23Q0216_SOL.pdf

PDF 1 MB Posted

Attached to
BATTERY TESTER Federal contract opportunity
Solicitation number
N0016423Q0216
Issued by
Department of the Navy Naval Sea Systems Command

About this file

This document is a solicitation for the purchase of battery test equipment and related accessories. The Navy seeks to procure one 40V battery tester with 16 channels, one 10V battery tester with 64 channels, two 5V battery test stations, two 60V battery test systems, one 9-400V battery tester, and options for additional 5V and 60V test systems. Delivery for most items is required within 28 weeks of award. The solicitation includes detailed specifications for the equipment and describes supplies, inspection and acceptance terms, and delivery information. It also lists contract clauses typically included in federal contracts.

View the file

Other files for this federal contract opportunity

Other files attached to BATTERY TESTER, newest first.
File Type Posted
23Q0216_SOL_0002.pdf PDF
23Q0216_SOL_0001.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SEE ADDENDUM

(No Collect Calls)

N0016423Q0216 20-Jul-2023

b. TELEPHONE NUMBER

812-854-0000

8. OFFER DUE DATE/LOCAL TIME

12:00 PM 03 Aug 2023

5. SOLICITATION NUMBER 6. SOLICITATION ISSUE DATE

AUTHORIZED FOR LOCAL REPRODUCTION

PREVIOUS EDITION IS NOT USABLE

STANDARD FORM 1449 (REV. 2/2012)

Prescribed by GSA – FAR (48 CFR) 53.212

(TYPE OR PRINT)

(SIGNATURE OF CONTRACTING OFFICER)

ADDENDA ARE

26. TOTAL AWARD AMOUNT (For Gov t. Use Only )

23.

CODE 10. THIS ACQUISITION IS

SUCH ADDRESS IN OFFER

17b. CHECK IF REMITTANCE IS DIFFERENT AND PUT

BELOW IS CHECKED

TELEPHONE NO.

N001649. ISSUED BY

18b. SUBMIT INVOICES TO ADDRESS SHOWN IN BLOCK 18a. UNLESS BLOCK

7. FOR SOLICITATION

INFORMATION CALL:

a. NAME

BRAXTON W TAYLOR

2. CONTRACT NO. 3. AWARD/EFFECTIVE DATE 4. ORDER NUMBER

(TYPE OR PRINT)

30b. NAME AND TITLE OF SIGNER 30c. DATE SIGNED 31b. NAME OF CONTRACTING OFFICER

30a. SIGNATURE OF OFFEROR/CONTRACTOR 31a.UNITED STATES OF AMERICA

0 27a. SOLICITATION INCORPORATES BY REFERENCE FAR 52.212-1. 52.212-4. FAR 52.212-3. 52.212-5 ARE ATTACHED.

25. ACCOUNTING AND APPROPRIATION DATA

1. REQUISITION NUMBER

20.

ADDITIONAL SHEETS SUBJECT TO THE TERMS AND CONDITIONS SPECIFIED.

OFFEROR TO COMPLETE BLOCKS 12, 17, 23, 24, AND 30

SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL ITEMS

ARE NOT ATTACHED

27b. CONTRACT/PURCHASE ORDER INCORPORATES BY REFERENCE FAR 52.212-4. FAR 52.212-5 IS ATTACHED. ADDENDA ARE ARE NOT ATTACHED

(BLOCK 5), INCLUDING ANY ADDITIONS OR CHANGES WHICH ARE

SET FORTH HEREIN, IS ACCEPTED AS TO ITEMS:

. YOUR OFFER ON SOLICITATION

28. CONTRACTOR IS REQUIRED TO SIGN THIS DOCUMENT AND RETURN

% FOR:SET ASIDE:UNRESTRICTED OR X

SMALL BUSINESSX

17a.CONTRACTOR/ CODE FACILITY

OFFEROR CODE

NAVAL SURFACE WARFARE CENTER CRANE DIV

B. TAYLOR (812-854-1024)

BLDG 3373

300 HWY 361

CRANE IN 47522-5001

18a. PAYMENT WILL BE MADE BY CODE

RATED ORDER UNDER

DPAS (15 CFR 700)

13a. THIS CONTRACT IS AX

13b. RATING

DO-A7

CODE15. DELIVER TO CODE 16. ADMINISTERED BY

SEE SCHEDULE

12. DISCOUNT TERMS11. DELIVERY FOR FOB DESTINA-

TION UNLESS BLOCK IS

MARKED

SEE SCHEDULE

14. METHOD OF SOLICITATION

RFQ IFB RFPX

FAX:

TEL: 812-854-1024 SERVICE-DISABLED

VETERAN-OWNED

SMALL BUSINESS

8(A)

HUBZONE SMALL

BUSINESS

SIZE STANDARD:

NAICS:

334515

X

OFFER DATED

29. AWARD OF CONTRACT: REF.

DELIVER ALL ITEMS SET FORTH OR OTHERWISE IDENTIFIED ABOVE AND ON ANY

COPIES TO ISSUING OFFICE. CONTRACTOR AGREES TO FURNISH AND

EMAIL:

TEL:

31c. DATE SIGNED

SEE SCHEDULE

SCHEDULE OF SUPPLIES/ SERVICESITEM NO. QUANTITY UNIT UNIT PRICE AMOUNT

24.22.21.19.

WOMEN-OWNED SMALL BUSINESS (WOSB)

ELIGIBLE UNDER THE WOMEN-OWNED

SMALL BUSINESS PROGRAM

EDWOSB

32g. E-MAIL OF AUTHORIZED GOVERNMENT REPRESENTATIVE

SOLICITATION/CONTRACT/ORDER FOR COMMERCIAL ITEMS

(CONTINUED)

PAGE 2 OF91

ACCEPTED, AND CONFORMS TO THE CONTRACT, EXCEPT AS NOTED: ______________________________________________________

32a. QUANTITY IN COLUMN 21 HAS BEEN

RECEIVED INSPECTED

32b. SIGNATURE OF AUTHORIZED GOVERNMENT

REPRESENTATIVE

32c. DATE 32d. PRINTED NAME AND TITLE OF AUTHORIZED GOVERNMENT

REPRESENTATIVE

32e. MAILING ADDRESS OF AUTHORIZED GOVERNMENT REPRESENTATIVE 32f . TELEPHONE NUMBER OF AUTHORIZED GOVERNMENT REPRESENTATIVE

37. CHECK NUMBER

FINALPARTIALCOMPLETE

36. PAYMENT35. AMOUNT VERIFIED

CORRECT FOR

34. VOUCHER NUMBER

FINAL

33. SHIP NUMBER

PARTIAL

38. S/R ACCOUNT NUMBER 39. S/R VOUCHER NUMBER 40. PAID BY

41a. I CERTIFY THIS ACCOUNT IS CORRECT AND PROPER FOR PAYMENT

41b. SIGNATURE AND TITLE OF CERTIFYING OFFICER 41c. DATE

42a. RECEIVED BY (Print)

42b. RECEIVED AT (Location)

42c. DATE REC'D (YY/MM/DD) 42d. TOTAL CONTAINERS

STANDARD FORM 1449 (REV. 2/2012) BACK

Prescribed by GSA – FAR (48 CFR) 53.212

AUTHORIZED FOR LOCAL REPRODUCTION

PREVIOUS EDITION IS NOT USABLE

SEE SCHEDULE

20.

SCHEDULE OF SUPPLIES/ SERVICES

21.

QUANTITY UNIT

22. 23.

UNIT PRICE

24.

AMOUNT

19.

ITEM NO.

Section SAP - SAP Continuation Sheet

ITEM NO SUPPLIES/SERVICES QTY U/I UNIT PRICE AMOUNT

0001 40V TESTER

This acquisition is for the purchase of one (1) 40V Battery Tester, with applicable accessories. This purchase shall be IAW included Salient Characteristics.

Brand Name or Equal to:

Arbin Model LBT22023MC P/N: 917922

FFP

0001AA 40V TESTER

1 Each

In accordance with included Salient Characteristics

Brand Name or Equal to:

Arbin Model - LBT22023MC P/N: 917922

Accelerated or Partial delivery is acceptable by Government at no extra cost.

PSC: 6625

FOB: Destination

0001AB EQUIP CHARGE

Brand Name or Equal to:

Arbin Model - Standard Equipment Charge P/N: 600031

0001AC CHANNEL CABLE

16 Each

Brand Name or Equal to:

Arbin Model - IV Channel Cable P/N: Custom

0001AD BATTERY HOLDER

Brand Name or Equal to:

Arbin Model - Battery Holder-60A Universal Cylindrical Cell P/N: 409726

0001AE AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-Thermocouple Input Type-K-32CH P/N: 430518

0001AF AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-2nd Voltage-(-10~10V)-16CH P/N: 445080

0001AG AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-DIDO(Relay)-16CH P/N: 430576

0001AH AUX SYSTEM

3 Each

Brand Name or Equal to:

Arbin Model - Auxiliary System-MTCHI-ICH P/N: 449904

0001AJ CHANNEL CABLE

32 Each

Arbin Model - Auxiliary Channel Cable-K Type Thermocouple

P/N: 603744

0001AK CHANNEL CABLE

32 Each

Brand Name or Equal to:

Arbin Model - Auxiliary Channel Cable-Auxiliary Voltage/Digital I/O Cable P/N: 905740

0001AL CASTER WHEELS

4 Each

Brand Name or Equal to:

Arbin Model - Caster Wheels

0002 10V TESTER

This acquisition is for the purchase of one (1) 10V Battery Tester, along with applicable accessories. This purchase shall be IAW included Salient Characteristics.

Brand Name or Equal to:

Arbin Model - LBT21084HC P/N: 447790

0002AA 10V TESTER

Brand Name or Equal to:

Arbin Model - LBT21084HC P/N: 447790

0002AB EQUIPMENT CHARGE

Brand Name or Equal to:

Arbin Model - Standard Equipment Charge

0002AC CHANNEL CABLE

Brand Name or Equal to:

Arbin Model - I/V Channel Cable P/N: 410922

0002AD BATTERY HOLDER

64 Each

Brand Name or Equal to:

Arbin Model - Battery Holder-10A Universal Cylindrical Cell P/N: 409722

0002AE AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-Thermocouple Input Type-K-64CH P/N: 430522

0002AF AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-2nd Voltage-(-10~10V)-8CH P/N: 445078

0002AG AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-DIDO (Relay)-8CH P/N: 430574

0002AH AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-DIDO (Relay)-16CH

0002AJ CHANNEL CABLE 64 Each

Arbin Model - Auxiliary Channel Cable-K Type Thermocouple P/N: 603744

0002AK CHANNEL CABLE

48 Each

Brand Name or Equal to:

Arbin Model - Auxiliary Channel Cable-Auxiliary Voltage / Digital I/O Cable P/N: 905740

0002AL CASTER WHEELS

Brand Name or Equal to:

Arbin Model - Caster Wheels

0003 5V TEST STATION

This acquisition is for the purchase of two (2) 5V Battery Test Stations, along with applicable accessories. This purchase shall be IAW included Salient Characteristics.

Brand Name or Equal to:

Arbin Model - LBT21014 P/N: 920544

This line item must be delivered no earlier than twenty-eight (28) weeks after date of award.

0003AA 5V TESTER

2 Each

Brand Name or Equal to:

Arbin Model - LBT21014 P/N: 920544

0003AB EQUIP CHARGE

Brand Name or Equal to:

Arbin Model - Standard Equipment Charge

0003AC MAIN CABLE

10 Each

Brand Name or Equal to:

Arbin Model - Main I/V Cable-500A, 12ft P/N: 507826

0003AD AUX SYSTEM

Arbin Model - Auxiliary System-Thermocouple Input Type-T-16CH P/N: 430508

0003AE CSA INSPECTION

Brand Name or Equal to:

Arbin Model - CSA (NRTL) Factory Inspection P/N: 601772

0003AF TRAINING

1 Lot

Brand Name or Equal to:

Arbin Model - On-site Training P/N: 603786

0004 60V TEST SYSTEM

This acquisition is for the purchase of two (2) 60V Battery Test Stations, along with applicable accessories.

This purchase shall be IAW included Salient Characteristics.

Brand Name or Equal to:

Arbin Model - RBT32013 P/N: 921500

0004AA 60V TESTER

Brand Name or Equal to:

Arbin Model - RBT32013

0004AB EQUIP CHARGE

Brand Name or Equal to:

Arbin Model - Standard Equipment Charge

0004AC MAIN CABLE

5 Each

Brand Name or Equal to:

Arbin Model - Main I/V Cable-500A, 12ft P/N: 410608

0004AD AUX SYSTEM 1 Each

Arbin Model - Auxiliary System-Thermocouple Input Type-T-48CH P/N: 430512

0004AE AUX SYSTEM

2 Lot

Brand Name or Equal to:

Arbin Model - Auxiliary System-2nd Voltage-(-5~5V)-16CH P/N: 445076

0004AF CSA INSPECTION

2 Lot

Brand Name or Equal to:

Arbin Model - CSA (NRTL) Factory Inspection P/N: 601772

0004AG TRAINING

Brand Name or Equal to:

Arbin Model - On-site Training

0005 9-400V TESTER

This acquisition is for the purchase of one (1) 9-400V Battery Test Stations, along with applicable accessories.

This purchase shall be IAW included Salient Characteristics.

Brand Name or Equal to:

Arbin Model - RBT33022 P/N: 452446

0005AA 9-400V TESTER

Brand Name or Equal to:

Arbin Model - RBT33022 P/N: 452446

0005AB EQUIP CHARGE

Brand Name or Equal to:

Arbin Model - Standard Equipment Charge

0005AC MAIN CABLE

8 Each

Arbin Model - Main I/V Cable-200A, 20ft P/N: 453734

0005AD AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-Thermocouple Input, Type-T-48CH P/N: 430512

0005AE AUX SYSTEM

8 Each

Brand Name or Equal to:

Arbin Model - Auxiliary System, CAN FD, ICH P/N: 449902

0005AF AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System, Digital I/O, Relay, 16CH

0005AG AUX SYSTEM

Brand Name or Equal to:

Arbin Model - Auxiliary System-MTCI-1CH P/N: 449904

0005AH CHANNEL CABLE

Brand Name or Equal to:

Arbin Model - Auxiliary Channel Cable P/N: 178202

0005AJ CSA INSPECTION

Brand Name or Equal to:

Arbin Model - CSA (NRTL) Factory Inspection P/N: 601772

0005AK ONSITE SUPPORT

Brand Name or Equal to:

Arbin Model - On-site Support for New Systems

0006 5V TEST STATION

OPTION In accordance with included Salient Characteristics

Brand Name or Equal to:

Arbin Model - LBT21014 (along with applicable accessories) P/N: 920544

Option Exercise Period: From time of award to one year thereafter

Period of Performance: 28 weeks After Option Exercise

0007 60V TEST SYSTEM

OPTION In accordance with included Salient Characteristics

Arbin Model - RBT32013 (along with applicable accessories)

Option Exercise Period: From time of award to one year thereafter

Period of Performance: 28 weeks After Option Exercise

40V SALIENT CHARACTERISTICS

Specification for:

40V Battery Safety Test System

1.0 SCOPE

1.1 This specification describes the requirements for a rechargeable battery test systems. The system shall contain a minimum of sixteen channels with a minimum voltage range of 40V and a minimum current range of 20A.

2.0 REQUIREMENTS

2.1 Overall Test System Requirements.

2.1.1 The system shall be capable of charging and discharging batteries simultaneously and independently of each other. If required, the contractor shall perform the final installation. All electrical and other utilities to the unit will be attached by the user.

2.1.2 All sixteen identical 40V-channels shall be mounted in single test rack. Each rack shall be no more than 30” wide. The test system shall be portable and shall have wheels to permit movement from facility to facility. The system shall be air cooled.

2.1.3 The rack shall be able to be expanded in the field if required by the user.

2.1.4 All data acquisition devices shall have a minimum of 24-bit resolution and meet the enclosed requirements.

2.1.5 Each channel shall be capable of reading over a voltage range of 0 to 40V DC.

Under either charge or discharge the 40V DC channels shall provide a minimum of 20A.

2.1.6 If multiple current ranges are provided the channels shall seamlessly transition between these ranges.

2.1.7 Each multichannel test rack shall permit paralleling of up to sixteen channels to allow for higher discharge and charge rates of up to 320A.

2.1.8 Each channel shall be able to run a totally independent test profile. The channels shall be capable of charging and discharging different batteries to different profiles and different start and stop times.

2.1.9 Each channel shall be able to charge or discharge batteries in the following modes: constant current, constant power, constant resistance as well as perform current ramps (triangular pulses). This system shall be able to use auxiliary inputs such as cell or pack voltages and temperature for these functions such as ending charging or discharging based on any auxiliary input.

2.1.10 Each channel shall be able to change loads in 5 milliseconds (ms) or less from zero to maximum value or from maximum value to zero. During load changes the current shall not overshoot. When the current is increased during a load change the current during transition shall never (including overshoot) exceed the final current value plus the tolerance of ±0.2% of the range + 0.2% of the set point. When the current is being decreased the current value during transition shall never (including overshoot) fall below the final specified value minus the tolerance of ±0.2% of the range + 0.2% of the set point for the appropriate current range.

2.1.11 Each channel shall be able to provide a controlled current during charge or discharge over a range from 0 A to maximum current as follows. At 20A shall be a minimum of + 10mA and at 100mA a minimum of + 50.0uA.

2.1.12 Each channel shall be able to perform constant power charge of 0 W to maximum power with an accuracy of + 0.02% FSD.

2.1.13 Each channel shall be able to perform constant resistance operation over a range of 0 Ohms to 1 Mega Ohms with an accuracy of + 0.02% FSD.

2.1.14 Each channel shall be capable of performing pulses of widths 10 ms or less during which time data shall be recorded at 1 ms rate internally.

2.1.15 All calculated control values Amp hour or watt hour capacity shall be continuously updated continually at 1 msec. interval.

2.2 Data Measurement and Control.

2.2.1 Primary measurement by the system (excluding auxiliary inputs) shall use 4 point measurements.

2.2.2 The system shall be able to measure, adjust, and control within a time period of

1 millisecond all the channels.

2.2.3 Voltage resolution for the 40V channel the resolution shall be 200 uV. The meters shall have a minimum accuracy of 0.001% FSR.

2.2.4 Voltage resolution for auxiliary voltmeters 1 mV at + 40Vdc range and an accuracy of 0.003% FSR.

2.2.5 The current accuracy for each channel shall be at a minimum 0.02%FSD +

0.001%RDG.

2.2.6 Each channel shall be able to measure and record temperatures using K Type thermocouples.

2.2.7 The test system and rack shall be air cooled.

2.2.8 The test systems shall use Ethernet to communicate with PCs.

2.2.9 The systems shall use 220V 3 Phase power circuit.

2.2.10 Battery connections for each channel shall be a true 4-wire connection. The end user shall provide cables such that the two wires that carry the load shall be terminated at the battery terminals, and the two wires that measure the battery voltage shall be terminated at the battery terminals.

2.3 Watchdog Timer and Redundant Safety Limits

2.3.1 Each channel shall go to an open or suspend state if for any reason the channel is unable to communicate with the test computer.

2.3.2 Each channel shall be able to open or suspend testing if any of the user defined voltage or current, or temperature limits have been exceeded.

2.3.3 The system shall go to an open or suspend state if for any reason the system cannot communicate with an individual channel, or any of the critical components such as system cooling loop, test system temperature, power faults, etc.

2.3.4 Each channel shall have a current limiting device with a programmable trip in time to prevent accidental cell or battery shorting due to system failure.

2.3.5 Once the system or channel is down it will not be permitted to restart until the fault is cleared. The system shall be able to restart at the point at which it was halted automatically without having to either reenter data or advance any cycle or step indexes.

2.4 Auxiliary voltage measurements.

2.4.1 The system shall be able to read a minimum of 24 auxiliary voltage readings with a range of +10V to -10V.

2.4.2 The auxiliary voltage channels shall be able to be mapped to any combination of channels in any quantity.

2.5 Auxiliary digital inputs and outputs.

2.5.1 The system shall be provided with a minimum of 24 digital inputs and 8 digital outputs.

2.5.2 The auxiliary digital channels shall be able to be mapped to any combination of channels in any quantity.

2.6 Test control computer

2.6.1 Three identical test control computers shall be provided to permit individual users to run test on each tester.

2.6.2 All test control computers shall be provided with either Windows 10 Ultimate or

Windows 10 Enterprise.

2.6.3 All test control computers will have the same test system software.

2.7 Test System Software.

2.7.1 Software shall permit each channel to perform the following types of charge or discharge:

a. Constant Current.

b. Constant Voltage.

c. Constant Power.

d. Constant Resistance

e. Voltage ramps or scans

f. Perform DC Internal Resistance Measurements.

2.7.2 Each test step shall have at a minimum use the following conditions to control or end the step; volts, amps, temperature, or time but shall also be able to use integrated values (C-Rate capacity, capacity {C-Rate capacity, Ampere hours [Ah], Watt hours [Wh], Last Cycle Ah, Last Cycle Wh}, and calculated values (first and second derivatives of dV/dt, dT/dt).

2.7.3 The test system software shall be able to show the current real time status of all channels on a screen.

2.7.4 The test system software shall be able to show all the current real time test parameters of any of the channels.

2.7.5 The software shall be able to create test profiles by importing time vs. current, time vs. voltage, and time vs. power data from other sources without having to create individual steps.

2.8 The company shall ensure that the system is functioning properly after shipment and installation. The systems test channel calibrations shall be verified at the site and if required calibrated on-site. After delivery and installation the system shall be certified to conform to the specifications.

2.9 The tester shall be able to control test temperatures in a minimum of two temperature chambers.

2.9.1 The tester shall be able to communicate with the following chambers or controllers; Cincinnatti sub zero EZT570S, ESPEC P300, and Watlow F4T.

2.9.2 The tester shall be able adjust temperature set points in the middle of the test program and to wait until the chamber is acclimated before performing steps.

2.10 A minimum of one full day of onsite operational training shall be provided.

2.11 A set of operating manuals shall be provided as either a paper copy or PDF.

2.12 Test System Calibration.

2.12.1 The test system shall include detailed channel calibration procedures and an easily accessible calibration program. Test system calibration shall be able to be performed by the user.

2.12.2 The calibration procedure shall be automated and shall only require a calibrated multi meter, a calibrated current shunt, and an isolated voltage source.

2.12.3 The systems shall be configured to permit the calibration of all the channels in a single system simultaneously.

2.12.4 The system shall be delivered calibrated or calibrated on site to NIST traceable standards with a calibration that is valid for a full year.

2.13 Test system voltage requirement is 240Vdc 3 Phase.

3.0 Supplies

3.1 Main Test System Cables

3.1.1 Each channel shall be provided with a 20ft four wire test cable with ring lugs that connect to locking cylindrical cell holder.

3.1.2 Each cable will use one set of leads to carry current and a second set for sense lines.

3.1.3 Cables shall be designed for 20A loads.

3.1.4 Each cable shall be provided with a camlock battery holder that is adjustable.

The holder will use two connection points to cell or battery to be tested. A large one will be used for carrying the current and a smaller pointed one will be used for voltage sense. The camlock device will be able to be adjusted to handle sells or batteries up to 6” long and 2” in diameter.

3.2 Auxiliary Cables

3.2.1 Each individual K type thermocouple channel shall be provided with a 20ft long

K type thermocouple cable.

3.2.2 Each auxiliary voltage or digital input or output channel shall be provided with a

20 ft cable.

3.3 Calibration supplies

3.3.1 All cabling and supplies required to interface the test system to user supplied calibrated multi meter, user supplied calibrated current shunt, and user supplied isolated voltage source shall be provided.

10V SALIENT CHARACTERISTICS

10V Battery Safety Test System

4.0 SCOPE

4.1 This specification describes the requirements for a rechargeable battery test systems. The system shall contain a minimum of sixty-four channels with a minimum voltage range of 10V and a minimum current range of 10A.

5.0 REQUIREMENTS

5.1 Overall Test System Requirements.

5.1.1 The system shall be capable of charging and discharging batteries simultaneously and independently of each other. If required, the contractor shall perform the final installation. All electrical and other utilities to the unit will be attached by the user.

5.1.2 All sixty-four identical 10V-channels shall be mounted in single test rack. Each rack shall be no more than 30” wide. The test system shall be portable and shall have wheels to permit movement from facility to facility. The system shall be air cooled.

5.1.3 The rack shall be able to be expanded in the field if required by the user.

5.1.4 All data acquisition devices shall have a minimum of 24-bit resolution and meet the enclosed requirements.

5.1.5 Each channel shall be capable of reading over a voltage range of 0 to 10V DC.

Under either charge or discharge the 10V DC channels shall provide a minimum of 10A.

5.1.6 If multiple current ranges are provided the channels shall seamlessly transition between these ranges.

5.1.7 Each multichannel test rack shall permit paralleling of up to sixteen channels to allow for higher discharge and charge rates of up to 160A for the 10V tester.

5.1.8 Each channel shall be able to run a totally independent test profile. The channels shall be capable of charging and discharging different batteries to different profiles and different start and stop times.

5.1.9 Each channel shall be able to charge or discharge batteries in the following modes: constant current, constant power, constant resistance as well as perform current ramps (triangular pulses). This system shall be able to use auxiliary inputs such as cell or pack voltages and temperature for these functions such as ending charging or discharging based on any auxiliary input.

5.1.10 Each channel shall be able to change loads in 5 milliseconds (ms) or less from zero to maximum value or from maximum value to zero. During load changes the current shall not overshoot. When the current is increased during a load change the current during transition shall never (including overshoot) exceed the final current value plus the tolerance of ±0.2% of the range + 0.2% of the set point. When the current is being decreased the current value during transition shall never (including overshoot) fall below the final specified value minus the tolerance of ±0.2% of the range + 0.2% of the set point for the appropriate current range.

5.1.11 Each channel shall be able to provide a controlled current during charge or discharge over a range from 0 A to maximum current as follows. For the 10V 10A channels the current control at 10A shall be a minimum of + 5mA and at 5mA a minimum of + 2.5uA.

5.1.12 Each channel shall be able to perform constant power charge of 0 W to maximum power with an accuracy of + 0.02% FSD.

5.1.13 Each channel shall be able to perform constant resistance operation over a range of 0 Ohms to 1 Mega Ohms with an accuracy of + 0.02% FSD.

5.1.14 Each channel shall be capable of performing pulses of widths 10 ms or less during which time data shall be recorded at 1 ms rate internally.

5.1.15 All calculated control values Amp hour or watt hour capacity shall be continuously updated continually at 1 msec. interval.

5.2 Data Measurement and Control.

5.2.1 Primary measurement by the system (excluding auxiliary inputs) shall use 4 point measurements.

5.2.2 The system shall be able to measure, adjust, and control within a time period of

1 millisecond all the channels.

5.2.3 Voltage resolution for the 10V channel the resolution shall be 200 uV. The meters shall have a minimum accuracy of 0.001% FSR.

5.2.4 Voltage resolution for auxiliary voltmeters 1 mV at + 10Vdc range and an accuracy of 0.003% FSR.

5.2.5 The current accuracy for each channel shall be at a minimum 0.02%FSD +

0.001%RDG.

5.2.6 Each channel shall be able to measure and record temperatures using K Type thermocouples.

5.2.7 The test system and rack shall be air cooled.

5.2.8 The test systems shall use Ethernet to communicate with PCs.

5.2.9 The systems shall use 220V 3 Phase power circuit.

5.2.10 Battery connections for each channel shall be a true 4-wire connection. The end user shall provide cables such that the two wires that carry the load shall be terminated at the battery terminals, and the two wires that measure the battery voltage shall be terminated at the battery terminals.

5.3 Watchdog Timer and Redundant Safety Limits

5.3.1 Each channel shall go to an open or suspend state if for any reason the channel is unable to communicate with the test computer.

5.3.2 Each channel shall be able to open or suspend testing if any of the user defined voltage or current, or temperature limits have been exceeded.

5.3.3 The system shall go to an open or suspend state if for any reason the system cannot communicate with an individual channel, or any of the critical components such as system cooling loop, test system temperature, power faults, etc.

5.3.4 Each channel shall have a current limiting device with a programmable trip in time to prevent accidental cell or battery shorting due to system failure.

5.3.5 Once the system or channel is down it will not be permitted to restart until the fault is cleared. The system shall be able to restart at the point at which it was halted automatically without having to either reenter data or advance any cycle or step indexes.

5.4 Auxiliary voltage measurements.

5.4.1 The system shall be able to read a minimum of 24 auxiliary voltage readings with a range of +10V to -10V.

5.4.2 The auxiliary voltage channels shall be able to be mapped to any combination of channels in any quantity.

5.5 Auxiliary digital inputs and outputs.

5.5.1 The system shall be provided with a minimum of 24 digital inputs and 8 digital outputs.

5.5.2 The auxiliary digital channels shall be able to be mapped to any combination of channels in any quantity.

5.6 Test control computer

5.6.1 Three identical test control computers shall be provided to permit individual users to run test on each tester.

5.6.2 All test control computers shall be provided with either Windows 10 Ultimate or

Windows 10 Enterprise.

5.6.3 All test control computers will have the same test system software.

5.7 Test System Software.

5.7.1 Software shall permit each channel to perform the following types of charge or discharge:

a. Constant Current.

b. Constant Voltage.

c. Constant Power.

d. Constant Resistance

e. Voltage ramps or scans

f. Perform DC Internal Resistance Measurements.

5.7.2 Each test step shall have at a minimum use the following conditions to control or end the step; volts, amps, temperature, or time but shall also be able to use integrated values (C-Rate capacity, capacity {C-Rate capacity, Ampere hours [Ah], Watt hours [Wh], Last Cycle Ah, Last Cycle Wh}, and calculated values (first and second derivatives of dV/dt, dT/dt).

5.7.3 The test system software shall be able to show the current real time status of all channels on a screen.

5.7.4 The test system software shall be able to show all the current real time test parameters of any of the channels.

5.7.5 The software shall be able to create test profiles by importing time vs. current, time vs. voltage, and time vs. power data from other sources without having to create individual steps.

5.8 The company shall ensure that the system is functioning properly after shipment and installation. The systems test channel calibrations shall be verified at the site and if required calibrated on-site. After delivery and installation the system shall be certified to conform to the specifications.

5.9 The tester shall be able to control test temperatures in a minimum of two temperature chambers.

5.9.1 The tester shall be able to communicate with the following chambers or controllers; Cincinnatti sub zero EZT570S, ESPEC P300, and Watlow F4T.

5.9.2 The tester shall be able adjust temperature set points in the middle of the test program and to wait until the chamber is acclimated before performing steps.

5.10 A minimum of one full day of onsite operational training shall be provided.

5.11 A set of operating manuals shall be provided as either a paper copy or PDF.

5.12 Test System Calibration.

5.12.1 The test system shall include detailed channel calibration procedures and an easily accessible calibration program. Test system calibration shall be able to be performed by the user.

5.12.2 The calibration procedure shall be automated and shall only require a calibrated multi meter, a calibrated current shunt, and an isolated voltage source.

5.12.3 The systems shall be configured to permit the calibration of all the channels in a single system simultaneously.

5.12.4 The system shall be delivered calibrated or calibrated on site to NIST traceable standards with a calibration that is valid for a full year.

5.13 Test system voltage requirement is 240Vdc 3 Phase.

6.0 Supplies

6.1 Main Test System Cables

6.1.1 Each channel shall be provided with a 20 ft four wire test cable with ring lugs that connect to locking cylindrical cell holder.

6.1.2 Each cable will use one set of leads to carry current and a second set for sense lines.

6.1.3 Cables shall be designed for 10A loads.

6.1.4 Each cable shall be provided with a camlock battery holder that is adjustable.

The holder will use two connection points to cell or battery to be tested. A large one will be used for carrying the current and a smaller pointed one will be used for voltage sense. The camlock device will be able to be adjusted to handle sells or batteries up to 6” long and 2” in diameter.

6.2 Auxiliary Cables

6.2.1 Each individual K type thermocouple channel shall be provided with a 20 ft long

K type thermocouple cable.

6.2.2 Each auxiliary voltage or digital input or output channel shall be provided with a

20 ft cable.

6.3 Calibration supplies

6.3.1 All cabling and supplies required to interface the test system to user supplied

5V SALIENT CHARACTERISTICS

5V Battery Cycling Test System

7.0 SCOPE

7.1 This specification describes the requirements for a rechargeable battery test system. The system shall contain a minimum of five channels with a minimum voltage range of 5V and a minimum current range of 500A.

8.0 REQUIREMENTS

8.1 Overall Test System Requirements.

8.1.1 The system shall be capable of charging and discharging batteries simultaneously and independently of each other. If required, the contractor shall perform the final installation. All electrical and other utilities to the unit will be attached by the user.

8.1.2 All five identical 5V-channels shall be mounted in single test rack. Each rack shall be no more than 40” wide. The test system shall be portable and shall have wheels to permit movement from facility to facility. The system shall be air cooled.

8.1.3 All data acquisition devices shall have a minimum of 24-bit resolution and meet the enclosed requirements.

8.1.4 Each channel shall be capable of reading over a voltage range of 0 to 5V DC.

Under either charge or discharge the 5V DC channels shall provide a minimum of 500A.

8.1.5 If multiple current ranges are provided the channels shall seamlessly transition between these ranges.

8.1.6 Each multichannel test rack shall permit paralleling of up to five channels to allow for higher discharge and charge rates of up to 2000A for the 5V tester.

8.1.7 Each channel shall be able to run a totally independent test profile. The channels shall be capable of charging and discharging different batteries to different profiles and different start and stop times.

8.1.8 Each channel shall be able to charge or discharge batteries in the following modes: constant current, constant power, and constant voltage. This system shall be able to use auxiliary inputs such as cell or pack voltages and temperature for these functions such as ending charging or discharging based on any auxiliary input.

8.1.9 Each channel shall be able to change loads in 10 milliseconds (ms) or less from zero to maximum value or from maximum value to zero. During load changes the current shall not overshoot. When the current is increased during a load change the current during transition shall never (including overshoot) exceed the final current value plus the tolerance of ±0.2% of the range + 0.2% of the set point. When the current is being decreased the current value during transition shall never (including overshoot) fall below the final specified value minus the tolerance of ±0.2% of the range + 0.2% of the set point for the appropriate current range.

8.1.10 Each channel shall be able to provide a controlled current during charge or discharge over a range from 0 A to maximum current as follows. For the 5V 500A channels the current control at 500A shall be a minimum of + 100mA and at 500mA a minimum of + 1mA.

8.1.11 Each channel shall be able to perform constant power charge of 0 W to maximum power with an accuracy of + 0.02% FSD.

8.1.12 All calculated control values Amp hour or watt hour capacity shall be continuously updated continually at 10 msec. interval.

8.2 Data Measurement and Control.

8.2.1 The system shall be able to measure, adjust, and control within a time period of

5 millisecond all the channels.

8.2.2 Voltage resolution for the 5V channel the resolution shall be 1mV. The meters shall have a minimum accuracy of 0.02% FSR.

8.2.3 The current accuracy for each channel shall be at a minimum 0.02%FSD

8.2.4 Each channel shall be able to measure and record temperatures using K Type thermocouples.

8.2.5 The test system and rack shall be air cooled.

8.2.6 The test systems shall use Ethernet to communicate with PCs.

8.2.7 The systems shall use 208V 3 Phase power circuit.

8.2.8 Battery connections for each channel shall be a true 4-wire connection. The end user shall provide cables such that the two wires that carry the load shall be terminated at the battery terminals, and the two wires that measure the battery voltage shall be terminated at the battery terminals.

8.3 Watchdog Timer and Redundant Safety Limits

8.3.1 Each channel shall go to an open or suspend state if for any reason the channel is unable to communicate with the test computer.

8.3.2 Each channel shall be able to open or suspend testing if any of the user defined voltage or current, or temperature limits have been exceeded.

8.3.3 The system shall go to an open or suspend state if for any reason the system cannot communicate with an individual channel, or any of the critical components such as system cooling loop, test system temperature, power faults, etc.

8.3.4 Each channel shall have a current limiting device with a programmable trip in time to prevent accidental cell or battery shorting due to system failure.

8.3.5 Once the system or channel is down it will not be permitted to restart until the fault is cleared. The system shall be able to restart at the point at which it was halted automatically without having to either reenter data or advance any cycle or step indexes.

8.4 Test control computer

8.4.1 A single test control computers shall be provided with either Windows 10

Ultimate or Windows 10 Enterprise.

8.4.2 The test control computer shall be able to control up to four 5V testers at the same time.

8.5 Test System Software.

8.5.1 The test system software shall be able to show the current real time status of all channels on a screen.

8.5.2 The test system software shall be able to show all the current real time test parameters of any of the channels.

8.5.3 The software shall be able to create test profiles by importing time vs. current, time vs. voltage, and time vs. power data from other sources without having to create individual steps.

8.6 The company shall ensure that the system is functioning properly after shipment and installation. The systems test channel calibrations shall be verified at the site and if required calibrated on-site. After delivery and installation the system shall be certified to conform to the specifications.

8.7 A minimum of one full day of onsite operational training shall be provided.

8.8 A set of operating manuals shall be provided as either a paper copy or PDF.

8.9 Test System Calibration.

8.9.1 The test system shall include detailed channel calibration procedures and an easily accessible calibration program. Test system calibration shall be able to be performed by the user.

8.9.2 The calibration procedure shall be automated and shall only require a calibrated multi meter, a calibrated current shunt, and an isolated voltage source.

8.9.3 The systems shall be configured to permit the calibration of all the channels in a single system simultaneously.

8.9.4 The system shall be delivered calibrated or calibrated on site to NIST traceable standards with a calibration that is valid for a full year.

8.10 Test system voltage requirement is 208Vdc 3 Phase.

9.0 Supplies

9.1 Main Test System Cables

9.1.1 Each channel shall be provided with a 20 ft four wire test cable with ring lugs

9.1.2 Each cable will use one set of leads to carry current and a second set for sense lines.

9.1.3 Cables shall be designed for 500A loads.

9.2 Auxiliary Cables

9.2.1 Each individual K type thermocouple channel shall be provided with a 20 ft long

K type thermocouple cable.

9.3 Calibration supplies

9.3.1 All cabling and supplies required to interface the test system to user supplied

60V SALIENT CHARACTERISTICS

60V Battery Cycling Test System

10.0 SCOPE

10.1 This specification describes the requirements for a rechargeable battery test system. The system shall contain a minimum of five channels with a minimum voltage range of 60V and a minimum current range of 500A.

11.0 REQUIREMENTS

11.1 Overall Test System Requirements.

11.1.1 The system shall be capable of charging and discharging batteries simultaneously and independently of each other. If required, the contractor shall perform the final installation. All electrical and other utilities to the unit will be attached by the user.

11.1.2 All five identical 60V-channels shall be mounted in single test rack. Each rack shall be no more than 42” wide. The test system shall be portable and shall have wheels to permit movement from facility to facility. The system shall be air cooled.

11.1.3 All data acquisition devices shall have a minimum of 16-bit resolution and meet the enclosed requirements.

11.1.4 Each channel shall be capable of reading over a voltage range of 9 to 60V DC.

Under either charge or discharge the 60V DC channels shall provide a minimum continuous power output of 26000W

11.1.5 If multiple current ranges are provided the channels shall seamlessly transition between these ranges.

11.1.6 Each multichannel test rack shall permit paralleling of up to five channels to allow for higher discharge and charge rates of up to 2000A and 60000W total power for the 60V tester.

11.1.7 Up to four 60V testers shall be able to be run in parallel to allow for higher discharge and charges rates of up to 5000A and 160000W total power.

11.1.8 Each channel shall be able to run a totally independent test profile. The channels shall be capable of charging and discharging different batteries to different profiles and different start and stop times.

11.1.9 Each channel shall be able to charge or discharge batteries in the following modes: constant current, constant power, and constant voltage. This system shall be able to use auxiliary inputs such as cell or pack voltages and temperature for these functions such as ending charging or discharging based on any auxiliary input.

11.1.10 Each channel shall be able to change loads in 20 milliseconds (ms) or less from zero to maximum value or from maximum value to zero. During load changes the current shall not overshoot. When the current is increased during a load change the current during transition shall never (including overshoot) exceed the final current value plus the tolerance of ±0.5% of the range + 0.5% of the set point. When the current is being decreased the current value during transition shall never (including overshoot) fall below the final specified value minus the tolerance of ±0.5% of the range + 0.5% of the set point for the appropriate current range.

11.1.11 Each channel shall be able to provide a controlled current during charge or discharge over a range from 0 A to maximum current as follows. For the 60V 500A channels the current control at 500A shall be a minimum of +250mA and at 125A a minimum of +70mA.

11.1.12 All calculated control values Amp hour or watt hour capacity shall be continuously updated continually at 10 msec. interval.

11.1.13 Tester shall be capable of communicating with a Battery Management System (BMS) via CAN bus protocol.

11.2 Data Measurement and Control.

11.2.1 The system shall be able to measure, adjust, and control within a time period of

5 millisecond all the channels.

11.2.2 Voltage resolution for the 60V channel the resolution shall be 30mV. The meters shall have a minimum accuracy of 0.05% FSR.

11.2.3 The current accuracy for each channel shall be at a minimum 0.05%FSD

11.2.4 Each channel shall be able to measure and record temperatures using K Type thermocouples.

11.2.5 The test system and rack shall be air cooled.

11.2.6 The test systems shall use Ethernet to communicate with PCs.

11.2.7 The systems shall use 480V 3 Phase power circuit.

11.2.8 Battery connections for each channel shall be a true 4-wire connection. The end user shall provide cables such that the two wires that carry the load shall be terminated at the battery terminals, and the two wires that measure the battery voltage shall be terminated at the battery terminals.

11.3 Watchdog Timer and Redundant Safety Limits

11.3.1 Each channel shall go to an open or suspend state if for any reason the channel is unable to communicate with the test computer.

11.3.2 Each channel shall be able to open or suspend testing if any of the user defined voltage or current, or temperature limits have been exceeded.

11.3.3 The system shall go to an open or suspend state if for any reason the system cannot communicate with an individual channel, or any of the critical components such as system cooling loop, test system temperature, power faults, etc.

11.3.4 Each channel shall have a current limiting device with a programmable trip in time to prevent accidental cell or battery shorting due to system failure.

11.3.5 Once the system or channel is down it will not be permitted to restart until the fault is cleared. The system shall be able to restart at the point at which it was halted automatically without having to either reenter data or advance any cycle or step indexes.

11.4 Test control computer

11.4.1 A single test control computers shall be provided with either Windows 10

Ultimate or Windows 10 Enterprise.

11.4.2 The test control computer shall be able to control up to four 60V testers at the same time.

11.4.3 All 60V testers controlled by the same control computer shall be able to run in parallel.

11.5 Test System Software.

11.5.1 The test system software shall be able to show the current real time status of all channels on a screen.

11.5.2 The test system software shall be able to show all the current real time test parameters of any of the channels.

11.5.3 The software shall be able to create test profiles by importing time vs. current, time vs. voltage, and time vs. power data from other sources without having to create individual steps.

11.6 The company shall ensure that the system is functioning properly after shipment and installation. The systems test channel calibrations shall be verified at the site and if required calibrated on-site. After delivery and installation the system shall be certified to conform to the specifications.

11.7 A minimum of one full day of onsite operational training shall be provided.

11.8 A set of operating manuals shall be provided as either a paper copy or PDF.

11.9 Test System Calibration.

11.9.1 The test system shall include detailed channel calibration procedures and an easily accessible calibration program. Test system calibration shall be able to be performed by the user.

11.9.2 The calibration procedure shall be automated and shall only require a calibrated multi meter, a calibrated current shunt, and an isolated voltage source.

11.9.3 The systems shall be configured to permit the calibration of all the channels in a single system simultaneously.

11.9.4 The system shall be delivered calibrated or calibrated on site to NIST traceable standards with a calibration that is valid for a full year.

11.10 Test system voltage requirement is 480Vdc 3 Phase.

12.0 Supplies

12.1 Main Test System Cables

12.1.1 Each channel shall be provided with a 20 ft four wire test cable with ring lugs

12.1.2 Each cable will use one set of leads to carry current and a second set for sense lines.

12.1.3 Cables shall be designed for 500A loads.

12.2 Auxiliary Cables

12.2.1 Each individual K type thermocouple channel shall be provided with a 20 ft long

K type thermocouple cable.

12.3 Calibration supplies

12.3.1 All cabling and supplies required to interface the test system to user supplied

400V SALIENT CHARACTERISTICS

60V Battery Cycling Test System

The test system shall incorporate 8 individual test stations that are able to test between 9V – 400V or from 9V to 50V for both charging and discharging profiles. Charging mode is to provide at least 100A charging current in a Constant Voltage mode, 150A discharge in a regenerative power discharge mode at a 70% minimum efficiency. This allows for the test system to be used on 270V and 28V batteries for testing while controlling accuracy.

Test system shall provide Power Output to charge and discharge Devices Under Test (DUT) and must be able to limit charge and discharge to provide either Constant Voltage, Constant Current, Constant Power, or Fixed Resistance levels to meet charging and discharging requirements.

Power output rating per channel is to exceed 52.5kW per channel between 5°C and 30°C room temperature or 150kW at any time across all 8 stations for a 45 minute duration.

Each test station shall be able to run independent tests or be combined in parallel fashion as needed to increase current requirements (parallel configurations must be in the same voltage range setting) to 1100A minimum for discharge durations not to exceed 30 seconds (Constant Voltage, Constant Power, or Fixed Resistor Discharges). Voltage limitation to prevent exceeding power dissipation levels can exist.

System response to reach the final Resistance or Voltage setting on Constant Voltage, Constant Power, or Fixed Resistor Discharges shall not exceed 20ms even when channels are combined in parallel up to 1050A.

Test system shall have automatic range limiting and switching of both charging and discharging functions to allow stability to ±1% max of voltage or current levels from the power supply mode (either 400V or 50V settings) while operating within its specification allowed current levels and power dissipation levels.

Testers to have integrated safety features that stop and abort testing when exceeding the current level or power dissipation levels of the test station. This should shut down only the individual stations with a reason clearly documented.

Testers to have integrated safety features that stop and abort testing when exceeding the current level or power dissipation levels of the entire system if system level requirements are exceeded such as 4 stations each pulling 50kW of power at the same time.

The fluctuation in current will not exceed ±2% of the programed / operating current for any charge or discharge rates for the duration of the test.

Test system shall be able to record Actual Test time, Actual date and time, step charge times, step discharge times, actual total charge times, actual total discharge times, Temperature readings, Voltage readings, and Current readings for each test station. This data is to be stored in the system and can be saved off to excel compatible data files for analysis by Engineers.

Each data reading being recorded should be programmable from 1ms to at least 10 minutes (we can accept longer durations if it is already part of the system) depending on the need and accuracy of the data being recorded

The test system will be able to bypass readings and still perform the required testing of the above inputs if they are not connected or being used for the particular test.

When tests are interrupted for any reason the tester shall remember its location in the test and be able to be resumed with the subsequent data automatically added to the original test data, with the reason for the interruption documented in the break (loss of power; Channel 3 temperature out of limit, Channel 5 voltage out of limit, etc.).

The Test System Shall require only electrical power connections to operate. No Cooling water or other items that prohibit or restrict relocation to other areas is allowed.

Power to be 3 Phase; 50 Hz– 60 Hz; 380V – 480V AC power at 250 A Maximum and 220V single phase power at 30A maximum (Target values we can discuss if needed).

The test system shall be able to be programmed with up to 300 steps of various charging and discharging functions per profile with each function having controlled limit tolerances and safety limit tolerances incorporated for test integrity. Repeat commands need to be available to run the programmed profile again up to 100,000 cycles

The test system shall be able to retain at least 200 profiles in its system memory. The ability to print out the profile in a manner that can be graphed, or the graph of the profile for simulation and verification is preferred.

If this feature is currently not available, it is expected that when the feature is created for our…