QUESTIONS & ANSWERS RE:

SOLICITATION # 1645BC-18-R-00005

1. QUESTION: Will you accept an instrument with a 300mm Goniometer Radius. The 300mm instrument will be fully capable of meeting the NIOSH 7500 testing protocol.

ANSWER: Yes. See Amended solicitation.

1. QUESTION: Will you accept a sample holder minimum capacity of 48 samples (60 samples is specified) ANSWER: Yes. See Amended solicitation.

1. QUESTION: The bid describes the choice of two types of detectors:

2. high speed compound silicon strip detector (minimum 100 channels)

OR

2. Gas proportional or scintillation type detector Which type of detector is to be specified (or both) since there is a cost difference associated with these two types of detectors ANSWER: See Amended solicitation. The solicitation has been amended to require item a (silicon strip detector) and make b (gas proportional or scintillation type detector) an “optional” line item. The price for item b will not be evaluated and may or may not be purchased by the Government.