Salient Characteristics

Palo Alto Pkg. 0022 - Applied Biosystems QuantStudio 7 Flex Real-Time PCR (Brand Name or Equal)

QTY: 1

General The QuantStudio 7 Flex Real-Time PCR is a system capable of qPCR – typical PCR (Polymerase Chain Reaction) is a technique used in labs to make millions of copies (amplify) of a particular section of DNA from a very small amount of DNA. Real-Time PCR, also referred to as Quantitative PCR (or qPCR) is a procedure based upon the traditional PCR techniques, but it can also detect nucleic acid in a precise and rapid manner, plus it is able to do analysis with the help of various protocols, reagents, and software (options) for a broad range of applications.

Warranty Requirements Manufacturer must offer a (1) one-year minimum warranty that product is free from defects in materials and workmanship. All defects will be replaced or repaired by the manufacturer at no cost to the buyer.

Key Applications

a. Gene Expression

b. MicroRNA Profiling

c. SNP Genotyping

d. Copy Number Variation

e. Protein Thermal Shift

f. High Resolution Melt

g. Pathogen Detection

h. Pharmacogenomics

Specifications

a. Block Configuration: 96-well 96-well Fast 384-well TaqMan Array

b. Dimensions:73.8 cm (h) x 50.1 cm (w) x 66.1 cm (d)
c. Block Change Design:Block changes from front in less than 1 minute; no tools required
d. Run-time:30 min (96-well Fast)

35 min (384-well)

e. Ramp-up Rates: 3.9°C/sec (96-well) 6.5°C/sec (96-well Fast) 3.5°C/sec (384-well) 4.3°C/sec (TAC

f. Ramp-down Rates: 3.6°C/sec (96-well) 6.0°C/sec (96-well Fast) 3.0°C/sec (384-well) 3.7°C/sec (TAC)

g.Well-to-well Variability:+/- 0.25°C
h.Excitation Source:OptiFlex System lamp
i.Detection Channels:Excitation 455-672 nm
Emission 505-723 nm