILS & SCHEDULES
GENERAL CONSTRUCTION NOTES
8. CONTRACTOR SHALL PROVIDE WOOD BLOCKING AT TOILET ACCESSORIES, HAND
RAILS, BUMPER RAILS, SHELVING, CASEWORK, MOUNTING BRACKETS AND OTHER
WALL MOUNTED ITEMS, AS INDICATED IN THE CONTRACT DOCUMENTS AND AS
RECOMMENDED BY PRODUCT MANUFACTURERS. ALL WOOD BLOCKING SHALL BE
FIRE RETARDANT TREATED.
9. ALL EXPOSED NEW WORK IS TO RECEIVE NEW FINISHES UNLESS SPECIFICALLY
NOTED OTHERWISE. iF NO FINISH IS INDICATED AT A PARTICULAR SURFACE, CONTRACTOR SHALL PROVIDE FINISH(S) AS INDICATED AT SIMILAR CONDITIONS.
10. CONTRACTOR SHALL PROVIDE COMPLETE AND FULLY OPERATIONAL SYSTEMS WHICH
COMPLY WITH STATED CODES AND REGULATIONS. WHERE THE OMISSION OF A PART
OR ELEMENT OF A SYSTEM WOULD RESULT IN THE NON-OPERATION, OR INCORRECT
OPERATION OF A SYSTEM, CONTRACTOR SHALL INCLUDE SAID PART OR ELEMENT AS
PART OF THE WORK.
11. ALL PENETRATIONS THROUGH NEW AND/OR EXISTING SMOKE OR FIRE RATED WALLS
SHALL BE SMOKE/FIRE STOPPED AS REQUIRED BY THE RATING OF THE WALL.
12. FIRE PROTECTION CONTRACTOR'S SPRINKLER DESIGN WILL BE SUBJECT TO AN
INDEPENDENT REVIEW BY A CERTIFIED FIRE PROTECTION ENGINEER RETAINED BY
THE OWNER. THE FIRE PROTECTION CONTRACTOR IS RESPONSIBLE FOR MAKING
ANY REQUIRED CHANGES TO THE DESIGN AS IDENTIFIED BY THE OWNER'S FIRE
PROTECTION ENGINEER AS NECESSARY FOR COMPLIANCE WITH SPECIFIED NFPA
REQUIREMENTS.
13. GC TO HIRE LABORATORY CERTIFICATION SERVICES INC, COLUMBUS, OHIO (614)
486-078 CLEAN ROOM CERTIFIER TO CERTIFY CLEAN ROOMS PRIOR TO TURNING
OVER TO VA FOR USE. IF ROOM DOES NOT PASS LCS’S CERTIFICATION GC IS
RESPONSIBLE TO MAKING REPAIRS TO BE CERTIFIED AND RE-CERTIFY AFTER
DEFICIENCIES HAVE BEEN CORRECTED.
14. GC TO HIRE LABORATORY CERTIFICATION SERVICES INC, COLUMBUS, OHIO (614)
486-078 CLEAN ROOM CERTIFIER TO CERTIFY HOOD AFTER RE-INSTALLATION IN
NEWLY CONSTRUCTED SPACE.
* PLEASE NOTE THAT THE TERM PROJECT ENGINEER REFERS TO THE CONTRACTING
OFFICER'S TECHNICAL REPRESENTATIVE.
1. ALL WORK SHALL COMPLY WITH THE LATEST NFPA 101 LIFE SAFETY CODE,
INTERNATIONAL BUILDING CODE, AND OTHER COES AS IDENTIFIED IN VHA PROGRAM
GUIDE PG-18-3, TOPIC 01, CODES, STANDARDS AND EXECUTIVE ORDERS WHICH CAN
BE FOUND ON THE DEPARTMENT OF VETERAN AFFAIRS, TECHNICAL INFORMATION
LIBRRARY WEBSITE.
2. ALL WORK SHALL COMPLY WITH REQUIREMENTS OF THE LATEST ARCHITECTURAL
BARRIERS ACT ACCESSIBILITY STANDARD (ABAAS) AND DEPARTMENT OF VETERANS
AFFAIRS BARRIER FREE DESIGN GUIDE PG-18-13 WHICH CAN BE FOUND ON THE
DEPARTMENT OF VETERANS AFFAIRS TECHNICAL INFORMATION LIBRARY WEBSITE.
3. CONTRACTOR IS RESPONSIBLE TO VISIT THE SITE, EXAMINE AND ACCEPT EXISTING
CONDITIONS PRIOR TO BIDDING. SITE VISITS TO BE COORDINATED WITH THE
CONTRACTING OFFICER'S REPRESENTATIVE.
4. CONTRACTOR IS RESPONSIBLE TO REPAIR AND/OR REFINISH, TO MATCH ADJACENT
EXISTING SURFACES, ANY EXISTING MATERIALS TO REMAIN THAT ARE DAMAGED
DURING THE COURSE OF DEMOLITION OR NEW WORK.
5. WHERE REMOVAL OF EXISTING WORK IS REQUIRED FOR INSTALLATION OF NEW
WORK, CONTRACTOR SHALL REPAIR EXISTING CONSTRUCTION TO REMAIN OR
PROVIDE NEW CONSTRUCTION TO MATCH EXISTING ADJACENT SURFACES TO
REMAIN.
6. CONTRACTOR IS RESPONSIBLE TO FIELD VERIFY ALL CONDITIONS AND DIMENSIONS
PRIOR TO ORDERING, FABRICATING OR INSTALLING NEW MATERIALS.
7. CONTRACTOR IS RESPONSIBLE TO VERIFY SIZES OF ALL EQUIPMENT, ETC. TO BE
INSTALLED AS PART OF THIS PROJECT, AND WHERE NECESSARY, MAKE SPECIAL
PROVISIONS TO INSTALL EQUIPMENT THAT IS TOO LARGE TO FIT THROUGH FINISHED
OPENINGS.
DRAWING INDEXVICINITY MAP
GRAPHIC SYMBOLS
Revisions Date
0-HR Smoke Partition - OBC 710
1-HR Smoke Barrier - OBC 709
1-HR Fire Barrier - OBC 706
2-HR Fire Barrier - OBC 706
Means of Egress
Existing Wall to Remain
New Wall
CODE LEGEND
Non Hazardous
Drug Compounding
2-307
Ante
2-303B
Sterile Hazardous
Drug Compounding
2-303C
Hazardous Drug Storage
2-303A Entry
2-303
Elect.
2-304
Toilet
2-305
Staff Toilet
2-306
Anti-Neoplasty
2-308
Inpatient Pharmacy
2-310
Outpatient Pharmacy
2-300
Narcotics Safe
2-302
Anti-Neoplasty
2-301
Stair #2
S-202
Outpatient Supervisor
2-204
Social Worker
2-205
Vacant
2-206
Corridor
2C002
Tele Clos.
2-203
Elevator
2-300F
86'-6"
4'-0"
/2
4'-0"
T1.0
T1.0 T1.0
T1.0
T1.0T1.0
T1.0
Area of Work Elevator use for construction.
Contractor to provide protection for elevator as stipulated in the specification.
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
XX
JPA Project No.
Dwg. of
Pharmacy IV Prep Room Renovation
10-26-2018
CODE DATA AND PHASING PLANS
G100
SOJR
18003
538-18-201
Chillicothe VA
N
Scale:
PROJECT
1/8" = 1'-0"
CODE PLAN- SECOND FLOOR
Code Review Chillicothe VA B31 IV Prep Room 2018 NFPA 101 Life Safety Code Review
Chapter 6- Classification of Occupancy and Hazard of Content
6.1.5 Business (Existing)- No Change
6.2.2.- Hazard Classification: Ordinary- No change Chapter 7- Means of Egress: No Change 7.3.1- Occupant Load: No Change
Chapter 8- Features of Fire Protection: No Change Chapter 9- Building Service and Fire Protection Equipment: No Change Chapter 10- Interior Finish, Contents, and Furnishes 10.2.3.1- Interior wall and Ceiling Finish Materials tested in accordance with
NFPA 286.
10.2.3.1- Interior wall and ceiling finish materials shall be classified in accordance with NFPA 286 and comply with 10.2.3.2.
10.2.3.1.2.- Materials tested in accordance with 10.2.3.1.1 and complying with 10.2.3.2 shall be considered also to comply with the requirements of Class A, Class B, or Class C in accordance with 10.2.3.3.
(1) During the 40 kW expsure, flames shall not spread to the ceiling.
(2) The flame shall not spread to the outer extremity of the sample on any wall or ceiling.
(3) Flashover, as described in NFPA 286, shall not occur.
(4) The peak heat release rate throughout the test shall not exceed 800 kW.
(5) For new installations, the total smoke released through-out the test shall not exceed 1000 m2.
10.2.3.3- Wall and ceiling- Classifications characteristics.
(1) Class A: Flame Spread: 0-25; Smoke Dev. : 0-450.
(2) Class B: Flame Spread: 26-75; Smoke Dev. : 0-450
(3) Class C: Flame Spread: 76-200; Smoke Dev. : 0-450
10.2.7.3- Interior Floor Finish- Shall be classified based on NFPA 253 or
ASTEM E 648.
10.2.3.4.- Interior Floor- Classifications Characteristics:
(1) Class I: Rad Flux: Fin .45 W/CM2 (2 Class II: Rad Flux: Fin .22 W/CM2, and max. .45 W/cm2 10.2.8.1- Wall and Ceiling finishes, where sprinkled:
Class C permitted in any location where Class B is req’d.
Class B permitted in any location where Class A is req’d.
10.2.8.2.- Floor Finishes, where sprinkled:
Class II permitted in any location where Class I is req’d.
No critical Rad. Flux permitted in any location where Class II is req’d.
Chapter 39- Existing Business Occupancies:
39.1.1.1- The requirements of this chapter shall apply to existing buildings or portions thereof currently occupied as business occupancies.
39.2.2.7.- Sensor- release of electrical locking systems compling with
7.2.1.6.2 shall be permitted.
39.2.5.2- Dead- end corridors shall not exceed 50 ft.
39.2.5.3- Limitations on common path of travel shall be in accordance with 39.2.5.3.1.
39.2.5.3.1.- Common path of travel shall not exceed 100 ft. on a story protected throughout by an approved automatic sprinkler system in accordance with 9.7.1.1 (1).
39.2.6- Travel distance to exits. Travel distance shall comply with 39.2.6.3 through 39.2.6.3.
39.2.6.3.- Travel distance shall not exceed 300 ft in business occupancies protected throughout by an approved, supervised automatic sprinkler system in accordance with Section 9.7.
39.2.9.1- Emergency lighting shall be provided in accordance with Section
7.9 in any building where any one of the following conditions exists:
(1) The building is three or more stories in height.
(2) The occupancy is subject to 100 or more occupants above or below the level of exit discharge.
(3) The occupancy is subject to 1000 or more total occupancies.
39.3.3. Interior Finish
39.3.3.1. General. Interior finish shall be in accordance with Section 10.2.
39.3.3.2.1.- Interior wall and ceiling finish materials complying Section 10.2.
shall be Class A or Class B in exits and in exit access corridors.
39.3.3.2.2.- Interior wall and ceiling shall be Class A, Class B, Class C in areas other than those specified in 39.3.3.2.1.
39.3.4.1. General. A fire alarm system in accordance with Section 9.6 shall be provided in all business occupancies where any one of the following conditions exists:
(1) The building is three or more stories in height.
(2) The occupancy is subject to 100 or more occupants above or below the level of exist discharge.
3. The occupancy is subject to 1,000 or more total occupants.
39.3.4.2 Initiation. Initiation of the required fire alarm system shall be by one of the following means:
(1) Manual means in accordance with 9.6.2.1(1)
(2) Approved automatic fire detection system in accordance with 9.6.2.1 (2) that provides protection throughout the building and the provision of 9.6.2.6.
shall apply.
(3) Approved automatic sprinkler system in accordance with 9.6.2.1. (3) that provides protection throughout the building and the provision on 9.6.2.6 shall provide.
39.3.5 Extinguishment Requirements. Portable fire extinguishers shall be provided in every business occupancy in accordance with Section 9.9.
A. For temporay partition locations see Phasing Plans. Temporary partitions are suggested locations.Coordinate final location in field with C.O.R.
B. Verify all conditions in the field prior to fabrication, erection and construction.
C. All asbestos containing material (ACM) must be abated per the direction of the C.O.R. and project specification 017419 and local state and federal regulations. Abatement if required is to be scheduled in appropriate sequence with demolition, new construction and project phasing.
D. Coordinate phase completion and start-up with requirements as set forth by the C.O.R. to maintain daily operations to near as normal as possible.
E. Coordinate location of job dumpster with the C.O.R. so as to disrupt daily operations as little as possible.
F. All constructrion material deliveries must be delived to the Area of Work from 6 pm-10 pm.
F. Construction working hours are restritected to 2 pm-10 pm. Debris removals from Area of Work is restricted to 6pm-10pm. Loud construction activities restricted to 6pm-10pm. No loud music.
G. Temporary door locations and sizes are as required by the General Contractor. Doors will interfere with daily operations as little as possible. Confirm all locations with the C.O.R.
H. Existing HVAC, medical gases, sprinklers and water supply to be maintained throughout as required during each phase of construction.Coordinate phasing with Mechanical and Plumbing contractor. Support and protect all lines during each phase of construction. See MEP Drawings.
J. Provide and maintain negative air pressure at all construction.
Coordinate with HVAC contractor.
K. Temporary partition to structural deck above. See A102 for temporary partition type.
L. Existing power must be maintained throughout as required during each phase of construction.Coordinate phasing with Electrical Contractor. Coordinate power turnover with requirements as dictated by the C.O.R. See Electrical Drawings.
M
Coordinate sequence construction with C.O.R. so that only one side of any corridor in use is obstructed at any one time.
P. Utilities must be maintained. Any required utility disruptions are restricted to 6pm-10pm. Contractor to coordinate with C.O.R for any disruption of utilities.
N
Scale:
PROJECT
1/8" = 1'-0"
PHASING PLAN- SECOND FLOOR
GENERAL PHASING NOTES
12 1314 15
60 636465
Sterile Hazardous
Drug Compounding
2-303C
Entry
2-303
Hazardous Drug Storage
2-303A
Ante
2-303B
Non Hazardous
Drug Compounding
2-307
Anti-Neoplasty
2-308
Staff Toilet
2-306
Toilet
2-305
Inpatient Supervisor
2-309
Inpatient Pharmacy
2-310
Anti-Neoplasty
2-301
Stair #2
S-202
Narcotics Safe
2-302
Elect.
2-304
Outpatient Pharmacy
2-300
9'-7 1/8" +/-
4 1/4"
7'-6 7/8" +/-
/8
/8
/4
/4
7'-6 1/4" +/-
6'-9" +/-
4 7/8"
10'-4 3/8"
2-303A
2-307A
2-303
2-307
S2.0
2'-5 1/8"
4 7/8"
1'-1" +/-
7' - 0"
Align
Align
Align
B
A
A F
8.09
8.09
S1.1
S2.0
S2.0
/4
1'-7 1/4"
1'-5 1/4"
7.04
7.04
S2.0
7.04
S2.0
S2.0
S2.0
S2.0
Align
2.04
2'-8"
27.01
27.01
8.64 8.63
8.63
9.13
9.13
9.13
2-303
2-303A
2-303C
2-303B
2-307
2-300
9.30
8.14
8.158.15
8.148.14
8.15
12 1314 15
60 61 62 636465
2.12
2.10
2.08 2.072.07
2.07
2.132.13
2.11
2.11
2.13
2.092.09
2-303A
2-303
2-303C
2-303B
2-307
2.04
2.142.14
2.14
REFLECTED CEILING PLAN LEGEND
Supply Air Diffuser
Return Air Grille
2' x 4' fluorescent fixture
2' x 2' acoustical tile ceiling
Gypsum board / portland cement plaster ceiling / soffit - painted
Access panel
S1.0S1.1S2.0
FLOOR SLAB
FIN. CEILING
STRUCTURE ABOVE
4 7/8"
9.219.21
9.02
9.80
7.82
7.82
9.219.21
9.03
9.80
7.82
7.82
7 1/4"
9.02
7.13
7.82
7.82
4 1/4"
9.80
T1.0
7.82
4 1/4"
7.82
9.02
9.21
9.21
31-SECOND FLOOR
754' - 11 5/128"
/8
/8
2'-5 1/4"
8" 1' - 1 3/8" 8"
8.09
8.26
A102
A102
31-SECOND FLOOR
754' - 11 5/128"
/8
/8
1'-6 7/8" 1'-3 1/2" 1'-6 7/8"
4'-5 1/4" +/-
8.09
8.26
A102
A102
SEE SCHED.
S E
E S
C H
E D
U L E
F
2"
SEE
SCHED. 2"
S E
E S
C H
E D
U L E
A
SEE PLAN
1/2" 1/2"
H1 J1
8.02
9.21
9.80
8.02
9.09
9.21
9.80
9.08
7.82 7.82 1/2" 1/2"
SD
SEE SCHEDULE
S E
E S
C H
E D
U L E
9.21
9.80
6.01
9.02
9.14
1/2" 1/2"
8.26
7.83
S1
SEE PLAN
9.21
9.02
9.21
9.39
DEMOLITION PLAN LEGEND
Wall to be Demolished
Existing Wall to Remain
New Wall
S E
E P
L A
N
9.21
9.80
6.01
9.02
9.21
8.26
7.83
S2
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
XX
JPA Project No.
Dwg. of
Pharmacy IV Prep Room Renovation
10-26-2018
FLOOR PLANS
A102
SOJR
18003
538-18-201
Chillicothe VA
GENERAL PLAN NOTES
A. All dimensions are taken to face of gypsum wallboard or unit masonry.
B. Edge of all door jambs at hinge side not otherwise indicated are to be 6" from the face of intersecting walls. Edge of all door jambs at strike side not otherwise located are to be 18" minimum from the face of intersecting walls.
C. All wood blocking to be fire retardent treated.
D. All partitions shall be type 'S1.0' unless noted otherwise.
N
Scale:
PROJECT
1/4" = 1'-0"
NEW WORK PLAN- SECOND FLOOR
N
Scale:
PROJECT
1/8" = 1'-0"
REFLECTED CEILING PLAN- SECOND FLOOR
N
Scale:
PROJECT
1/8" = 1'-0"
DEMOLITION PLAN- SECOND FLOOR
DOOR SCHEDULE
DOOR
BETWEEN DOOR DESCRIPTION FRAME DESCRIPTION
FIRE
RATING
HDWE
SET REMARKSRM RM TYPE MAT'L
SIZE
TYPE MAT'L HEAD JAMB# W H T
2-303 2-303B 2-303C 7' - 0" 7' - 0" 4 5/8" Automatic Sliding Door- See door type and frame SD
2-303A 2-303 2-303A F WD 1 3' - 0" 7' - 0" 1 3/4" 1 HM 17
2-307 2-307 2-303B 7' - 0" 7' - 0" 4 5/8" Automatic Sliding Door- See door and frame type SD
2-307A 2-303 2-303B 7' - 0" 7' - 0" 4 5/8" Automatic Sliding Door-See door and frame type SD
WALL TYPES
A B
N.T.S.Scale:
DOOR TYPES
N.T.S.Scale:
FRAME TYPES
N.T.S.Scale:
H.M. FRAME DETAILS
PLAN NOTES
2.04 Existing electrical panel to remain.
2.07 Remove existing bracket mounted casework.
2.08 Remove existing basecabinets.
2.09 Salvage existing hoods for reinstallation.
2.10 Remove existing gypsum ceiling.
2.11 Remove existing ACT ceiling.
2.12 Remove existing sheet flooring.
2.13 Remove existing LVT.
2.14 Remove existing drywall and metal framing from existing window openings.
6.01 2" x 4" wood blocking.
7.04 Install new drywall in existing window openings. Install sealant around window. Remove any existing selant prior to installing new.
7.13 Batt thermal wall insulation. R-11 min.
7.82 Continuous acoustic sealant @ non-rated walls / fire-rated sealant @ rated walls each side.
7.83 Cont. sealant each side.
8.02 Hollow metal frame.
8.09 Pass Through Window with interlock feature.
8.14 18" x 18" Access Doors
8.15 12" x 12" Access Doors
8.26 [4-1/2"] Aluminum storefront window system with 1/4" tempered glass.
8.63 Handwave motion sensor, both sides.
8.64 Handwave motion sensor.
9.02 3 5/8" non-structural metal framing @ 16" o.c.
9.03 6" non-structural metal framing @ 16" o.c.
9.08 Metal framing box header.
9.09 Double metal framing at jambs, typ.
9.13 Coved corners at ceiling wall intersection.
9.14 Solid Surface window trim.
9.21 5/8" Type X gypsum board.
9.30 Gypsum board bulkhead.
9.39 2" radius min. cove ceiling to wall transition moulding.
9.80 Acoustic insulation
27.01 Carder Reader. See technology drawings.
Scale: 1 1/2" = 1'-0"3
DETAIL
A. All wood blocking shall be fire retardant.
B. Light fixtures, sprinkler heads, speakers, and HVAC devices shall be located as close as possible to the center of acoustical ceiling panels.
Confirm any deviation from this with the COR.
C. See P, M, E and FP drawings for additional information regarding final information and quantites of fixtures/devices to be installed in ceiling.
D. See Electrical Drawings for lighting specifications.
E. All gypsum board ceilings to be painted P-4 unless noted otherwise.
F. Ceiling height to be 8'-0" AFF, unless noted otherwise.
GENERAL RCP NOTES
Scale: 1 1/2" = 1'-0"1
TYP. COVED CEILING DETAIL
Provide coved ceiling detail at ceiling permiters of rooms 2303B, 2303C, 2307. See floor plan for wall types.
A. Remove all existing partitions shown with dashed line. Repair wall and floor finishes to match adjacent surfaces at areas of removal or scheduled new surfaces as required.
B. Remove all existing borrowed lights, doors and hollow metal frames where indicated with dashed line. Unused hardware shall be turned over to Owner. Box hardware before giving to Owner.
C. See Plumbing, HVAC and Electrical Drawings for removal work associated with those trades.
D. Computer and communication cables and/or other wiring, piping, tubing, etc. serving other areas and functions located above ceiling or in walls of area to be remodeled shall be maintained during demolition/construction process. Coordinate any required changes to these systems with the C.O.R.
E. Repair existing walls to match adjacent undisturbed wall where cabinets, countertops, etc. are shown to be removed.
F. Prepare walls and floor slabs as required to receive new fininsh materials where existing finish is indicated to be removed.
G. Remove existing corner guards in areas to be remodeled.
GENERAL DEMO NOTES
Scale: 1 1/2" = 1'-0"2 DETAIL A. All wood blocking shall be fire retardant.
B. See Finish Plans for typical sill conditions at unlike floor finishes.
C. Seal joints between finish material and bottom of door frames.
D. Paint window frames at door windows to match door frame.
GENERAL DOOR NOTES
Note: All glazing in project to be tempered. Including storefront, sliding doors, and pass through windows.
1,300 sq. ft.
TC-1
AREA OF WORK
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
XX
JPA Project No.
Dwg. of
Pharmacy IV Prep Room Renovation
10-26-2018
ATTIC PLAN
A103
SOJR
18003
538-18-201
Chillicothe VA
N
Scale:
PROJECT
1/8" = 1'-0"
ATTIC WATERPROOFING PLAN
N.T.S.Scale:
Attic Photos N.T.S.Scale:
Attic Photos N.T.S.Scale:
Attic Photos N.T.S.Scale:
Attic Photos
N.T.S.Scale:
Attic Photos N.T.S.Scale:
Attic Photos
A. Existing MEP equipment to remain. Contractor to protect existing equipment during construction.
B. MEP systems to continue to operate during construction. Coordinate with C.O.R. any potential disruptions of service during construction.
C. Contractor to provide trims and curbs, compliant with traffic coating system, at all pipe penetrations, conduits penetrations, duct penetrations, floor mounted ducts and mechanical units, and ect. Contractor to remove and replace existing pipe, and duct insulation, as required to install traffic coating system trims and curbs.
D. Existing attic space has limited clearance and space.
Review attic space prior to bid, coordinate with C.O.R.
ATTIC NOTES
Ante
2-303B
Sterile Hazardous
Drug Compounding
2-303C
Hazardous Drug Storage
2-303A
Entry
2-303
Non Hazardous
Drug Compounding
2-307
Outpatient Pharmacy
2-300
LVT-1
LVT-1
RES-1
RES-1
RES-1
LVT-1
Ante
2-303B
Sterile Hazardous
Drug Compounding
2-303C
Hazardous Drug Storage
2-303A
Entry
2-303
Non Hazardous
Drug Compounding
2-307
Outpatient Pharmacy
2-300
P-2
P-3
EG-1
SP-1
SP-1SP-1
2'-9" 2'-9"
B2B1
PL-1
SS-1
RB-1
P-2
Drawer
B2B1
12.20
12.19
2'-1"
2'-1"
/2
/2
3"
12.19
12.20
9.60
12.21
12.31
12.22
12.29
12.30
3"
12.32
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
XX
JPA Project No.
Dwg. of
Pharmacy IV Prep Room Renovation
10-26-2018
FINISH PLANS
AF101
LPJR
18003
538-18-201
Chillicothe VA
PLAN NOTES
9.60 Resilient base. See Room Finish Schedule.
12.19 Applied solid surface backsplash.
12.20 Solid surface material over 3/4" plywood, with 1/4" eased edge.
12.21 Plastic laminate over 3/4" particle board.
12.22 Adjustable shelf. Plastic laminate over 3/4" particle board.
12.29 4" wire pull.
12.30 Self-latching, key-operated lock.
12.31 Concealed hinge.
12.32 Wire management grommet. Coordinate exact location with COR.
N
Scale:
PROJECT
1/8" = 1'-0"
FLOOR FINISH PLANN
Scale:
PROJECT
1/8" = 1'-0"
WALL FINISH PLAN
FINISH LEGEND
FINISH CATEGORY CODE FINISH TYPE MANUFACTURER PRODUCT / COLOR COMMENTS
CASEWORK FINISHES
CASEWORK
FINISHES
PL-1 PLASTIC LAMINATE NEVAMAR CARMEL SAGAWOOD WM8350T
CASEWORK
FINISHES
SS-1 SOLID SURFACE FORMICA 758 BIANCO MINERAL
CEILING FINISHES
CEILING FINISHES ACT-1 ACOUSTICAL CEILING TILE ARMSTRONG "ULTIMA HEALTH ZONE" 1935, SQUARE LAY-IN, 24" x 24", WHITE GRID: PRELUDE XL 15/16" EXPOSED TEE
CEILING FINISHES P-4 PAINT SHERWIN WILLIAMS SW 7006 EXTRA WHITE GLOSS LEVEL 1
CEILING FINISHES SP-2 HIGH BUILD SPECIALTY COATING STONHARD "STONGLAZE VSE" TO MATCH P-4
FLOOR FINISHES
FLOOR FINISHES LVT-1 LUXURY VINYL TILE SHAW CONTRACT "LEVEL HEXAGON" 0551V DEGREE 51103, 25" x 29" x 14" RANDOM INSTALLATION
FLOOR FINISHES RES-1 RESINOUS EPOXY FLOORING STONHARD "STONTEC ERF" 1/16" MOJAVE BEIGE INTEGRAL COVE BASE
FLOOR FINISHES TC-1 TRAFFIC COATING SHERWIN WILLIAMS EPO-FLEX MER II SYSTEM 3/32"
WALL FINISHES
WALL FINISHES EG-1 END WALL GUARD C-S ACROVYN FSC-25N 103 BEIGE TO CEILING
WALL FINISHES P-1 PAINT SHERWIN WILLIAMS SW 6142 MACADAMIA GLOSS LEVEL 3
WALL FINISHES P-2 PAINT SHERWIN WILLIAMS SW 6228 REFUGE GLOSS LEVEL 3
WALL FINISHES P-3 PAINT SHERWIN WILLIAMS MATCH EXISTING MATCH EXISTING
WALL FINISHES P-5 PAINT SHERWIN WILLIAMS SW 6143 BASKET BEIGE GLOSS LEVEL 4
WALL FINISHES RB-1 RUBBER BASE JOHNSONITE 6" COVE BASE, 09 CLAY WB
WALL FINISHES SP-1 HIGH BUILD SPECIALTY COATING STONHARD "STONGLAZE VSE" TO MATCH P-1
ROOM FINISH SCHEDULE
ROOM # ROOM NAME FLOOR BASE
WALLS CASEWORK
CEILINGNorth South East West Cabinet Countertop
2-300 Outpatient Pharmacy LVT-1/EXIST. - - P-3 P-3 - - - -
2-303 Entry LVT-1 RB-1 P-1 P-1,2 P-1 P-2 PL-1 SS-1 ACT-1
2-303A Hazardous Drug Storage LVT-1 RB-1 P-1 P-1 P-1 P-1 - - P-4
2-303B Ante RES-1 RES-1 SP-1 SP-1 SP-1 SP-1 - - SP-2
2-303C Sterile Hazardous Drug Compounding RES-1 RES-1 SP-1 SP-1 SP-1 SP-1 - - SP-2
2-307 Non Hazardous Drug Compounding RES-1 RES-1 SP-1 SP-1 SP-1 SP-1 - - SP-2
GENERAL FINISH PLAN NOTES
A. See Sheet AF101 for Finish Materials Legend.
B. See Sheet AF101 for Room Finish Schedule.
C. See Interior Elevations on Sheet AF101 for further information.
D. All door frames and hollow metal doors to be painted P-5, unless otherwise noted.
Scale: 1/4" = 1'-0"1
2-300B ENTRY
N.T.S.Scale:
CASEWORK SECTIONS
12 1314 15
60 636465
A
A
B
A
B
A
A
JOHN
A.
POE
IO
HO
FOET
A T
S
R E G
IS
E T
R
ED RA
C H
TI
E C T
Drawing Title
Approved: Project Director
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number o n e-ei g h th in ch o n e fo o t o n e-q u a rt er in ch o n e fo o t th re e-e ig h th s in c h o n e fo o t o n e-h a lf in c h o n e fo o t th re e-q u a rt er s in ch o n e fo o t o n e in c h o n e fo o t o n e a n d o n e-h a lf in ch e s o n e fo o t th re e in ch es o n e fo o t
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
FULLY SPRINKLERED
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
XX
JPA Project No.
Dwg. of
Pharmacy IV Prep Room Renovation
10-26-2018
SIGNAGE PLAN
IF102-S
SOJR
18003
538-18-201
Chillicothe VA
N.T.S.Scale:
SIGN TYPE A - Room ID Sign
N.T.S.Scale:
SIGN TYPE B - Equipment Sign
N
Scale:
PROJECT
1/4" = 1'-0"
SIGNAGE PLAN- SECOND FLOOR
SIGNAGE PLAN
Type Mark Type Comments Count
A 51" A.F.F. TO BOTTOM 5
B 51" A.F.F TO BOTTOM 2
A1
P1 P1OR
A1
OH
P2
B
P2
GENERAL FLOOR PLAN NOTES
RISER OR STACK NUMBER
DETAIL: B = DETAIL DESIGNATION
P2 = SHEET WHERE DETAIL IS LOCATED
FIRE SUPPRESSION HAZARD CLASSIFICATION AND HAZARD
CLASSIFICATION GROUP
SECTION: 1 = SECTION DESIGNATION
P2 = SHEET WHERE DETAIL IS LOCATED
EQUIPMENT REFERENCE. LETTER DESIGNATION VARIES. REFER TO
SCHEDULES.
EQUIPMENT, DEVICE, OR PLUMBING FIXTURE MARK. LETTER
DESIGNATIONS REFER TO SCHEDULES.
CONNECT TO EXISTING
PLAN NOTE. APPLIES ONLY TO THE SHEET WHICH IT IS SHOWN
UNLESS NOTED OTHERWISE.
DETAIL NOTE. APPLIES ONLY TO THE ASSOCIATED DETAIL.
"UP TO" SYMBOL (ITEM ON FLOOR ABOVE)
10" 10"
ELEV: 8' - 0"ELEV: 8' - 0"
TOE: 3' - 0"
BOE: 0' - 6"
APPROXIMATE DIMENSION ABOVE FINISHED FLOOR TO TOP OR
BOTTOM OF EQUIPMENT, UNLESS NOTED OTHERWISE
APPROXIMATE DIMENSION ABOVE FINISHED FLOOR TO CENTERLINE
OF PIPE, UNLESS NOTED OTHERWISE
FIRE SUPPRESSION SYMBOLS
DOUBLE LINE SINGLE LINE
CONCEALED PENDENT SPRINKLER
FS
PIV
S
FIRE HYDRANT
FLOW SWITCH
GATE VALVE OS&Y
INSTITUTIONAL PENDENT SPRINKLER
PENDENT SPRINKLER
POST INDICATOR VALVE
RECESSED PENDENT SPRINKLER
SIDE WALL SPRINKLER
SUPERVISED VALVE
UPRIGHT SPRINKLER
PIPING SYMBOLS
DOUBLE LINE SINGLE LINE
BOTTOM CONNECTION (45°)
BRANCH TEE CONNECTION (NOTE:
BULLHEAD TEE'S ARE NOT PERMITTED)
DIRECTION OF PITCH
DROP
ELBOW DOWN
ELBOW UP
EXISTING PIPE TO BE REMOVED
EXISTING PIPE TO REMAIN
FLOW DIRECTION DESIGNATION
PUMP
RISE
TOP CONNECTION (45°)
TOP CONNECTION (90°)
BOTTOM CONNECTION (90°)
DD
RR
SPECIALTY GAS SYMBOLS
AIR OUTLET - LAB TYPE
VC
VC
CARBON DIOXIDE OUTLET - HOSPITAL TYPE
MEDICAL AIR OUTLET - HOSPITAL TYPE
MEDICAL VACUUM INLET - HOSPITAL TYPE
NITROGEN OUTLET - HOSPITAL TYPE
NITROUS OXIDE OUTLET - HOSPITAL TYPE
OXYGEN OUTLET - HOSPITAL TYPE
WAGD INLET - HOSPITAL TYPE
NATURAL GAS OUTLET - LAB TYPE
VACUUM OUTLET - LAB TYPE
VACUUM CLEANING STATION - WALL
VACUUM CLEANING STATION - FLOOR
PLUMBING AND FIRE SUPPRESSION
PIPING DESIGNATIONS
DOMESTIC COLD WATER
DOMESTIC HOT WATER
DOMESTIC HOT WATER RETURN
EXISTING PIPE TO REMAIN
EXISTING PIPE TO BE REMOVED
HOT WATER PIPE (140 DEGS. F.)
HOT WATER PIPE (180 DEGS. F.)
HOT WATER RETURN PIPE (140 DEGS. F.)
HOT WATER RETURN PIPE (180 DEGS. F.)
ARGON PIPE
COMPRESSED AIR PIPE
MEDICAL CARBON DIOXIDE PIPE
COMBINATION FIRE SUPPRESSION AND DOMESTIC
WATER SERVICE
DIESEL FUEL PIPE
SPRINKLER PIPE (DRY)
FIRE SUPPRESSION (STANDPIPE / SPRINKLER MAIN)
FIRE SERVICE
FUEL OIL FLOW LINE
FUEL OIL GAUGE LINE
FUEL OIL RETURN LINE
FUEL OIL SUPPLY LINE
NATURAL GAS PIPE
GARAGE DRAINAGE PIPE
GAS SERVICE
HYDROGEN PIPE
HELIUM PIPE
MEDICAL INSTRUMENT AIR PIPE
KITCHEN WASTE PIPE
LABORATORY COMPRESSED AIR PIPE
LABORATORY VENT PIPE
LABORATORY VACUUM PIPE
LABORATORY WASTE PIPE
MEDICAL NITROGEN PIPE
MEDICAL COMPRESSED AIR PIPE
MEDICAL-SURGICAL VACUUM PIPE
GASOLINE PIPE (NON-LEAD)
MEDICAL NITROUS OXIDE PIPE
NON-POTABLE
MEDICAL OXYGEN PIPE
(OVERFLOW) SECONDARY STORM DRAINAGE PIPE
PROPANE GAS PIPE
PUMP DISCHARGE PIPE
PRE-ACTION / DELUGE SPRINKLER PIPE
PURE WATER PIPE
SPRINKLER PIPE (WET)
SANITARY DRAINAGE PIPE
SOFT COLD WATER
SPRINKLER DRAIN PIPE
STORM DRAINAGE PIPE
FUEL TANK VENT PIPE
TRAP PRIMER DISCHARGE PIPE
TEMPERED WATER PIPE
SANITARY SEWER VENT
WASTE ANESTHESIA GAS DISPOSAL PIPE
WATER SERVICE
LV
140R
180R
AR
CA
CO2
CWS
DF
DS
F
FS
FOF
FOG
FOR
FOS
G
GD
GS
H2
HE
IA
IV
IW
LA
LV
LW
N2
MA
MV
NG
NO
NPW
O2
OD
P
PD
PS
PW
S
SAN
SCW
SD
STM
T
TP
TW
V
WAGD
WS
INDIRECT VENT PIPE
INDIRECT WASTE PIPE
K
CHECK VALVE
T
P
FOB
FOT
SHUTOFF VALVE (REFER TO SPECIFICATIONS FOR
REQUIRED TYPE BASED ON APPLICATIONS)
CONCENTRIC PIPE REDUCER
ECCENTRIC PIPE REDUCER
PRESSURE GAUGE
TEMPERATURE GAUGE OR THERMOMETER
UNION
CLEANOUT
STRAINER
STRAINER WITH A BLOW DOWN VALVE AND HOSE
CONNECTION
DRAIN VALVE WITH HOSE END CONNECTION
AUTOMATIC FLOW CONTROLLER WITH P/T PLUG
IN AND OUT
EXPANSION JOINT
PRESSURE REGULATING VALVE
SAFETY RELIEF VALVE. PIPE DISCHARGE AIR
GAPPED TO FLOOR DRAIN UNLESS NOTED
OTHERWISE.
PRESSURE AND TEMPERATURE TEST PLUG
TRAP PRIMER
VACUUM GAUGE WITH STOP
CLEANOUT TO GRADE OR FINISHED FLOOR
END CAP
MIXING FAUCET
PLUG
HOSE BIB
WALL HYDRANT
PLUG VALVE
SHUTOFF VALVE AND BOX
SHUTOFF VALVE ON RISER
V
VB
TP
CO
COMBINATION SHUTOFF AND BALANCING
VALVE (REFER TO SPECIFICATIONS FOR
REQUIRED TYPE BASED ON APPLICATIONS)
BALL VALVE
BUTTERFLY VALVE
GATE VALVE
VALVES AND FITTINGS
DOUBLE LINE SINGLE LINE
SOLENOID VALVE
S
P/T PRESSURE AND TEMPERATURE SAFETY RELIEF
VALVE. PIPE DISCHARGE AIR GAPPED TO FLOOR
DRAIN UNLESS NOTED OTHERWISE.
FIRE DEPARTMENT VALVE
WATER METERM
ACID VENT PIPEAV
ACID WASTE PIPEAW
ABBREVIATIONS
AAP - AREA ALARM PANEL (MEDICAL GAS)
AC - AIR COMPRESSOR OR AIR CONDITIONER
ACC - ACCESS
ACCU - AIR COOLED CONDENSING UNIT
AD - ACCESS DOOR OR AREA DRAIN
ADB - ACID DILUTION BASIN
ADJ - ADJUSTABLE
AFF - ABOVE FINISHED FLOOR
AFG - ABOVE FINISHED GRADE
ALT - ALTERNATE
AP - ACCESS PANEL
APPROX - APPROXIMATE
AR - AIR RECEIVER OR ARGON
ARCH - ARCHITECT OR ARCHITECTURAL
ASSY - ASSEMBLY
AV - ACID VENT
AW - ACID WASTE
BDD - BACK DRAFT DAMPER
BFP - BACKFLOW PREVENTER
BLDG - BUILDING
BOB - BOTTOM OF BEAM
BOD - BOTTOM OF DUCT
BOE - BOTTOM OF EQUIPMENT
BOF - BOTTOM OF FOOTING
BOG - BOTTOM OF GRILLE
BOP - BOTTOM OF PIPE
BOT - BOTTOM
BT - BATHTUB
BTU - BRITISH THERMAL UNIT
BTUH - BRITISH THERMAL UNIT PER HOUR
BTWN - BETWEEN
CA - COMPRESSED AIR
CB - CATCH BASIN
CBD - COUNTER BALANCED BACKDRAFT DAMPER
CFCI - CONTRACTOR FURNISHED CONTRACTOR INSTALLED
CFM - CUBIC FEET PER MINUTE
CHS - CHILLED WATER SUPPLY
CHR - CHILLED WATER RETURN
CI - CAST IRON
CK - CLINICAL SINK
CLG - CEILING
CMU - CONCRETE MASONRY UNIT
CO - CLEAN OUT
CO2 - MEDICAL CARBON DIOXIDE
CONN - CONNECT OR CONNECTION
CONTR - CONTRACTOR
COR - CONTRACTING OFFICER'S REPRESENTATIVE
CORR - CORRIDOR
CS - CLINICAL SINK OR COLD SOFT WATER
CTR - CENTER
CU - COPPER
CW - COLD WATER
CWS - COMBINATION WATER SERVICE
OR CONDENSER WATER SUPPLY
CWR - CONDENSER WATER RETURN
D - DEPTH OR DRAIN LINE
DD - DECK DRAIN
DET - DETAIL
DF - DRINKING FOUNTAIN OR WATER COOLER
OR DIESEL FUEL
DFU - DRAINAGE FIXTURE UNIT
DI - DEIONIZED WATER
DIA - DIAMETER
DIM - DIMENSION
DN - DOWN
DS - DOWN SPOUT OR SPRINKLER (DRY)
DT - PERFORATED DRAIN TILE
DWG - DRAWING
EA - EACH
EC - ELECTRICAL CONTRACTOR (DIVISION 26)
EEW - EMERGENCY EYE WASH
EJ - EXPANSION JOINT
ELEC - ELECTRICAL
ELEV - ELEVATOR
EQ - EQUAL
EQUIP - EQUIPMENT
ET - EXPANSION TANK
ETR - EXISTING TO REMAIN
ES - EMERGENCY SHOWER
EQS - EQUIPMENT SUPPLIER
EWC - ELECTRICAL WATER COOLER
EXH - EXHAUST AIR
EXP - EXPANSION
EXT - EXTERIOR
EX - EXISTING
F - FIRE SUPPRESSION (STANDPIPE/SPRINKLER MAIN)
FCE - FIRE CONTROL EQUIPMENT
FCO - FLOOR CLEANOUT
FD - FLOOR DRAIN
FDC - FIRE DEPARTMENT CONNECTION
FDV - FIRE DEPARTMENT VALVE
FF - FINISHED FLOOR ELEVATION
FHC - FIRE HOSE CABINET
FLR - FLOOR
FM - FORCE MAIN
FOB - FLAT ON BOTTOM
FOF - FUEL OIL FLOW
FOG - FUEL OIL GAUGE
FOR - FUEL OIL RETURN
FOS - FUEL OIL SUPPLY
FOT - FLAT ON TOP
FPM - FEET PER MINUTE
FR - FIRE RISER
FS - FLOOR SINK OR FIRE SERVICE
FSC - FIRE SUPPRESSION CONTRACTOR (DIVISION 21)
FT - FEET
FTG - FOOTING
G - GAS OR NATURAL GAS
GA - GAUGE
GAL - GALLON
GALV - GALVANIZED
GC - GENERAL TRADES CONTRACTOR
GD - GARAGE DRAINAGE
GPM - GALLONS PER MINUTE
GS - GAS SERVICE
GW - GREASE WASTE
H2 - HYDROGEN
HB - HOSE BIBB
HC - HVAC CONTRACTOR (DIVISION 23)
HD - HUB DRAIN
HE - HELIUM
HG - REFRIGERANT HOT GAS
HP - HORSEPOWER OR HIGH POINT
HPC - HIGH PRESSURE CONDENSATE RETURN
HPS - HIGH PRESSURE STEAM SUPPLY
HPW - HIGH PURITY WATER
HPWR - HIGH PURITY WATER RETURN
HR - HOSE REEL
HT - HEAT TRACE
HTR - HEATER
HVAC - HEATING, VENTILATING, AND AIR CONDITIONING
HW - HOT WATER
HWR - HEATING HOT WATER RETURN
HWS - HEATING HOT WATER SUPPLY
IA - MEDICAL INSTRUMENT AIR
ID - INSIDE DIAMETER
INV - INVERT ELEVATION
IN - INCHES
IV - INDIRECT VENT
IW - INDIRECT WASTE
JS - JANITOR SINK
K - KITCHEN WASTE
KEC - KITCHEN EQUIPMENT CONTRACTOR
L - LENGTH
LA - LABORATORY COMPRESSED AIR
LAV - LAVATORY
LBS - POUNDS
LCW - LABORATORY COLD WATER
LEC - LABORATORY EQUIPMENT CONTRACTOR
LHW - LABORATORY HOT WATER
LHWR - LABORATORY HOT WATER RETURN
LPC - LOW PRESSURE CONDENSATE RETURN
LPS - LOW PRESSURE STEAM SUPPLY
LV - LABORATORY VACUUM OR LABORATORY VENT
LW - LABORATORY WASTE
MA - MEDICAL COMPRESSED AIR
MAP - MASTER ALARM PANEL (MEDICAL GAS)
MAX - MAXIMUM
MB - MOP BASIN
MC - MECHANICAL CONTRACTOR (DIVISION 23)
MEZZ - MEZZANINE
MFR - MANUFACTURER
MH - MANHOLE
MIN - MINIMUM OR MINUTE
MISC - MISCELLANEOUS
MTD - MOUNTED
MTG - MOUNTING
MPC - MEDIUM PRESSURE CONDENSATE RETURN
MPS - MEDIUM PRESSURE STEAM SUPPLY
MU - WATER MAKE-UP
MV - MEDICAL SURGICAL VACUUM
N2 - MEDICAL NITROGEN
N/C - NORMALLY CLOSED
NG - GASOLINE (NON-LEAD)
NIC - NOT IN CONTRACT
N/O - NORMALLY OPEN
NO - MEDICAL NITROUS OXIDE
NOM - NOMINAL
NPW - NON-POTABLE WATER
NPT - NATIONAL PIPE THREAD
NTS - NOT TO SCALE
02 - MEDICAL OXYGEN
OA - OUTDOOR AIR
OD - OUTSIDE DIAMETER OR OVERFLOW DRAIN
OFCI - OWNER FURNISHED CONTRACTOR
INSTALLED
OFOI - OWNER FURNISHED OWNER INSTALLED
P - PROPANE GAS
PC - PLUMBING CONTRACTOR (DIVISION 22)
OR PUMPED CONDENSATE RETURN
PD - PUMP DISCHARGE OR PARAPET DRAIN
PIV - POST INDICATOR VALVE
PLBG - PLUMBING
PS - PRE-ACTION/DELUGE SPRINKLER
PRESS - PRESSURE
PRV - PRESSURE REGULATING VALVE
PSF - POUNDS PER SQUARE FOOT
PSI - POUNDS PER SQUARE INCH
PSV - PRESSURE SUSTAINING VALVE
PSIG - POUNDS PER SQUARE INCH GAUGE
PW - PURE WATER
RA - RETURN AIR
RAD - RADIUS
RCP - REINFORCED CONCRETE PIPE
RD - ROOF DRAIN
REC - RECESSED
REQD - REQUIRED
RI - ROUGH IN
RL - REFRIGERANT LIQUID
ROS - REVERSE OSMOSIS WATER SUPPLY
ROR - REVERSE OSMOSIS WATER RETURN
RPM - REVOLUTIONS PER MINUTE
RS - REFRIGERANT SUCTION
RV - RELIEF VALVE
S - SPRINKLER (WET)
SA - SHOCK ARRESTOR OR SUPPLY AIR
SAN - SANITARY OR SANITARY DRAIN
SCH - SCHEDULE
SCW - SOFT COLD WATER
SD - SPRINKLER DRAIN OR SUBSOIL DRAIN
SH - SHOWER
SHT - SHEET
SK - SINK
SPEC - SPECIFICATIONS
SQ - SQUARE
SR - SUPPLY RISER
SS - SANITARY STACK (SOIL OR WASTE) OR
STAINLESS STEEL
STD - STANDARD
STM - STORM OR STORM DRAINAGE
STRUC - STRUCTURAL OR STRUCTURE
SUC - SITE UTILITY CONTRACTOR
T - FUEL TANK VENT
TD - TRENCH DRAIN
TEMP - TEMPERATURE
TOB - TOP OF BEAM
TOD - TOP OF DUCT
TOE - TOP OF EQUIPMENT
TOF - TOP OF FOOTING
TOJ - TOP OF JOIST
TOP - TOP OF PIPE
TOS - TOP OF SLAB OR TOP OF STEEL
TF - TRAP FILLER
TP - TRAP PRIMER OR TRAP PRIMER DISCHARGE
TW - TEMPERED WATER
TYP - TYPICAL
UR - URINAL
UNO - UNLESS NOTED OTHERWISE
V - VENT OR SANITARY SEWER VENT
VAC - VACUUM
VC - VACUUM CLEANING
VCV - VACUUM CLEANING VALVE
VE - VACUUM EXHAUST
VEL - VELOCITY
VIB - VALVE IN BOX
VOL - VOLUME
VP - VACUUM PUMP
VS - VENT STACK
VTR - VENT THROUGH ROOF
VR - VENT RISER
W/ - WITH
W/0 - WITHOUT
W - WASTE
WAGD - WASTE ANESTHESIA GAS DISPOSAL
WC - WATER CLOSET
WCO - WALL CLEANOUT
WH - WALL HYDRANT OR WATER HEATER
WIV - WALL INDICATOR VALVE
WS - WATER SERVICE
YCO - YARD CLEANOUT
ZVC - ZONE VALVE CABINET
PIPE RISER
NOTE: ALL SYMBOLS AND ABBREVIATIONS ARE SUBJECT
TO MODIFICATIONS ON OTHER DRAWINGS.
ALL SYMBOLS OR ABBREVIATIONS MIGHT NOT
NECESSARILY BE USED ON THIS PROJECT.
FIRE SUPPRESSION DESIGN NOTES
A PROVIDE A COMPLETE FIRE SUPPRESSION SPRINKLER SYSTEM IN THE RENOVATED AREA
AS INDICATED ON THE DRAWINGS.
B COORDINATE WORK WITH VAMC, PHASING PLAN AND OTHER TRADES. CONTRACTOR
SHALL BE RESPONSIBLE FOR OFFSETTING AND/OR RELOCATING EXISTING SERVICES TO
ACCOMMODATE NEW WORK.
C FOLLOW ALL VA STANDARD DESIGN PRACTICES, DETAILS, SPECIFICATIONS AND SITE
SPECIFIC COR REQUIREMENTS AS WELL AS WORK SHOWN ON THESE DOCUMENTS. ALL
DEVIATIONS FROM AFOREMENTIONED REQUIREMENTS SHALL BE APPROVED BY THE
COR.
D VERIFY THAT THERE IS ADEQUATE SPACE TO INSTALL ALL NEW PIPING REQUIRED TO
COMPLETE THIS WORK.
E ALL SPRINKLER PIPING IS LOCATED ABOVE THE CEILING (AT THE BOTTOM OF STRUCTURE
IN EXPOSED STRUCTURAL AREAS) UNLESS OTHERWISE INDICATED.
F THE BASIS OF DESIGN OF SPACING AND SIZING FOR THE COMPLETE FIRE SUPPRESSION
SPRINKLER SYSTEM SHALL BE FORE A WET PIPE SYSTEM HYDRAULICALLY CALCULATED
FOR A :
1 LIGHT HAZARD AREA LIMITED TO A MAXIMUM OF 225 SQ. FT. PER SPRINKLER WITH
THE DENSITY OF 0.10 GPM/SQ. FT. OVER THE MOST REMOTE 1500 SQ. FT. PROVIDE
AN ADDITION OF 100 GPM HOSE ALLOWANCE.
2 ALL MECHANICAL, ELECTRICAL AND SIMILAR SPACE AREAS SHALL BE ORDINARY
HAZARD (GROUP 1) LIMITED TO A MAXIMUM OF 130 SQ. FT. PER SPRINKLER WITH A
DENSITY OF 015 GPM/ SQ. FT. PROVIDE AN ADDITIONAL OF 250 GPM HOSE
ALLOWANCE.
3 ALL PHARMACY, STORAGE, TRASH, FILE STORAGE AND SIMILAR SPACE AREAS SHALL
BE ORDINARY HAZARD (GROUP 2) LIMITED TO A MAXIMUM OF 130 SQ. FT. PER
SPRINKLER WITH A DENSITY OF 0.20 GPM/ SQ. FT. PROVIDE AN ADDITION OF 250 GPM
HOSE ALLOWANCE.
G PROVIDE ALL SPRINKLER LOCATIONS AND RELATED PIPING.
H THE ENTIRE SPRINKLER SYSTEM SHALL CONFORM TO VAMC, LOCAL, STATE AND NFPA
STANDARD, RULES AND REGULATIONS.
I CAREFULLY COORDINATE THE LOCATION OF ALL SPRINKLERS AND PIPING WITH ALL
OTHER TRADES AND THE ARCHITECT.
J VERIFY THE LOCATION OF ALL LIGHTING FIXTURES AND AIR DEVICES BEFORE
INSTALLATION OF SPRINKLERS.
K PROVIDE SPRINKLER PIPING LAYOUT AND SIZING.
L ALL SPRINKLER LOCATIONS MUST BE COORDINATED WITH ALL EQUIPMENT KINDRED TO
OTHER TRADES. SPRINKLER LOCATIONS MUST BE APPROVED BY THE ARCHITECT
BEFORE WORK COMMENCES. REFER TO THE ARCHITECTURAL AND REFLECTIVE CEILING
PLANS FOR ADDITIONAL INFORMATION ON ALL CEILING TYPES, HEIGHTS AND OTHER
ARCHITECTURAL FEATURES.
M MAKE NECESSARY FLOW TESTS, LAYOUT SYSTEMS, OBTAIN APPROVALS FROM THE
AUTHORITIES HAVING JURISDICTION PRIOR TO BEGINNING ANY FABRICATION OR
INSTALLATION WORK.
N SPRINKLER PIPING SHALL BE 1" MINIMUM SIZE.
O SPRINKLERS (UNLESS OTHER NOTED):
EXPOSED AREAS - UPRIGHT OR PENDENT TYPE - STANDARD BRASS
CEILING AREA - RECESSED PENDENT TYPE
ALL SPRINKLERS SHALL BE UL LISTED AND FM APPROVED.
P SPRINKLERS SHALL BE "QUICK RESPONSE" TYPE IN ALL AREAS EXCEPT HIGH
TEMPERATURE AREAS.
Q ALL PENDENT TYPE SPRINKLERS LOCATED IN SUSPENDED CEILING TILE AREAS SHALL BE
LOCATED IN THE CENTER OF THE CEILING TILE.
R ANNULAR SPACE AT PIPE AND OTHER SIMILAR PENETRATIONS OF FIRE RATED
ASSEMBLIES SHALL BE FIRESTOPPED. REFER TO SPECIFICATION SECTION 07 84 00
FIRESTOPPING. REFER TO ARCHITECTURAL DRAWINGS FOR CONSTRUCTION, LAYOUT
AND FIRE RATING OF FLOORS, WALLS AND PARTITIONS AND OTHER BUILDING ELEMENTS.
S A FLOW TEST IS REQUIRED. REFER TO SPECIFICATION SECTION 21 13 13 FOR FLOW
TEST INFORMATION.
T ALL EXISTING SMOKE COMPARTMENTS SHALL BE MAINTAINED. ALL NEW SPRINKLER
ZONES SHALL MATCH SMOKE COMPARTMENTS.
U REFER TO ARCHITECTURAL DRAWINGS FOR INFORMATION ON SMOKE COMPARTMENTS,
SMOKE PARTITIONS AND FIRE BARRIERS.
V PHASING: REFER TO ARCHITECTURAL DRAWINGS FOR AN OUTLINE OF PROJECT
PHASING. ADDITIONAL NOTES DESCRIBING PHASING DETAILS MAY BE INCLUDED
THROUGHOUT THIS DRAWING SET.
W ANY NEW OR RENOVATION WORK REQUIRED OUTSIDE OF THE BOUNDARIES/ EXTENTS
OF SPECIFIC PHASE (OR TEMPORARY CONSTRUCTION WALLS PER DIVISION 1) SHALL BE
BID AS "NIGHT WORK" AND SHALL BE AS DIRECTED BY THE COR.
X ALL CORE DRILL AND PIPE EXTENSIONS THROUGH EXISTING STRUCTURAL FLOOR SLAB
OR STRUCTURAL WALLS SHALL BE X-RAYED TO DETERMINE THE EXACT LOCATION OF
"RE-BARS" IN STRUCTURAL FLOOR/ WALL. THE CONTRACTOR SHALL NOT CUT THE
LOWER (BOTTOM) "RE-BARS" IN THE STRUCTURAL FLOOR SLAB.
Y REQUIRED SHUT-DOWNS OF THE SMOKE DETECTION SYSTEM AND FIRE SUPPRESSION
SPRINKLER SYSTEM SHALL NOT OCCUR CONCURRENTLY AND SHALL COMPLY WITH THE
JOINT COMMISSION ON ACCREDITATION OF HEALTHCARE ORGANIZATIONS (JCAHO)
INTERIM LIFE SAFETY MEASURES (ILSM). COORDINATION BETWEEN THE COR AND ALL
TRADES SHALL BE REQUIRED BEFORE SHUT-DOWN OF THE SMOKE DETECTION SYSTEM
OR FIRE SUPPRESSION SPRINKLER SYSTEM.
Z AUTOMATIC SPRINKLER SYSTEM SHALL BE MAINTAINED IN FULLY OPERATIONAL
CONDITION. SPRINKLER SYSTEM MAY BE TAKEN OFF-LINE FOR UP TO 8 HOURS, AS
REQUIRED TO MAKE MODIFICATIONS/CONNECTIONS TO EXISTING SYSTEM. SYSTEM SHALL
BE BROUGHT BACK ON-LINE AT THE END OF EVERY WORK SHIFT. COORDINATE ALL
DOWNTIME AND FIREWATCH WITH COR MINIMUM 2 WEEKS IN ADVANCE.
MATTHEW S.
OH
A
T
S T
E
OF
I O
GRAHAM
E-72670
IEF
O
R P
A
SS
NOI GNEL
SIGER DERET
E
N
R E
Drawing Title
Reviewed for Concept Only
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number
Dwg. of on e-ei gh th in ch o ne fo ot on e-qu ar te r i nc h on e fo ot th re e-ei gh th s in ch o ne fo ot on e-ha lf in ch o ne fo ot th re e-qu ar te rs in ch o ne fo ot on e in ch o ne fo ot on e an d on e-ha lf in ch es o ne fo ot th re e in ch es o ne fo ot
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
JPA Project No.
FULLY SPRINKLERED
Revision Date
HEAPY PROJECT No.: 2018-05066 OHIO COA FIRM No.: 01528
10.23.18
/1
/2
:4
:5
P
M
Pharmacy IV Prep Room Renovation
10/23/2018
LEGEND, ABBREVIATIONS, INDEX
OF DRAWINGS
FX001
KMBEWL
18003.00
538-18-201
Chillicothe VA
INDEX OF DRAWINGS - FIRE SUPPRESSION
Sheet Number Sheet Name
FX001 LEGEND, ABBREVIATIONS, INDEX OF DRAWINGS
FX102 PARTIAL SECOND FLOOR PLAN
BASIS OF DESIGN FLOW TEST DATA
LOCATION: HYDRANT #3 NEAR BUILDING 31
STATIC PRESSURE: 490 kPa (71 PSI) RESIDUAL PRESSURE: 427 kPa (62 PSI) FLOW: 78 L/s (1,233 GPM)
DATE: 07/07/2015 TIME: 2:30 PM
NOTE: PERFORM A FLOW TEST OF HYDRANTS #2 & #3 TO VERIFY AVAILABLE
PRESSURE AND FLOW.
12 14 15 13
2-307
Non Hazardous
Drug Compounding
2-303B Ante
2-303C
Sterile Hazardous
Drug Compounding
2-303A
Hazardous Drug
Storage
OH
OH
OH
OH
OH
Entry 2-303
MATTHEW S.
OH
A
T
S T
E
OF
I O
GRAHAM
E-72670
IEF
O
R P
A
SS
NOI GNEL
SIGER DERET
E
N
R E
Drawing Title
Reviewed for Concept Only
Location
Building Number
Project Title
Checked DrawnDate
Project No.
Drawing Number
Dwg. of on e-ei gh th in ch o ne fo ot on e-qu ar te r i nc h on e fo ot th re e-ei gh th s in ch o ne fo ot on e-ha lf in ch o ne fo ot th re e-qu ar te rs in ch o ne fo ot on e in ch o ne fo ot on e an d on e-ha lf in ch es o ne fo ot th re e in ch es o ne fo ot
F E
D C
B A
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
F E
D C
B A
VA Project No.
ARCHITECT/ENGINEERS:CONSULTANTS:
Office of Construction and Facilities Management
JPA Project No.
FULLY SPRINKLERED
Revision Date
HEAPY PROJECT No.: 2018-05066 OHIO COA FIRM No.: 01528
10.23.18
/1
/2
:4
:5
P
M
Pharmacy IV Prep Room Renovation
10/23/2018
PARTIAL SECOND FLOOR PLAN
FX102
KMBEWL
18003.00
538-18-201
Chillicothe VA
N
Scale:
PROJECT
1/4" = 1'-0"
PARTIAL SECOND FLOOR PLAN
GENERAL NOTES
A REFER TO SHEET FX001 FOR LEGEND, ABBREVIATIONS & INDEX OF
DRAWINGS.
PLAN NOTES
1 EXISTING TO REMAIN.
2 LIMITS OF RENOVATED AREA REQUIRING EXISTING SPRINKLER SYSTEM
MODIFICATIONS. MODIFICATIONS INCLUDE, BUT ARE NOT LIMITED TO,
REPLACING ALL SPRINKLERS AND BRANCH PIPING TO CONFORM TO NEW
ARCHITECTURAL LAYOUT WHERE NECESSARY. EXISTING PIPING SHALL BE
RE-USED AS MUCH AS POSSIBLE.
A1
P1 P1OR
A1
OH
P2
B
P2
GENERAL FLOOR PLAN NOTES
RISER OR STACK NUMBER
DETAIL: B = DETAIL DESIGNATION
P2 = SHEET WHERE DETAIL IS LOCATED
FIRE SUPPRESSION HAZARD CLASSIFICATION AND HAZARD
CLASSIFICATION GROUP
SECTION: 1 = SECTION DESIGNATION
P2 = SHEET WHERE DETAIL IS LOCATED
EQUIPMENT REFERENCE. LETTER DESIGNATION VARIES. REFER TO
SCHEDULES.
EQUIPMENT, DEVICE, OR PLUMBING FIXTURE MARK. LETTER
DESIGNATIONS REFER TO SCHEDULES.
CONNECT TO EXISTING
PLAN NOTE. APPLIES ONLY TO THE SHEET WHICH IT IS SHOWN
UNLESS NOTED OTHERWISE.
DETAIL NOTE. APPLIES ONLY TO THE ASSOCIATED DETAIL.
"UP TO" SYMBOL (ITEM ON FLOOR ABOVE)
10" 10"
ELEV: 8' - 0"ELEV: 8' - 0"
TOE: 3' - 0"
BOE: 0' - 6"
APPROXIMATE DIMENSION ABOVE FINISHED FLOOR TO TOP OR
BOTTOM OF EQUIPMENT, UNLESS NOTED OTHERWISE
APPROXIMATE DIMENSION ABOVE FINISHED FLOOR TO CENTERLINE
OF PIPE, UNLESS NOTED OTHERWISE
FIRE SUPPRESSION SYMBOLS
DOUBLE LINE SINGLE LINE
CONCEALED PENDENT SPRINKLER
FS
PIV
S
FIRE HYDRANT
FLOW SWITCH
GATE VALVE OS&Y
INSTITUTIONAL PENDENT SPRINKLER
PENDENT SPRINKLER
POST INDICATOR VALVE
RECESSED PENDENT SPRINKLER
SIDE WALL SPRINKLER
SUPERVISED VALVE
UPRIGHT SPRINKLER
PIPING SYMBOLS
DOUBLE LINE SINGLE LINE
BOTTOM CONNECTION (45°)
BRANCH TEE CONNECTION (NOTE:
BULLHEAD TEE'S ARE NOT PERMITTED)
DIRECTION OF PITCH
DROP
ELBOW DOWN
ELBOW UP
EXISTING PIPE TO BE REMOVED
EXISTING PIPE TO REMAIN
FLOW DIRECTION DESIGNATION
PUMP
RISE
TOP CONNECTION (45°)
TOP CONNECTION (90°)
BOTTOM CONNECTION (90°)
DD
RR
SPECIALTY GAS SYMBOLS
AIR OUTLET - LAB TYPE
VC
VC
CARBON DIOXIDE OUTLET - HOSPITAL TYPE
MEDICAL AIR OUTLET - HOSPITAL TYPE
MEDICAL VACUUM INLET - HOSPITAL TYPE
NITROGEN OUTLET - HOSPITAL TYPE
NITROUS OXIDE OUTLET - HOSPITAL TYPE
OXYGEN OUTLET - HOSPITAL TYPE
WAGD INLET - HOSPITAL TYPE
NATURAL GAS OUTLET - LAB TYPE
VACUUM OUTLET - LAB TYPE
VACUUM CLEANING STATION - WALL
VACUUM CLEANING STATION - FLOOR
PLUMBING AND FIRE SUPPRESSION
PIPING DESIGNATIONS
DOMESTIC COLD WATER
DOMESTIC HOT WATER
DOMESTIC HOT WATER RETURN
EXISTING PIPE TO REMAIN
EXISTING PIPE TO BE REMOVED
HOT WATER PIPE (140 DEGS. F.)
HOT WATER PIPE (180 DEGS. F.)
HOT WATER RETURN PIPE (140 DEGS. F.)
HOT WATER RETURN PIPE (180 DEGS. F.)
ARGON PIPE
COMPRESSED AIR PIPE
MEDICAL CARBON DIOXIDE PIPE
COMBINATION FIRE SUPPRESSION AND DOMESTIC
WATER SERVICE
DIESEL FUEL PIPE
SPRINKLER PIPE (DRY)
FIRE SUPPRESSION (STANDPIPE / SPRINKLER MAIN)
FIRE SERVICE
FUEL OIL FLOW LINE
FUEL OIL GAUGE LINE
FUEL OIL RETURN LINE
FUEL OIL SUPPLY LINE
NATURAL GAS PIPE
GARAGE DRAINAGE PIPE
GAS SERVICE
HYDROGEN PIPE
HELIUM PIPE
MEDICAL INSTRUMENT AIR PIPE
KITCHEN WASTE PIPE
LABORATORY COMPRESSED AIR PIPE
LABORATORY VENT PIPE
LABORATORY VACUUM PIPE
LABORATORY WASTE PIPE
MEDICAL NITROGEN PIPE
MEDICAL COMPRESSED AIR PIPE
MEDICAL-SURGICAL VACUUM PIPE
GASOLINE PIPE (NON-LEAD)
MEDICAL NITROUS OXIDE PIPE
NON-POTABLE
MEDICAL OXYGEN PIPE
(OVERFLOW) SECONDARY STORM DRAINAGE PIPE
PROPANE GAS PIPE
PUMP DISCHARGE PIPE
PRE-ACTION / DELUGE SPRINKLER PIPE
PURE WATER PIPE
SPRINKLER PIPE (WET)
SANITARY DRAINAGE PIPE
SOFT COLD WATER
SPRINKLER DRAIN PIPE
STORM DRAINAGE PIPE
FUEL TANK VENT PIPE
TRAP PRIMER DISCHARGE PIPE
TEMPERED WATER PIPE
SANITARY SEWER VENT
WASTE ANESTHESIA GAS DISPOSAL PIPE
WATER SERVICE
LV
140R
180R
AR
CA
CO2
CWS
DF
DS
F
FS
FOF
FOG
FOR
FOS
G
GD
GS
H2
HE
IA
IV
IW
LA
LV
LW
N2
MA
MV
NG
NO
NPW
O2
OD
P
PD
PS
PW
S
SAN
SCW
SD
STM
T
TP
TW
V
WAGD
WS
INDIRECT VENT PIPE
INDIRECT WASTE PIPE
K
CHECK VALVE
T
P
FOB
FOT
SHUTOFF VALVE (REFER TO SPECIFICATIONS FOR
REQUIRED TYPE BASED ON APPLICATIONS)
CONCENTRIC PIPE REDUCER
ECCENTRIC PIPE REDUCER
PRESSURE GAUGE
TEMPERATURE GAUGE OR THERMOMETER
UNION
CLEANOUT
STRAINER
STRAINER WITH A BLOW DOWN VALVE AND HOSE
CONNECTION
DRAIN VALVE WITH HOSE END CONNECTION
AUTOMATIC FLOW CONTROLLER WITH P/T PLUG
IN AND OUT
EXPANSION JOINT
PRESSURE REGULATING VALVE
SAFETY RELIEF VALVE. PIPE DISCHARGE AIR
GAPPED TO FLOOR DRAIN UNLESS NOTED
OTHERWISE.
PRESSURE AND TEMPERATURE TEST PLUG
TRAP PRIMER
VACUUM GAUGE WITH STOP
CLEANOUT TO GRADE OR FINISHED FLOOR
END CAP
MIXING FAUCET
PLUG
HOSE BIB
WALL HYDRANT
PLUG VALVE
SHUTOFF VALVE AND BOX
SHUTOFF VALVE ON RISER
V
VB
TP
CO
COMBINATION SHUTOFF AND BALANCING
VALVE (REFER TO SPECIFICATIONS FOR
REQUIRED TYPE BASED ON APPLICATIONS)
BALL VALVE
BUTTERFLY VALVE
GATE VALVE
VALVES AND FITTINGS
DOUBLE LINE SINGLE LINE
SOLENOID VALVE
S
P/T PRESSURE AND TEMPERATURE SAFETY RELIEF
VALVE. PIPE…