UV Sterilizer QandA.pdf

PDF 104 KB Posted

Attached to
UV Sterilizer Federal contract opportunity
Solicitation number
12305B22Q0061
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Northeast Area

View the file

Other files for this federal contract opportunity

Other files attached to UV Sterilizer, newest first.
File Type Posted
UV Sterilizer QandA updated.pdf PDF
Specification Sheet.pdf PDF
Specification Sheet.pdf PDF
Sterilizer Clauses.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

UV Sterilizer Q&A

-1. Please confirm that the intent of item number 4 is that the control panel cabinets for the two new UV sterilizers are to be NEMA 4X? Yes, this is an absolute requirement as they will be mounted in wet spaces with possible splash.

In the Specification Sheet, Specifications for (4) Kits section. As the existing Emperor Aquatic model COM6260-STD are obsolete and the company no longer exists can the following information be provided:

• Length of the existing 65 Watt UV bulbs? 61” see attached photos

• Length and diameter of the existing quartz sleeves? 1” diameter, 65” length; the unit housing is made from 6” PVC and is approximately 6’ long.

-2. Would the government consider a quotation for the 2 new UV systems only, and not the 4 rebuild kits? No Or is it an all or nothing purchase? It is an all or nothing purchase.

File details come from the government source that posted it. Updated .