BioTek Synergy™ Hybrid Multi-mode Microplate Reader or Equal Salient Characteristics Project: VAPAHCS Research VMU Features, Design and Construction Common to Hybrid Multi-mode Microplate Readers:
· The hybrid multi-mode microplate reader shall include the following detection modes: top and bottom fluorescence intensity, UV-visible absorbance, luminescence, fluorescence polarization, TRF & TR-FRET.
· The hybrid multi-mode microplate reader shall include temperature control to 45 ̊C and Shaking.
· The hybrid multi-mode microplate reader shall include one filter cube with filters and data analysis software.
· The data analysis software shall offer advanced reader control functionality, powerful data analysis and flexible exporting/reporting tools combined in an easy-to-use software package.
· The data analysis software shall include software upgrades to meet windows operation system updates through the life of equipment.
· The hybrid multi-mode microplate reader shall be a GSA labeled product.
· The hybrid module microplate reader shall include the dual reagent injector module accessory.
· The dual reagent injector module shall include two (2) syringes, two (2) reagent bottles, holders and inlet tubing, and feedthrough assembly with straight dispense tips.
Warranty Common to Hybrid Multi-mode Microplate Readers:
· The vendor shall provide a warranty of twelve months of parts and labor.
Certifications, Environmental Impact and Testing Common to Hybrid Multi-mode Microplate Readers:
· CE and TUV marked. RoHS compliant.
Critical Characteristics Unique to Hybrid Multi-mode Microplate Readers:
· Absorbance:
· Light Source: Xenon Flash
· Detector: photodiode
· Wavelength selection: monochromator
· Wavelength range: 230-999 nm, 1 nm increments
· Monochromator bandwidth: 4 nm, 8 nm
· Dynamic range: 0 – 4.0 OD
· Resolution: .0001 OD
· Pathlength correction: yes
· Monochromator wavelength accuracy: ±2 nm
· Monochromator wavelength repeatability: ±0.2 nm
· OD accuracy: <1% at 2.0 OD; <3% at 3.0 OD
· OD linearity: <1% from 0 to 3.0 OD
· OD repeatability: <.5% at 2.0 OD
· Stray light: 0.03% at 230 nm
· Reading speed (kinetic): 96 wells: 11 seconds; 384 wells: 22 seconds
· Fluorescence Intensity:
· Light Source: Xenon Flash
· Detector: PMT for monochromator system; PMT for filter system
· Wavelength selection: Quad monochromators (top/bottom); Filters (top)
· Wavelength range: Monochromators: 250-700 nm; Filters: 200-700 nm
· Monochromator bandwidth: Fixed, 16 nm
· Dynamic range: 7 decades
· Sensitivity:
· Filters: Fluorescein 0.25 pM (0.025 fmol/well, 384-well plate);
· Quad Monochromator:Fluorescein 2.5 pM (0.25 fmol/well, 384-well plate) – top;
· Fluorescein 4 pM (0.4 fmol/well, 384-well plate) – bottom
· Reading speed (kinetric): 96 wells: 11 seconds; 384 wells: 22 seconds
· Luminescence:
· Wavelength range: 300-700 nm
· Dynamic range: >6 decades
· Sensitivity:
· Monos: 20 amol ATP (flash)
· Filters 10 amol ATP (flash), 100 amol (glow)
· Fluorescence Polarization:
· Light source: Xenon flash
· Detector: PMT
· Wavelength selection: Filters
· Wavelength range: 280-700 nm
· Sensitivity: 1.2 mP standard deviation at 1 nm fluorescein
· Time-Resolved Fluorescence:
· Light source: Xenon Flash
· Detector: PMT
· Wavelength selection: Quad monochromators (secondary mode); Filters (top)
· Wavelength range: Filters: 200-700 nm
· Sensitivity:
· Filters: Europium 40 fM (4 amol/well, 384-well plate);
· Monos: Europium 1200 fM (120 amol/well, 384-well plate)
· Reagent Dispensers:
· Supported detection modes: All modes
· Number: 2 syringe pumps
· Supported labware: 6- to 384-well microplates, Petri dishes
· Dead volume: 1.1 mL with back flush
· Dispense volume: 5 – 1000 µL in 1 µL increment
· Dispense accuracy: ± µL or 2%
· Dispense precision: <2% at 50-200 µL