Questions and Answers

Question # 1: We are unclear about the level of installation you will require. Is there already a tube support that will mount to this tube ? We need to understand how much fabrication and wiring into your existing door interlocks or xray on light is required.

NIST Answer: The following instructions have been added to the Amended NIST Requirements document: NIST will provide a tube support and mounting brackets to mount the system, interlock wiring to interface the Xray Worx controller with the NIST safety PLC (one channel for key and button arming control and one channel for magnetic door interlock), a CHRH single phase 30A door interlock to be wired to high voltage source, and power cord to wire to high voltage source.”