• Imaging Area at least: 23cm x 29 cm (9” x 11.5”)
• Pixel Pitch: 48 µm
• Spatial Resolution: 5 µm at 10X geometric magnification (10% MTF (Modulation Transfer
Function) at 48 µm)
• Digital Detector: Cesium Iodide (CsI)
• Energy Range: 10 – 100 kV
• Tube Current: 0.3 mA max (isowatt limited to 12W)
• Focal Spot Size, Nominal: 8 µm
• Window Filtration: 0.25 mm Be, tungsten source
• Automatic Exposure Control
• Safety features with no OSHA license, nor radiological test required
• Must not require any additional x-ray shielding (shielding must be fully built-in)
• Stainless steel interior chamber for easy cleaning
• Power Requirements: 100 – 240 VAC ±10%, 50/60 Hz, 150VA
• A cesium iodide scintillator for its sharper images
• Software to enable us to automatically measure seed size, shell thickness and to count seeds.