Brand_Name_Specifications.docx
DOCX document 149 KB Posted
- Attached to
- Upgrades to 2 Linkam stage systems Federal contract opportunity
- Solicitation number
- 0040396025
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Brand_Name_Salient_Characteristics.docx | DOCX document | |
| Clauses_and_Provisions_SBSA.rtf | RTF text file |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
BRAND NAME SPECIFICATIONS
The sellers of the Brand Name system upgrades shall offer following Brand Name Specifications in order to meet or exceed the attached Brand Name Salient Characteristics.
This purchase is for two Linkam stage upgrades for our existing Linkam TMSG600 stages (SNs. 06R2599 and 06K1739). One includes software, controller, connector, and pump. The second is the same four items listed above with the addition of a camera.
Upgrade software (of which we require 2 copies) must do the following:
· Be compatible with our existing Linkam TMSG600 stages (SNs. 06R2599 and 06K1739)
· Easy to use and intuitive interface
· Full control of all Linkam Stages
· Manual and automated control of temperature and other parameters
· Real Time chart of temperature and other measured parameters
· Row by Row control of Vacuum, Humidity, Tensile Force and Shear modes controllable and synchronised with Temperature
· Parameters of connected Linkam sensors displayed and stored - e.g. Vacuum, Humidity, Force, XY position (dependent on connected stage type)
· Trigger I/O for synchronisation with other equipment
· Triggers recorded and plotted row by row
· A profile can contain up to 100 ramps
· Full five point temperature calibration
· Data export to standard CSV format for further analysis
· USB 2.0 to T96
· Native 64 bit application The upgraded controllers (of which we require two units) must:
· Be compatible with our existing Linkam TMSG600 stages (SNs. 06R2599 and 06K1739)
· Temperature ramps of up to 200°C/min (dependant on stage attached).
· Temperature measurement using a PT100 platinum Sensor with better than 0.01°C resolution.
· The standard USB interface allows the controller to be connected to a PC.
The upgraded connector (of which we require two) must:
· Connect our existing Linkam TMSG600 stages (SNs. 06R2599 and 06K1739) to the upgraded controllers.
The upgraded pump (of which we require two units) must:
· Be compatible with our existing Linkam TMSG600 stages (SNs. 06R2599 and 06K1739)
· Precisely control the liquid nitrogen flow in the Linkam TMSG600 stages (SNs. 06R2599 and 06K1739) we already have.
· Enable the stages to be controlled at linear cooling rates as fast as 100°C/min or as slow as 0.01°C/min.
· Be compatible with the updated Linkam controller.
· allow cooling right down to -196°C.
Camera (of which we require one) upgrade must:
· Be compatible with the upgraded LINK software.
· 2.8 MP
· Color
· USB 3.0
· 2/3" CCD
Include set-up and user training, including travel to Denver Federal Center.
image1.emf
File details come from the government source that posted it.